140
Thu Jul 20 13:21:57 2006
Willy Andersson
Quench Recovery
started // warm return flow.
TIC121 was down to 27K
warm return valves were wide open
opened supply valves CV143 and CV146.
141
Thu Jul 20 13:27:21 2006
Willy Andersson
Quench Recovery
cold return mode continues...
cr_c1 and cr_12 are both at about 45K and
heading slowly down, so the
gamble of going to // cold return worked.
142
Thu Jul 20 13:34:26 2006
Willy Andersson
Quench Recovery
cold return mode continues...
Even though cr_c1 and cr_c12 have begun to
tick down, cr_c7 and cr_c6 are
actually ticking up. This is probably a wave
143
Fri Jul 21 07:39:37 2006
Doug Preddy
Quench Recovery
Cool down continues
Raised PIC229 to 11Bar. Started to close HIC155.
Cooling rates improve.
144
Fri Jul 21 14:08:04 2006
Doug Preddy
Quench Recovery
switch to series mode
Coils temps <10K.Switching to series mode.
145
Fri Jul 21 17:18:19 2006
Willy Andersson
Quench Recovery
Opened HIC142 fully open
Power leads have flatlined but still building
LHe level.
146
Fri Jul 21 23:36:21 2006
Doug Preddy
Quench Recovery
Final cooldown problems
80K shield is down to <60K Liquid level not
building. Power leads rippling
and warming. T1 inlet 85K with 10%heater
147
Sat Jul 22 06:34:14 2006
Experimentor
Quench Recovery
looking good
Donald Arseneau reporting.
It is looking pretty good, so I came back
148
Fri Jul 28 20:39:40 2006
Doug Preddy
Quench Recovery
Restarting refrigerator
Solenoid quenched at about 6:00 tonight.
I restarted the compressor, then the cold
box and turbines. When cold box
149
Sat Jul 29 05:30:28 2006
Doug Preddy
Quench Recovery
switched to series mode
Temps bottomed out at ~16K. Switching to series
mode.
150
Sat Jul 29 09:38:06 2006
Doug Preddy
Quench Recovery
power leads cooled
opened power lead flow about 40 minutes ago.
~12K will let them stablize.
LHE level >20%
151
Tue Aug 8 07:27:49 2006
Doug Preddy
Quench Recovery
Solenoid Quench at about 6:00 this morning
No indication of problem. Turbines still running,
but no flow to solenoid.
Started flow. Solenoid ~20K.
152
Wed Aug 9 06:44:58 2006
Willy Andersson
Quench Recovery
// cold mode cooling continues.
After restarting flow to the solenoid yesterday
in // cold return mode
TIC121 just refused to go lower than about
153
Wed Aug 9 06:59:43 2006
Willy Andersson
Quench Recovery
more // cold return stuff
About 21:00 last night the solenoid temps
had not changed much, pretty well
flat lined. I guess my reducing CV155 stalled
154
Wed Aug 9 07:50:49 2006
Doug Preddy
Quench Recovery
Switched to series mode.
Cool down continues.
156
Thu Aug 10 16:33:11 2006
Willy Andersson
Quench Recovery
Flow to solenoid tripped on powerlead temperature.
Power lead temps were drifting up, tried to
tweak TIC121 and HIC141 to
unstall powerlead flow. Got quench. Restarted
157
Thu Aug 10 20:26:09 2006
Willy Andersson
Quench Recovery
solenoid energized
Powerleads seem stable.
158
Tue Aug 15 08:45:56 2006
Doug Preddy
Quench Recovery
restarted the compressor
Solenoid tripped off Sunday. Restarted compressor
and cool down of cold box.
159
Tue Aug 15 15:39:27 2006
Willy Andersson
Quench Recovery
switched to parallel cold return
Return flow temps from solenoid about 40K,
TIC121 at about 40K so switching
to //cold return.
160
Thu Aug 17 08:04:44 2006
Doug Preddy
Quench Recovery
Solenoid tripped at about 3:00 this morning
restarting system.
Compressor would not start. Idid not see
anything that would not let it