126
|
Wednesday, July 04, 2012, 15:37 |
Anders Mjos | FEBIAD anode end cap test | | | | | | In progress | Pumping down |
Replaced grid containing ring (target side) with new ring. No contact between target and anode measured in lab.
NOTE: While closing up the VCR fittings the tube rotated causing the tube to flatten in a bend. This may reduce water flow on the heat shield circuit. (see attachment)
Wednesday, July 04, 2012, 15:33: Started roughing.
Wednesday, July 04, 2012, 15:57 : Started turbos
Wednesday, July 04, 2012, 16:03: Turbos at speed. Vacuum at 3E-5 |
127
|
Thursday, July 05, 2012, 08:41 |
Anders Mjos | FEBIAD anode end cap test | | | | | | In progress | Water opened - reduced flow heat shield |
Anders Mjos wrote: |
Replaced grid containing ring (target side) with new ring. No contact between target and anode measured in lab.
NOTE: While closing up the VCR fittings the tube rotated causing the tube to flatten in a bend. This may reduce water flow on the heat shield circuit. (see attachment)
Wednesday, July 04, 2012, 15:33: Started roughing.
Wednesday, July 04, 2012, 15:57 : Started turbos
Wednesday, July 04, 2012, 16:03: Turbos at speed. Vacuum at 3E-5
|
Thursday, July 05, 2012, 08:39: Water supply open. Vacuum stable at 5.9E-7. Due to the kink in the heat shield cooling line, flow in the heat shield line is lower than other lines (visually inspecting speed of paddle wheels). |
128
|
Thursday, July 05, 2012, 10:59 |
Friedhelm Ames | | | | | | | Problems/Solutions | water tripping in front of HV power supplies |
Found some water tripping in front of the high voltage power supplies rack from the water lines at the ceiling. Plastic sheets have been put on top of the racks and MCR has been informed. |
129
|
Thursday, July 05, 2012, 11:00 |
Friedhelm Ames | | | | | | | In progress | source started up again |
tube heater ramping started |
130
|
Thursday, July 05, 2012, 18:02 |
Friedhelm Ames | | | | | | | Problems/Solutions | fluctuations in magnet current |
The current at the magnet power supply TMS:MB1 shows fluctuations every 1-2 minutes. The same can be seen in the Hall probe read back and in the beam current at Fc6. |
131
|
Thursday, July 05, 2012, 18:10 |
Friedhelm Ames | | | | | | | Problems/Solutions | fluctuations in magnet current |
Friedhelm Ames wrote: |
The current at the magnet power supply TMS:MB1 shows fluctuations every 1-2 minutes. The same can be seen in the Hall probe read back and in the beam current at Fc6.
|
a fault report has been issued to power supplies group. Dan Louie will look into it. |
132
|
Friday, July 06, 2012, 11:04 |
Friedhelm Ames | | | | | | | | source tripped off over night |
A big vacuum excursion tripped the source off at about 1am. The vacuum did not fully recover afterwards only down to about 1.3e-6T. Ne leak has been closed and vacuum went down to 6e-7T within a few minutes. After the trip the current of the magnet stayed constant. Fluctuations may have been caused by high voltage sparks. |
133
|
Friday, July 06, 2012, 11:05 |
Friedhelm Ames | | | | | | | | start source again |
heater ramping up again with Ne leak closed. |
134
|
Friday, July 06, 2012, 11:18 |
Friedhelm Ames | | | | | | | | TSG:TP1B off source off again. |
turbo pump TSG:TP1B has tripped off. Cannot be restarted. Fault report to vacuum. Everything switched off again
|
135
|
Friday, July 06, 2012, 15:36 |
Friedhelm Ames | | | | | | | Done | switched everything off |
switched everything off to prepare for scheduled power outage on Monday |
136
|
Monday, July 09, 2012, 11:14 |
Aurelia Laxdal | | | | | | | | Test Stand status |
Restarted the PC at the Test Stand.
Restarted the Archivers; currently pumping down the separator area. |
137
|
Monday, July 09, 2012, 12:57 |
Anders Mjos | | | | | | | Done | Vented |
|
138
|
Tuesday, July 10, 2012, 16:01 |
Anders Mjos | | | | | | | Problems/Solutions | Turbo replaced - cleaning chamber recommended |
A new turbo has been installed. Eddie said the other one had crashed but not catastrophically. He recommends the test stand be wiped clean before pumping because there is a fine layer of Aluminum dust on the chamber base.
|
139
|
Wednesday, July 11, 2012, 11:16 |
Anders Mjos | | | | | | | In progress | Pumping down - with new turbo pump |
|
140
|
Wednesday, July 11, 2012, 15:30 |
Anders Mjos | | | | | | | Problems/Solutions | Restarting turbos after power bump |
Wednesday, July 11, 2012, 15:27: Started TSGTP1A and 1B
Wednesday, July 11, 2012, 15:31: Started TSMTP2 and 6
Wednesday, July 11, 2012, 15:33: TMS tubos at speed. Vacuum 4.8E-7
Wednesday, July 11, 2012, 15:34: TSG turbos at speed. Vacuum 9.5E-6 |
141
|
Thursday, July 12, 2012, 09:05 |
Anders Mjos | | | | | | | Problems/Solutions | Shut down TSGTP1A - bearing noise |
Shut down TSGTP1A due to noise (possibly from bearings) after consulting Edi. |
142
|
Thursday, July 12, 2012, 09:44 |
Anders Mjos | | | | | | | In progress | Venting |
System will be vented and gas lines purged with air, then nitrogen.
Thursday, July 12, 2012, 09:43: Shut down TSGTP1B. |
143
|
Thursday, July 12, 2012, 12:23 |
Anders Mjos | | | | | | | Done | TSH:SV5 bypass on |
Thursday, July 12, 2012, 15:48: bypass removed |
144
|
Thursday, July 12, 2012, 13:41 |
Anders Mjos | | | | | | | Done | purging gas lines with air |
- gas lines purged with air for 60 min (through TSHUHV1 and TSHSV4).
- gas line allowed to fill with nitrogen to 300 torr and evacuated. repeated 3 times.
|
145
|
Thursday, July 12, 2012, 13:49 |
Anders Mjos | | | | | | | Problems/Solutions | Aluminum oxide found inside TSHSV1 |
Aluminum oxide found inside TSHSV1 and the copper line to TSHUHV1.
SV1 replaced and copper line cleaned. |