Solenoid logbook, Page 5 of 15
New
|
Reply
|
Find
|
Delete
|
Last day
|
Help
|
Config
Full
| Summary |
Threaded
286 Entries
Goto page
Previous
1
,
2
,
3
,
4
, 5,
6
...
13
,
14
,
15
Next
All
ID
Date
Author
Operational Mode
Subject
Text
206
Wed Jul 4 11:13:12 2007
Doug Preddy
Quench Recovery
test
This is a test to make sure the e-mail works.
205
Sun Jul 1 16:09:37 2007
Doug Preddy
Normal Running
M9B:DP2 tripped off
Diffusion pump tripped off and vacuum (and
power lead temps) going up. I
had an experimentor cycle HIC 142 at 4:15
204
Tue Jun 19 21:15:41 2007
Doug Preddy
Quench Recovery
Today's quench
power lead temps caused a quench at 8:30 this
morning. Recovery of system
and flow to solenoid was started by 11:00.
203
Sat Jun 16 15:05:03 2007
Doug Preddy
Normal Running
Solenoid power lead temp problem
Power lead temps rippling and increasing.
I asked an operator to cycle
HIC142. This seems to have fixed the problem
202
Thu Jun 14 00:10:10 2007
Doug Preddy
Quench Recovery
recovering from earlier quench
Solenoid quenched on power lead temperature
last night at about 6:00. I
restarted flow to the solenoid in parallel
201
Tue Jun 12 12:48:25 2007
Doug Preddy
Maintenance
added one bottle UHP helium to system
200
Thu May 31 11:51:49 2007
Doug Preddy
Quench Recovery
Switched to series mode
199
Thu May 31 07:11:05 2007
Doug Preddy
Quench Recovery
switched to parallel cold return
switched to parallel cold return earlier this
morning. Vacuum is 5 x 10-5.
198
Wed May 30 11:27:32 2007
Doug Preddy
Quench Recovery
Problems recovering
Various start up problems :
MV51 stuck open. Finally got it closed by
hitting the valve with a hammer.
197
Wed May 30 08:38:38 2007
Doug Preddy
Quench Recovery
Solenoid quenched this morning
Solenoid quenched this morning on power lead
temp (I believe).
I restarted it at ~7:00, but the capacity
196
Wed May 16 10:02:35 2007
Doug Preddy
Maintenance
warming up power leads
Power lead temps have been slowly increasing
over the last wekk. I ramped
the solenoid down and am warming the power
195
Wed May 2 16:28:16 2007
Doug Preddy
Normal Running
Energizing solenoid to 500 amps
Power supply initially would not ramp up.
I exercised the rate control
selector switch and found that the power
194
Mon Apr 30 09:30:26 2007
Doug Preddy
Normal Running
problems over weekend
The PLC Limit switch power supply was tripped
off on Saturday. I came in
and reset it and restarted all the vacuum
193
Fri Apr 27 10:32:07 2007
Doug Preddy
Quench Recovery
Reset quench relay logic box
The quench relay logic controller was not
functioning correctly. It showed
a "Trip" light and no other lights. I cycled
192
Fri Apr 27 10:16:44 2007
Kevin Trithardt
Quench Recovery
Test
Don't Panic -- There was no quench. I am testing
the e-log.
I have added Kevin's name and e-mail to the
191
Fri Apr 27 10:10:07 2007
Doug Preddy
Normal Running
Solenoid Cold
Solenoid cold, power leads cold. Alarm handler
started.
190
Tue Apr 24 10:21:49 2007
Doug Preddy
Start Up
Started flow to turbines
Started flow to turbines at about 10:00 in
parallel warm mode.
189
Mon Apr 23 16:31:02 2007
Doug Preddy
Start Up
Started compressor
Startewd compressor and purifer today.
188
Fri Apr 20 12:15:12 2007
Doug Preddy
Start Up
remote PLC switch
I could not control the vacuum devices using
EPICS. The buttons could be
pushed, but nothing would respond. The PLC
187
Thu Apr 19 15:18:25 2007
Doug Preddy
Start Up
preparing to start the refrigerator
Starting the vacuum systems in the coldbox
and solenoid.