83
Wed Nov 24 07:28:38 2004
Other
Mission Impossible
180
Tue Nov 7 10:35:32 2006
Doug Preddy
Normal Running
Maintenance day
I ramped down the solenoid power suplly and
shut off flow to the power
leads. I will allow them to warm up and (hopefully)
262
Sun Oct 5 11:12:48 2008
Doug Preddy
Normal Running
M9B:V1 interlocks and a IOC crash
M9A:V1 was removed this shutdown. It's "in"
limit is required to allow M9B:V1 to be opened.
I have bypassed the interlocks on M9B:V1
205
Sun Jul 1 16:09:37 2007
Doug Preddy
Normal Running
M9B:DP2 tripped off
Diffusion pump tripped off and vacuum (and
power lead temps) going up. I
had an experimentor cycle HIC 142 at 4:15
111
Wed Nov 2 10:17:27 2005
Willy Andersson
Start Up
M9B helium refrigerator and solenoid cooldown
Started M9B MYCOM helium compressor circulating
helium through system last
Friday. Right hand purifier was regenerated
51
Fri Oct 1 11:39:55 2004
Doug Preddy
Maintenance
Loading of compressor not working
When the compressor was restarted on Monday,
the remote load button would
not respond. The compressor was shut off.
129
Fri Jun 16 08:12:04 2006
Doug Preddy
Normal Running
LI108 indicates high on EPICS screen
LI108 indicates high on EPICS screen (150%)
Controller shows ~64%
258
Tue Sep 9 15:04:23 2008
Doug Preddy
Maintenance
Jumpered interlock for diff pump water flow
Jumpered interlock for diff pump water flow
in Proteus box.
9
Wed Apr 21 06:48:31 2004
Doug Preddy
Start Up
JT bypass closed
Closed cv155
28
Fri May 21 09:27:39 2004
Doug Preddy
Maintenance
Idle mode
Power supply was ramped down yesterday as
there are no users.
The Sundance server was rebooted yesterday.
249
Sun Jul 13 20:17:24 2008
Doug Preddy
Other
Helium refrigerator problems
In early June the helium compressor was running
with the solenoid cold. A vavlve broke off
at the suction side of the compressor. The
284
Mon Aug 9 14:27:12 2010
Doug Preddy
Shut Down
Helium plant shut down
156
Thu Aug 10 16:33:11 2006
Willy Andersson
Quench Recovery
Flow to solenoid tripped on powerlead temperature.
Power lead temps were drifting up, tried to
tweak TIC121 and HIC141 to
unstall powerlead flow. Got quench. Restarted
179
Fri Nov 3 12:09:51 2006
Doug Preddy
Normal Running
Fine tuning
The power lead temperatures have been climbing
at a rate of ~ 0.02K very 8
hours. The negative lead started to show
1
Fri Dec 19 11:34:10 2003
Doug Preddy
Quench Recovery
Final home of this E-log
http://isac.triumf.ca/m9sol/
Chris has set up the M9 solenoid e-log at
the above web address. Play with
146
Fri Jul 21 23:36:21 2006
Doug Preddy
Quench Recovery
Final cooldown problems
80K shield is down to <60K Liquid level not
building. Power leads rippling
and warming. T1 inlet 85K with 10%heater
195
Wed May 2 16:28:16 2007
Doug Preddy
Normal Running
Energizing solenoid to 500 amps
Power supply initially would not ramp up.
I exercised the rate control
selector switch and found that the power
41
Sat Aug 7 23:40:47 2004
Willy Andersson
Quench Recovery
Energizing solenoid
23:39 Powerleads cold, LHe level over 40%
and rising. Energizing solenoid
to 500 Amps, ramp rate 60. Checking compressor
178
Tue Oct 31 08:37:34 2006
Doug Preddy
Quench Recovery
Energizing solenoid
211
Sun Jul 29 04:58:27 2007
Doug Preddy
Quench Recovery
Energizing solenoid
LHe level~50%. Still have a slight ripple
in the power leads, but they are
cold enough to power the solenoid. LHe_in