| ID | Date | Author | Operational Mode | Subject | Text |  | 
| 166 | Thu Oct  5 14:53:54 2006 | Doug Preddy | Start Up | Switched to series mode |  |  | 
| 165 | Thu Oct  5 11:51:53 2006 | Doug Preddy | Start Up | Adde 2 bottles UHP Helium | Valved off purifiers, but left LN2 on. |  | 
| 164 | Thu Oct  5 10:26:26 2006 | Doug Preddy | Start Up | switched to parallel cold return |  |  | 
| 163 | Wed Oct  4 18:10:18 2006 | Willy Andersson | Start Up | Quick check on system | System in good // warm return mode, temps coming down nicely. BCP2 cryopump
 does not look too healthy, temperature is
 |  | 
| 162 | Wed Oct  4 07:29:04 2006 | Doug Preddy | Start Up | Starting cool down | Started compressor yesterday with flow through the purifier overnight.
 Starting tubines this morning in parallel
 |  | 
| 161 | Thu Aug 17 19:45:27 2006 | Willy Andersson | Quench Recovery | switch to series cooling | Temps of flow returning from solenoid flat lined at about 10K.
 Gradually closed CV155.
 |  | 
| 160 | Thu Aug 17 08:04:44 2006 | Doug Preddy | Quench Recovery | Solenoid tripped at about 3:00 this morning | restarting system. Compressor would not start. Idid not see
 anything that would not let it
 |  | 
| 159 | Tue Aug 15 15:39:27 2006 | Willy Andersson | Quench Recovery | switched to parallel cold return | Return flow temps from solenoid about 40K, TIC121 at about 40K so switching
 to //cold return.
 |  | 
| 158 | Tue Aug 15 08:45:56 2006 | Doug Preddy | Quench Recovery | restarted the compressor | Solenoid tripped off Sunday. Restarted compressor and cool down of cold box.
 |  | 
| 157 | Thu Aug 10 20:26:09 2006 | Willy Andersson | Quench Recovery | solenoid energized | Powerleads seem stable. |  | 
| 156 | Thu Aug 10 16:33:11 2006 | Willy Andersson | Quench Recovery | Flow to solenoid tripped on powerlead temperature. | Power lead temps were drifting up, tried to tweak TIC121 and HIC141 to
 unstall powerlead flow. Got quench. Restarted
 |  | 
| 155 | Thu Aug 10 08:33:44 2006 | Doug Preddy | Normal Running | Running low on helium | Temperture okay, but there are fluctuation in the power leads. PIC229 is
 trying to charge. The buffer tank pressure
 |  | 
| 154 | Wed Aug  9 07:50:49 2006 | Doug Preddy | Quench Recovery | Switched to series mode. | Cool down continues. |  | 
| 153 | Wed Aug  9 06:59:43 2006 | Willy Andersson | Quench Recovery | more // cold return stuff | About 21:00 last night the solenoid temps had not changed much, pretty well
 flat lined. I guess my reducing CV155 stalled
 |  | 
| 152 | Wed Aug  9 06:44:58 2006 | Willy Andersson | Quench Recovery | // cold mode cooling continues. | After restarting flow to the solenoid yesterday in // cold return mode
 TIC121 just refused to go lower than about
 |  | 
| 151 | Tue Aug  8 07:27:49 2006 | Doug Preddy | Quench Recovery | Solenoid Quench at about 6:00 this morning | No indication of problem. Turbines still running, but no flow to solenoid.
 Started flow. Solenoid ~20K.
 |  | 
| 150 | Sat Jul 29 09:38:06 2006 | Doug Preddy | Quench Recovery | power leads cooled | opened power lead flow about 40 minutes ago. ~12K will let them stablize.
 LHE level >20%
 |  | 
| 149 | Sat Jul 29 05:30:28 2006 | Doug Preddy | Quench Recovery | switched to series mode | Temps bottomed out at ~16K. Switching to series mode.
 |  | 
| 148 | Fri Jul 28 20:39:40 2006 | Doug Preddy | Quench Recovery | Restarting refrigerator | Solenoid quenched at about 6:00 tonight. I restarted the compressor, then the cold
 box and turbines. When cold box
 |  | 
| 147 | Sat Jul 22 06:34:14 2006 | Experimentor | Quench Recovery | looking good | Donald Arseneau reporting. 
 It is looking pretty good, so I came back
 |  |