Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  Test-Stand  Not logged in ELOG logo
Message ID: 607     Entry time: Friday, October 06, 2017, 12:14
Author: Aaron Schmidt 
Target/No.#: SiC#37 
Source: HP-FEBIAD 
Station: ITE 
Module: TM4 
Target Oven W.O#: 39530-10 
Heat Shield W.O#: 41751-4 
Status: In progress 
Subject: Conditioning SiC#37 

NOTE* : The right coil terminal terminates into the heat shield. DO NOT TURN ON COIL.

Friday, October 06, 2017, 12:12: Roughing. * Note on installation. Middle TBHT prong 'stuck' to insulator slightly and was moved to fit. Fit ok in Jig.

Friday, October 06, 2017, 13:53: Turbo pumps on.

Friday, October 06, 2017, 14:00: Ion gauge, RGA on.


Tuesday, October 10, 2017, 07:45: Waterlines opened. No response on RGA. Vacuum ok @ 1.35e-07T.

Tuesday, October 10, 2017, 07:49: RGA Fatal communications error. Restart RGA.

Tuesday, October 10, 2017, 07:57: Power/sources on.

Tuesday, October 10, 2017, 08:17: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Tuesday, October 10, 2017, 08:57: BIAS / EZL / EE @ 0 kV. Power/sources off. Locked out*

Tuesday, October 10, 2017, 09:30:  Install/swap gas bottles. Elog.

Tuesday, October 10, 2017, 10:36: Power on. Gas bottles opened. Await vacuum stabilization.

Tuesday, October 10, 2017, 10:40: Sources on.

Tuesday, October 10, 2017, 10:50: BIAS / EZL / EE @ 30 / 13 / 1.5 kV .

Unable to bring Anode to voltage. Power supply was changed and does not 'Float'. Test TGHT & TBHT. Then switch from Cathode Negative Bias to Anode positive*.

Tuesday, October 10, 2017, 11:37: Auto ramping TGHT / TBHT to 20 / 20 A @ 4A per min. (TEST then cool)

Tuesday, October 10, 2017, 11:45: Auto-cooling

Tuesday, October 10, 2017, 11:57: Sources off / BIAS / EZL / EE @ 0 kV.


Wednesday, October 11, 2017, 10:25: Anode & PS switched*. Power/Sources on.

Wednesday, October 11, 2017, 10:38: BIAS / EZL / EE @ 30 / 13 / 1.5 kV.

Wednesday, October 11, 2017, 10:40: Coil @ 0 A. Anode 200 V / 7.326mA.

Wednesday, October 11, 2017, 10:45: Gas bottles opened. Await vacuum stabilization.

Wednesday, October 11, 2017, 10:47: IV4 Opened. Vacuum @ 2.43e-07T

Device Temp
[C]
 Flow
[lpm]
 HS + MSP 24.6 3.78
 TBHT 24.6 2.75
 COIL + EE 24.4 4.84
 TGHT  24.6 5.44

Wednesday, October 11, 2017, 10:50: Auto ramping TGHT / TBHT to 580 / 290 A @ 2A per min.

 Wednesday, October 11, 2017, 14:21: Check. Vacuum @ 2.99e-06T. Outgassing.

Wednesday, October 11, 2017, 14:42: Note* Slight discrepancy on the TBHT. Direct read out off the PS @ 296A. Source page: TBHT set to 290.0A, shows 302.2A


Device Temp
[C]
 Flow
[lpm]
 HS + MSP 31.4 3.84
 TBHT 25.4 2.98
 COIL + EE 25.0 4.86
 TGHT 27.1 5.45

Thursday, October 12, 2017, 09:00. Mass scans ok. Spark @ 8:57.

Thursday, October 12, 2017, 10:48: Auto ramping TBHT to 310 A @ 2A per min.


Friday, October 13, 2017, 07:15: Anode short determined. Plotted. Occurred around 18:00 (6pm) yesterday.

Friday, October 13, 2017, 07:20: Auto-cooling @ 4A per min

Friday, October 13, 2017, 10:03: Sources off. Gas lines closed.

Friday, October 13, 2017, 10:06: RGA off. IV4 Closed. Reset IV4

Friday, October 13, 2017, 10:08: Ion gauge off. Turbos off.

Friday, October 13, 2017, 11:10: Venting.

Friday, October 13, 2017, 11:45: power off. Waterlines purged.

Friday, October 13, 2017, 11:49: Doors open,

Attachment 1: TS-2017-10-10-07-55.rga  2.260 MB  Uploaded Thursday, October 19, 2017, 12:02
Attachment 2: TS-2017-10-06-14-57.rga  2.728 MB  Uploaded Thursday, October 19, 2017, 12:02
Attachment 3: SiC#37.docx  2.980 MB  Uploaded Thursday, October 19, 2017, 12:02
Attachment 4: ER-580-290-30kV-m23.txt  3 kB  Uploaded Thursday, October 19, 2017, 12:02  | Show | Hide all
Attachment 5: ER-580-290-30kV-m28.txt  3 kB  Uploaded Thursday, October 19, 2017, 12:02  | Hide | Hide all
value (Voltage offset       0.0000)
     80.0000
      0.7000
    100.0000
  29978.0220
21
21
./emit.dat
 -0.1282 -0.0800 -0.0717 -0.0702 -0.0705 -0.0699 -0.0690 -0.0702 -0.0699 -0.0351 -0.0140 -0.0296 -0.0507 -0.0610 -0.0668 -0.0677 -0.0684 -0.0681 -0.0745 -0.0726 -0.0717
 -0.0699 -0.0748 -0.0656 -0.0641 -0.0671 -0.0668 -0.0644 -0.0638 -0.0699 -0.0656 -0.0214 -0.0247 -0.0476 -0.0589 -0.0610 -0.0623 -0.0647 -0.0653 -0.0702 -0.0687 -0.0690
 -0.0357 -0.0717 -0.0674 -0.0684 -0.0711 -0.0677 -0.0656 -0.0656 -0.0684 -0.0708 -0.0439 -0.0330 -0.0516 -0.0595 -0.0598 -0.0623 -0.0647 -0.0665 -0.0647 -0.0659 -0.0644
 -0.0412 0.0220 0.0235 -0.0403 -0.0690 -0.0687 -0.0717 -0.0726 -0.0711 -0.0650 -0.0677 -0.0546 -0.0531 -0.0549 -0.0604 -0.0613 -0.0650 -0.0632 -0.0620 -0.0632 -0.0696
 -0.0687 -0.0037 0.1111 0.1181 0.1062 -0.0452 -0.0793 -0.0809 -0.0781 -0.0735 -0.0708 -0.0690 -0.0635 -0.0595 -0.0610 -0.0601 -0.0644 -0.0620 -0.0595 -0.0644 -0.0674
 -0.0659 -0.0665 0.0455 0.2451 0.3592 0.4355 0.2582 -0.0067 -0.0882 -0.0809 -0.0751 -0.0705 -0.0677 -0.0653 -0.0626 -0.0616 -0.0647 -0.0601 -0.0616 -0.0650 -0.0684
 -0.0653 -0.0674 -0.0616 0.0934 0.4333 0.8362 1.0184 0.7513 0.4086 -0.0378 -0.0739 -0.0684 -0.0674 -0.0644 -0.0665 -0.0635 -0.0629 -0.0632 -0.0641 -0.0656 -0.0653
 -0.0586 -0.0656 -0.0638 -0.0665 0.2002 0.7745 1.6849 1.4682 0.9906 0.5087 0.2045 -0.0400 -0.0708 -0.0659 -0.0681 -0.0638 -0.0638 -0.0656 -0.0647 -0.0647 -0.0659
 -0.0195 -0.0208 -0.0259 -0.0400 -0.0351 0.4218 1.4456 2.9980 2.0459 1.0242 0.3888 0.1514 0.0729 -0.0595 -0.0699 -0.0632 -0.0653 -0.0653 -0.0662 -0.0635 -0.0650
 0.0217 0.0214 0.0552 0.0967 0.1715 0.2869 0.9854 3.1738 5.3613 2.0422 0.7935 0.2029 0.0650 0.0204 0.0153 -0.0613 -0.0629 -0.0653 -0.0729 -0.0632 -0.0665
 -0.0531 0.0513 0.0540 0.0793 0.1791 0.3333 0.7407 1.7371 5.2942 3.7866 1.0818 0.2380 0.0311 0.0104 0.0009 -0.0113 -0.0110 -0.0659 -0.0641 -0.0635 -0.0662
 -0.0690 -0.0616 -0.0299 0.0650 0.1306 0.1947 0.3656 0.7361 1.8848 5.6970 1.7746 0.5069 0.0656 -0.0485 -0.0485 -0.0412 -0.0375 -0.0284 -0.0421 -0.0641 -0.0681
 -0.0659 -0.0638 -0.0629 -0.0610 -0.0171 0.1617 0.2716 0.4413 0.8774 2.1158 4.6158 1.0266 0.3339 0.0119 -0.0644 -0.0638 -0.0632 -0.0549 -0.0473 -0.0375 -0.0510
 -0.0671 -0.0650 -0.0653 -0.0717 -0.0595 -0.0552 0.0220 0.2460 0.5835 0.9680 1.9342 2.9599 0.6287 0.2295 -0.0299 -0.0650 -0.0644 -0.0632 -0.0598 -0.0580 -0.0543
 -0.0668 -0.0644 -0.0653 -0.0656 -0.0616 -0.0626 -0.0626 -0.0513 -0.0192 0.3293 0.8240 1.3589 1.2018 0.3604 0.1443 -0.0436 -0.0623 -0.0623 -0.0693 -0.0659 -0.0671
 -0.0638 -0.0638 -0.0656 -0.0607 -0.0604 -0.0589 -0.0589 -0.0568 -0.0543 -0.0497 -0.0223 0.1840 0.3668 0.3961 0.1917 0.0891 -0.0522 -0.0641 -0.0635 -0.0629 -0.0656
 -0.0668 -0.0641 -0.0613 -0.0693 -0.0558 -0.0522 -0.0519 -0.0510 -0.0543 -0.0558 -0.0555 -0.0562 -0.0571 0.0098 0.1544 0.0937 0.0403 -0.0644 -0.0729 -0.0632 -0.0690
 -0.0699 -0.0653 -0.0595 -0.0653 -0.0568 -0.0482 -0.0433 -0.0430 -0.0494 -0.0534 -0.0549 -0.0677 -0.0644 -0.0647 -0.0601 -0.0476 0.0348 -0.0266 -0.0638 -0.0647 -0.0684
 -0.0674 -0.0659 -0.0684 -0.0674 -0.0571 -0.0488 -0.0391 -0.0314 -0.0409 -0.0491 -0.0522 -0.0641 -0.0610 -0.0653 -0.0632 -0.0629 -0.0616 -0.0632 -0.0677 -0.0632 -0.0677
 -0.0665 -0.0647 -0.0702 -0.0693 -0.0568 -0.0479 -0.0354 -0.0201 -0.0180 -0.0406 -0.0543 -0.0583 -0.0635 -0.0604 -0.0629 -0.0613 -0.0629 -0.0647 -0.0632 -0.0644 -0.0674
 -0.0635 -0.0623 -0.0595 -0.0647 -0.0537 -0.0458 -0.0323 -0.0116 0.0067 -0.0241 -0.0494 -0.0580 -0.0659 -0.0604 -0.0626 -0.0616 -0.0626 -0.0616 -0.0665 -0.0632 -0.0677
Attachment 6: ER-580-290-30kV-m38.txt  3 kB  Uploaded Thursday, October 19, 2017, 12:02  | Show | Hide all
Attachment 7: ER-580-310-30kV-m23.txt  3 kB  Uploaded Thursday, October 19, 2017, 12:03  | Show | Hide all
Attachment 8: ER-580-310-30kV-m38.txt  3 kB  Uploaded Thursday, October 19, 2017, 12:03  | Show | Hide all
Attachment 9: ER-580-310-30kV-m202.txt  3 kB  Uploaded Thursday, October 19, 2017, 12:03  | Show | Hide all
ELOG V2.9.2-2455