M9sol Logbooks Solenoid Daily log
  Solenoid logbook, Page 11 of 15  ELOG logo
Entry  Wed May 30 08:38:38 2007, Doug Preddy, Quench Recovery, Solenoid quenched this morning 
Solenoid quenched this morning on power lead temp (I believe).
I restarted it at ~7:00, but the capacity control for the compressor would 
not respond. I shut the system back down and checked as many relays as I 
could. I then restarted the system and it is now cooling. I will go 
through the wiring diagrams and see if I can locate the bad relay.
Entry  Wed May 30 11:27:32 2007, Doug Preddy, Quench Recovery, Problems recovering 
Various start up problems :
MV51 stuck open. Finally got it closed by hitting the valve with a hammer.
Vacuum is bad in cold box as MV902 tripped closed.
Vacuum in solenoid has gone bad. Suspect boil off from inside solenoid. So 
far it is not recovering.
Previously T1 would not spin up - frozen.
Entry  Thu May 31 07:11:05 2007, Doug Preddy, Quench Recovery, switched to parallel cold return 
switched to parallel cold return earlier this morning. Vacuum is 5 x 10-5.
Entry  Thu May 31 11:51:49 2007, Doug Preddy, Quench Recovery, Switched to series mode 
 
Entry  Tue Jun 12 12:48:25 2007, Doug Preddy, Maintenance, added one bottle UHP helium to system 
 
Entry  Thu Jun 14 00:10:10 2007, Doug Preddy, Quench Recovery, recovering from earlier quench 
Solenoid quenched on power lead temperature last night at about 6:00. I 
restarted flow to the solenoid in parallel cold return. Switched to series 
mode and am building liquid. 
Entry  Sat Jun 16 15:05:03 2007, Doug Preddy, Normal Running, Solenoid power lead temp problem 
Power lead temps rippling and increasing. I asked an operator to cycle 
HIC142. This seems to have fixed the problem (for now).
Entry  Tue Jun 19 21:15:41 2007, Doug Preddy, Quench Recovery, Today's quench 
power lead temps caused a quench at 8:30 this morning. Recovery of system 
and flow to solenoid was started by 11:00. Solenoid was switched to series 
cold at 3:00. Solenoid and power leads were cold at ~6:30. power supply 
was to be energized by experimenters.
Entry  Sun Jul 1 16:09:37 2007, Doug Preddy, Normal Running, M9B:DP2 tripped off  
Diffusion pump tripped off and vacuum (and power lead temps) going up. I 
had an experimentor cycle HIC 142 at 4:15 this morning. When I got in 
today I restarted the pump and it is back to 6.0 x 10-7. Power lead temps 
also are going down.
Entry  Wed Jul 4 11:13:12 2007, Doug Preddy, Quench Recovery, test 
This is a test to make sure the e-mail works.
Entry  Thu Jul 5 03:04:50 2007, Doug Preddy, Normal Running, Power lead problems 
The power leads temps are starting to cycle again. Earlier today M9 
experimentor cycled HIC142 to stablize the oscillations. They started 
again early this morning, so cycled HIC142 again. Oscillations died back, 
but are still present. Will try and warm the power leads later today when 
the experiment's sample is changed.
Entry  Sat Jul 28 15:45:17 2007, Doug Preddy, Quench Recovery, quench recovery 
The solenoid tripped off this morning due to pwer lead over temp. The 
compressor was also off. I restarted this morning at about 8:15 in 
parallel cold return. Switching to series mode at 3:40.
Entry  Sat Jul 28 22:30:13 2007, Doug Preddy, Quench Recovery, quench again 
recovered system and energized solenoid at about 6:00 this evening.
At 9:30 power lead temps quenched solenoid again. MV52 stuck open during 
start-up.
Entry  Sun Jul 29 04:58:27 2007, Doug Preddy, Quench Recovery, Energizing solenoid 
LHe level~50%. Still have a slight ripple in the power leads, but they are 
cold enough to power the solenoid. LHe_in varies between 10.7K - 11.3K.
Entry  Mon Aug 6 08:41:31 2007, Doug Preddy, Quench Recovery, Cold box trip at 6:00 
compressor off due to cold box trip at 6:00 this morning. Restarted system 
and started flow to solenoid in parallel cold return.
Entry  Tue Aug 14 10:15:22 2007, Doug Preddy, Maintenance, Solenoid ramped down 
Solenoid ramped down and locked out for M9A slits work.
Entry  Tue Aug 14 12:55:12 2007, Doug Preddy, Normal Running, Strip tool saves 
There are two strip tools saved:
coils
powerlead
Entry  Wed Sep 12 12:01:02 2007, Doug Preddy, Maintenance, regenerated adsorber 
regenerated adsorber
replaced MV902 "ON" switch (used nitrogen precooling on switch)
Entry  Fri Sep 14 12:44:52 2007, Doug Preddy, Start Up, started compressor 
running purifier over weekend
Entry  Wed Sep 19 20:53:18 2007, Doug Preddy, Maintenance, More diff pump news 
Cold box pumped out to ~ 190 mTorr overnight.Leak checked the pump again. 
Still leaks at dip stick.Tried resetting the o-ring on the dip stick. It 
still leaks baddd!!!
Machined the old seal off and installed a smaller o-ring onto dip stick. 
Installed 60 ml of new pump oil into pump. Pumping down again. 
After 4 hours the cold box is down to 160 mTorr.