28
|
Wednesday, September 12, 2012, 01:53 |
John Wong | Nb#7 | LP-SIS | | | 39165-6 | | Spent | Nb#7 target description |
Wednesday, September 12, 2012, 02:10
Nb#7 target description:
Total Nb foils = 400 discs
Total Nb mass (Nb foils + Ta wires) = 26.529g
Total length (after filled) = 4.6cm
(see attachment #1) |
Attachment 1: Nb#7_target_description.jpg
|
|
29
|
Thursday, October 11, 2012, 11:41 |
John Wong | NiO#1 | HP-FEBIAD | ITE | TM4 | 38481-7 | 39069-1/10(lid) | Spent | NiO target description |
(Should have posted this before my long vacation... Anyway, it's for documentation)
Thursday, October 11, 2012, 11:38
NiO/Ni #1 |
Average |
NiO + Ni (g) |
0.109 |
NiO - Ni foil (g) |
0.06 |
NiO + Ni thickness (mm) |
0.116 |
NiO - Ni Thickness (mm) |
0.091 |
Area (cm2) |
2.212 |
NiO + Ni - organics 8.95% (g) |
0.099 |
NiO - Ni - Organics (g) |
0.055 |
NiO Density (g/cm3) |
2.145 (32%) |
Target contains 500 NiO/Ni discs and is measured at 9.0 in length. |
31
|
Tuesday, March 19, 2013, 13:32 |
John Wong | SiC#28-RFQ | LP-SIS | | TM4 | LP-RFQ (NEW) | | Spent | SiC#28-RFQ target description |
Tuesday, March 19, 2013, 12:28
SiC#28 - RFQ target description:
Total SiC foils = 325 discs
Total SiC + C-grv foils = 38.48 g
Total length = 10.9cm
Average mass of SiC/disc (without graphite foil) = 0.083 g
Average thickness of SiC (without graphite foil) = 0.18 mm
No Rh foil added
See attachment #1: SiC#28 RFQ density measurements
See link for more details: https://documents.triumf.ca/docushare/dsweb/View/Collection-9771
|
Attachment 1: SiC#28_RFQ_density_measurements.xlsx
|
32
|
Thursday, April 11, 2013, 14:30 |
John Wong | Ta#39 | LP-SIS | ITW | TM1 | 39165-7 | | Spent | Ta#39 target description |
Ta#38 LP-SIS target description:
Total Ta foils = 400
Foil Thickness = 0.025mm
Target density = 16.69 g/cm3
Target length = 4.1cm
Total Mass (400 foils + Ta wire) = 43.15g
Disc area = 2.12 cm2
Ta thickness = 20.354g/cm2
See link for more details:
http://documents.triumf.ca/docushare/dsweb/View/Collection-9934
|
33
|
Thursday, May 30, 2013, 16:58 |
John Wong | UCx#6 | LP-SIS | | TM1 | 39165-6 | 39230-8 | Spent | UCx#6 - target description |
Thursday, May 30, 2013, 16:57
UCx#6 target specification, see attachment #1.
The target is currently undergoing the final conditioning in the Evaporator#1.
Wednesday, June 12, 2013, 17:10
Target is assembled and stored in the Ar glove box.
It's ready to go into the Hotcell.
|
Attachment 1: UCx#6_specifications.PNG
|
|
34
|
Wednesday, July 10, 2013, 18:41 |
John Wong | SiC#29 | HP-FEBIAD-CTL | ITE | TM4 | 38481-6 | 39956-4 | Spent | SiC#29 - Target description |
Wednesday, July 10, 2013, 18:32
SiC#29 HP-FEBIAD-CTL target description:
Total SiC foils = 510 discs
Total mass SiC + C_grv foils = 62 g (before sintering, expecting ~13% loss of organic mass after burnt off)
Average mass SiC/disc (without C_grv) = 0.0892 g
Average thickness SiC (without C_grv) = 0.21mm
No Rh foil and salt added for this run.
See attachment #1 for the density measurement.
Wednesday, July 10, 2013, 18:40
Target is currently in the Test-Stand. Once it's tested; it is ready to go online.
Happy Beaming!
|
Attachment 1: SiC#29_density_measurements_final.xlsx
|
35
|
Friday, August 30, 2013, 12:08 |
John Wong | UCx#7 | LP-SIS | ITW | TM1 | 38634-7 | 39230-4 | Spent | UCx#7 - target description |
Friday, August 30, 2013, 12:06
UCx#7 target specification, see attachment #1.
The target is assembled and going online soon.
|
Attachment 1: UCx#7_specifications.PNG
|
|
36
|
Friday, August 30, 2013, 14:45 |
John Wong | NiO#2 | HP-FEBIAD | | | 39530-6 | 39069-4 | Spent | NiO target description |
Friday, August 30, 2013, 14:38
NiO/Ni#2 Average
NiO+ Ni (g) 0.114
NiO - Ni foil (g) 0.066
NiO + Ni thickness (mm) 0.116
NiO - Ni thickness (mm) 0.104
Area (cm2) 2.212
NiO + Ni - organics (8.95%) (g) 0.104
NiO - Ni - organics (8.95%) (g) 0.060
NiO Density (g/cm3) 2.110
% Yield density 32%
# discs in target 530
Total target length (cm) 9.0
|
37
|
Friday, October 25, 2013, 17:35 |
John Wong | Ta#41 | HP-SIS | ITW | TM4 | 38817-10 | | Spent | Ta#41 target description |
Friday, October 25, 2013, 17:35
Ta#41 HP-SIS target description:
Total Ta foils = 400
Foil Thickness = 0.025mm
Target density = 16.69g/cm3
Target length = 4.0cm
Total Mass (400 foils + Ta wire) = 42.19g
|
38
|
Wednesday, November 27, 2013, 17:56 |
John Wong | UCx#8 | LP-IG-LIS | ITE | TM4 | 39814-IV | | Spent | UCx#8 target description |
Wednesday, November 27, 2013, 17:49
UCx#8 target description - see attachment#1
|
Attachment 1: UCx_8_specification.PNG
|
|
39
|
Monday, March 17, 2014, 16:15 |
John Wong | Ta#42 | LP-SIS | ITW | TM4 | 40282-1 | 39956-1 | Spent | Ta#42 Target description |
Monday, March 17, 2014, 16:14
Ta#42 LP-SIS target description:
Total Ta foils = 400
Foil Thickness = 0.025mm
Target density = 16.69 g/cm3
Target length = 4.2cm
Total Mass (400 foils + Ta wire) = 45.899g
|
40
|
Monday, March 31, 2014, 15:02 |
John Wong | TiC#4 | HP-SIS | | | 38817-8 | 39956-10 | Spent | Target Description |
Monday, March 31, 2014, 15:00
Target consists of 360 discs of TiC + graphite foils.
Total length = 11.7cm
Total green mass = 40.945g
(see attachment for more details)
|
Attachment 1: TiC#4_description.JPG
|
|
41
|
Wednesday, April 30, 2014, 16:36 |
John Wong | SiC#30 | HP-FEBIAD | ITE | TM3 | 38481-8 | | Spent | SiC#30 Target description |
Wednesday, April 30, 2014, 15:16
Note:
This target was built in 2011. The endcaps were replaced after they got vaporized in Evap#1 during the sintering process, and the target was re-conditioned.
See elogs for more details:
https://elog.triumf.ca/TIS/Evaporator_1/20
The target was inspected and was good for online use.
Total SiC foils = 555 discs
Total length (packed) = 17.5cm
Total SiC mass (plus graphite foils) = 52.902g
See attachment for density measurements
|
Attachment 1: SiC#30_density_measurements.xlsx
|
45
|
Monday, July 28, 2014, 15:57 |
John Wong | ZrC#6 | LP-SIS | ITW | TM1 | 40282-3 | | Spent | ZrC target description |
Monday, July 28, 2014, 15:52
See attachment#1 for ZrC target description.
Target is currently undergoing the sintering process
|
Attachment 1: ZrC#6.JPG
|
|
61
|
Friday, March 11, 2016, 10:18 |
Laura Lambert | Ta # 47 | | ITE | TM4 | | | Spent | Ta # 47 target desscription |
Friday, March 11, 2016, 10:13
Ta#47 LP-SIS [new small center block] target description:
Total Ta foils = 470
Foil Thickness = 0.025mm
Target density = 15.05 g/cm3
Target length = 3.4cm [exactly 7.8cm space on both sides]
Total Mass = 51.1533g [Incl foils + Ta wire on 6 packs]
Note: Target consists of 350 single discs and 6x20 packs of Ta discs
|