ID Date Authorup Target/No.# Source Station Module Target Oven W.O# Heat Shield W.O# Status Subject
  74   Tuesday, May 30, 2017, 15:15 Laura LambertSiC#36HP-FEBIAD-CTL TM241369-10 ReadySiC#36 target description

 Tuesday, May 30, 2017, 15:12

See attachment for target description.


 

Attachment 1: SiC#36_Target_Description.JPG
SiC#36_Target_Description.JPG
  75   Wednesday, May 31, 2017, 09:53 Laura LambertTa# 52LP-SIS TM442218-14 In progressTa#52 target description

 Wednesday, May 31, 2017, 09:51

Ta#52 LP-SIS target description:

Total Ta foils = 400 [4x20 packs + 320 single discs]

Foil Thickness = 0.025mm

Target density = 14.77 g/cm3

Target length = 2.9cm

Total Mass = 42.8228g [Incl foils + Ta wire on 4 packs]


 

  76   Monday, June 26, 2017, 13:36 Laura LambertTa #53HP-SIS TM242218-6 ReadyTa#53 target description

 Monday, June 26, 2017, 13:32

Ta#53 HP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density = 14.80 g/cm3

Target length = 3.4cm

Total Mass = 50.3105g [Incl foils + Ta wire on 6 packs]


 

  78   Monday, August 21, 2017, 11:13 Laura LambertUCx#21LP-SIS  42218-15 ReadyUCx#21 Target Description

 Monday, August 21, 2017, 11:09

Attached please find the UCx#21 target description.


 

Attachment 1: target_description.JPG
target_description.JPG
  79   Tuesday, September 26, 2017, 12:07 Laura LambertTa#54HP-SISITWTM242218-10 ReadyTa#54 target description

Tuesday, September 26, 2017, 09:31

Ta#54 HP-SIS target description:

Total Ta foils = 470 [6x20 packs + 350 single discs]

Foil Thickness = 0.025mm

Target density = 14.72 g/cm3

Target length = 3.4cm

Total Mass = 50.0372g [Incl foils + Ta wire on 6 packs]


 

  80   Wednesday, September 27, 2017, 09:14 Laura LambertSiC#37HP-FEBIADITETM439530-10 ReadySiC#37 target description

 Wednesday, September 27, 2017, 09:13

See attachment for target description.


 

Attachment 1: Target_Description.JPG
Target_Description.JPG
  81   Monday, April 16, 2018, 09:55 Laura LambertUCx#22LP-SIS  42218-13 ReadyUCx#22 Target Description

 Monday, April 16, 2018, 09:53

Attached please find the UCx#22 target description.


 

Attachment 1: target_description.JPG
target_description.JPG
  105   Tuesday, March 16, 2021, 17:05 Luca EgoritiITE-TM4-UC#32-P2N-SISHP-SISITETM4--In progressUO2 discs cutting

This afternoon, Darwin prepared in the hot lab fume hood the two UO2 slurries to be cut into P2N target discs. Later, I took over cutting both slurries. A total of 49 annular discs were cut and loaded into the P2N graphite container.

The discs thickness was measured by stacking 5 discs on each other, and measuring their thickness in two points.

#discs = 49

Average disc thickness = 0.453mm

Empty cartridge weight = 9.42 g

Full cartridge weight = 39.86 g

UO2+C weight = 30.44 g

The unused material plus the cutting scrap were left in a zip lock bag on a shelf in the right hand side of the fume hood. The center scraps and the loaded target were placed in a plastic tray on the top shelf. The used tools and scales were left on the right of the working area, the napkins on the bottom were put into the garbage bag (inside the fume hood) and replaced, while the only napkin left is the one with the target information (recorded in attached picture).

Attachment 1: UC32-P2N_material_info.jpg
UC32-P2N_material_info.jpg
Attachment 2: Fume_hood.jpg
Fume_hood.jpg
  88   Thursday, November 29, 2018, 14:23 Marla CervantesUCx#26LP-IG-LIS  42281-3 ReadyUCx#26 Target description

Thursday, November 29, 2018, 14:20

Attached please find the UCx#26 target description. 


 

Attachment 1: UCx_26-description.JPG
UCx_26-description.JPG
  90   Monday, July 22, 2019, 16:33 Marla CervantesUCx#27LP-SIS  49044-9 ReadyUCx#27 Target Description

 Monday, July 22, 2019, 16:31

Attached please find the UCx#27 target description. 


 

Attachment 1: UCx#27.JPG
UCx#27.JPG
  93   Wednesday, October 02, 2019, 11:18 Marla CervantesUCx#28LP-SIS  44062-4 ReadyUCx #28 Target description

 Wednesday, October 02, 2019, 11:17

Attached please find the UCx#28 target description. 

Attachment 1: UCx28-target_description.JPG
UCx28-target_description.JPG
  100   Thursday, June 04, 2020, 15:00 Marla CervantesC#1LP-SIS  43044-8 In progressTarget C#1 Description

 Thursday, June 04, 2020, 15:00

See attached picture for description 

Attachment 1: C1-Description.PNG
C1-Description.PNG
  101   Tuesday, September 29, 2020, 11:50 Marla CervantesC#2LP-SISITWTM244139-9 In progressTarget C#2 Description

 Tuesday, September 29, 2020, 11:49

See attached picture for description 

Attachment 1: C2-Description.JPG
C2-Description.JPG
  118   Tuesday, June 21, 2022, 15:34 Matthew GareauSiC#42, SiC#43  TM2  Problems/SolutionsMultipin guide pin issues

 While removing the SiC#42 it became apparent that the multipin guide pin had become loose again and was not allowing the target to dismount from the module.

The multipin shoulder bolts were removed and then the multipin was able to be pried off.

However the NHC manipulator were not able to reach the guide pin to remove it, which would be required before the SiC#43 could be installed.

The target module was then moved to the SHC where the guide pin could be removed and the installation of SiC#43 took place, along with the transfer of the parasitic attachment onto the SiC#43.

This problem seems to suggest that the multipin could use some redesign to ensure smoother target exchanges.

  26   Wednesday, August 01, 2012, 09:37 Pierre BricaultITE-TM4-UO#3LP-FEBIADITETM4  Problems/SolutionsShort has developed while lowering the cathode heater

 On tuesday July we lower the cathode heater to swap the anode-cathode bias. While lowering the cathode heater a short on the anode developed. 

The polarity swap was not possible because the target/tube assembly has also a short to common. 

We return the configuration has it was and we added two 100 Ohms: 5 W resistor between the common and the heaters to prevent current to flow back to the common. 

The anode short disappear during the cathode and target heating. 