Thursday, January 19, 2023, 09:57
Cartridge #8 loaded with 198 discs
To be reduced inside a Ta container tube. Tube wrapped in 3 Ta foil heatshields.
Reduction for 4 hours and then 10 hours @ 425A
Good Vacuum: 5.0e-04 Torr
Bad Vacuum: 5.0e-02 Torr
Initial pressure is 1.41e-06 Torr
Thursday, January 19, 2023, 10:02 Auto ramping to 120A @ 1A/min
Thursday, January 19, 2023, 18:58
Autoramping to 300A @ 4/min (2A/30s)
IGP1=3.34e-06 Torr
Friday, January 20, 2023, 07:37
left at 300 A for 8 hours and pressure dropped. Restarted autoramp at 2A/30s and pressure was increasing until 425A where it peaked. Reduced autoramp to 1A/min and observed slow but definite pressure drop. Will go to 450A and leave there for 4 hours.
Currently 440A, 5.3V on TGHT and 1.28e-05 Torr on IGP1
Friday, January 20, 2023, 12:06
After 4 hours at 450A IGP1= 3.49e-06 Torr
TGHT = 450A 5.4V
Will drop to 425A and leave for 10 hours
Friday, January 20, 2023, 21:53
TGHT 425A, 5.2V
IGP1= 1.81e-06 Torr
autoramping to 0 A @ -2A/min |