Message ID: 553     Entry time: Wednesday, November 20, 2013, 13:52
Author: Bevan Moss 
Category: Conditioning Station 
Type: Standard Operation 
Module: TM2 
Target/Number:  
Subject: HV testing of TM2 

 HV testing without the cover continued today. With the EINZEL lens at 30kV the target module was able to achieve 58.7kV. At this voltage any spark was reeking havoc on the electronics of the TCS. Vacuum gauges signals were occasionally lost, the thermocouple reading would occasionally be lost, turbo pump would turn off. All of these events caused the interlocks to trip resulting in the power supplies to be turned off. As a result I bypassed majority of the interlocks to prevent this from happening. Ultimately as spark took out the CANBUS controller first for the EINZEL lens and then for the BIAS. David Morris has been contacted to have them replaced. The damage of the CANBUS controller for these power supplies is becoming a common problem and should be investigated for if there is a better solution. During testing I experimented with raising the BIAS to near its limit followed by the EINZEL lens and vise versa. From this it was determined that the EINZEL must be raised to high voltage before the BIAS otherwise sparking occurs. Even a step from 0 to 500 volts on the ENIZEL lens with the BIAS at 55kV caused sparking. There are a few combinations of testing along this line I would like to continue with but initially my though is that the BIAS and ENIZEL do not have sufficient clearance for the BIAS to operate above 55kV. Attached is a screen shot of the strip tool and the data. I have not heard from David yet but I do not expect the TCS to be operational today. 

Attachment 1: 20112013_HV_conditioning_TM2.JPG  223 kB  | Show | Hide all
Attachment 2: 20112013_TM2_No_source_tray_HV_Test_no_cover.txt  2.648 MB  | Hide | Hide all
Time	TCS:IG1:RDVAC [T]	TCS:IG2:RDVAC [T]	TCS:IG1S:RDVAC [T]	TCS:BIAS:RDVOL [V]	TCS:BIAS:RDCUR [uA]	TCS:EL:RDVOL [V]	TCS:EL:RDCUR [mA]	
11/20/2013 10:34:48.414676	                      0.000000253                      0.000003120                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:34:49.413724	                      0.000000253                      0.000003098                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:34:50.414780	                      0.000000256                      0.000003120                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:34:51.414583	                      0.000000258                      0.000003120                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:34:52.414423	                      0.000000260                      0.000003120                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:34:53.415237	                      0.000000258                      0.000003120                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:34:54.416059	                      0.000000258                      0.000003098                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:34:55.506062	                      0.000000258                      0.000003098                           BadVal                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:34:56.506814	                      0.000000258                      0.000003120                      0.000001230                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:34:57.506697	                      0.000000256                      0.000003098                      0.000001230                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:34:58.506780	                      0.000000258                      0.000003120                      0.000001230                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:34:59.506511	                      0.000000260                      0.000003120                      0.000001252                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:00.507321	                      0.000000258                      0.000003120                      0.000001230                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:01.507271	                      0.000000258                      0.000003120                      0.000001230                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:02.507196	                      0.000000256                      0.000003098                      0.000001230                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:03.507937	                      0.000000256                      0.000003098                      0.000001230                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:04.508830	                      0.000000256                      0.000003098                      0.000001230                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:05.509815	                      0.000000253                      0.000003120                      0.000001230                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:06.510560	                      0.000000258                      0.000003120                      0.000001208                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:07.511442	                      0.000000260                      0.000003120                      0.000001208                           BadVal                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:08.511382	                      0.000000258                      0.000003098                      0.000001208                            27.46                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:09.511380	                      0.000000260                      0.000003120                      0.000001252                            26.32                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:10.511063	                      0.000000258                      0.000003120                      0.000001252                            26.32                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:11.510938	                      0.000000256                      0.000003098                      0.000001252                            27.46                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:12.510832	                      0.000000256                      0.000003120                      0.000001208                            27.46                           BadVal                           BadVal                           BadVal
11/20/2013 10:35:13.510688	                      0.000000258                      0.000003120                      0.000001208                            27.46                             0.42                           BadVal                           BadVal
11/20/2013 10:35:14.510527	                      0.000000258                      0.000003098                      0.000001230                            27.46                             0.42                           BadVal                           BadVal
11/20/2013 10:35:15.510410	                      0.000000258                      0.000003120                      0.000001230                            27.46                             0.42                           BadVal                           BadVal
11/20/2013 10:35:16.510321	                      0.000000256                      0.000003098                      0.000001230                            27.46                             0.42                           BadVal                           BadVal
11/20/2013 10:35:17.510214	                      0.000000256                      0.000003098                      0.000001230                            26.32                             0.42                           BadVal                           BadVal
11/20/2013 10:35:18.510025	                      0.000000256                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                           BadVal
11/20/2013 10:35:19.509934	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                           BadVal
11/20/2013 10:35:20.509860	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                           BadVal
11/20/2013 10:35:21.511728	                      0.000000258                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                           BadVal
11/20/2013 10:35:22.511632	                      0.000000260                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                           BadVal
11/20/2013 10:35:23.511584	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:24.591659	                      0.000000258                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:25.729112	                      0.000000253                      0.000003098                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:26.729088	                      0.000000258                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:27.729906	                      0.000000253                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:28.729930	                      0.000000260                      0.000003120                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:29.729660	                      0.000000258                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:30.729489	                      0.000000258                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:31.729372	                      0.000000258                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:32.730305	                      0.000000253                      0.000003098                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:33.853965	                      0.000000256                      0.000003120                      0.000001252                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:34.854006	                      0.000000258                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:35.949159	                      0.000000258                      0.000003098                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:36.949749	                      0.000000256                      0.000003098                      0.000001252                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:37.949644	                      0.000000256                      0.000003120                      0.000001252                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:38.949535	                      0.000000256                      0.000003120                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:39.950466	                      0.000000256                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:41.28198	                      0.000000256                      0.000003098                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:42.34724	                      0.000000256                      0.000003098                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:43.34971	                      0.000000256                      0.000003098                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:44.34865	                      0.000000260                      0.000003098                      0.000001252                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:45.34739	                      0.000000258                      0.000003120                      0.000001252                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:46.34576	                      0.000000258                      0.000003098                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:47.34484	                      0.000000256                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:48.35380	                      0.000000258                      0.000003120                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:49.35302	                      0.000000256                      0.000003120                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:50.36136	                      0.000000253                      0.000003120                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:35:51.35938	                      0.000000253                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:52.35873	                      0.000000260                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:53.35802	                      0.000000260                      0.000003098                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:54.35700	                      0.000000260                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:55.35442	                      0.000000260                      0.000003098                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:56.35354	                      0.000000256                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:57.35238	                      0.000000258                      0.000003098                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:58.36111	                      0.000000253                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:35:59.35984	                      0.000000260                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:00.35832	                      0.000000260                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:01.35689	                      0.000000260                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:02.144399	                      0.000000256                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:03.144478	                      0.000000258                      0.000003098                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:04.144502	                      0.000000256                      0.000003098                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:05.145332	                      0.000000258                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:06.145089	                      0.000000253                      0.000003098                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:07.145209	                      0.000000260                      0.000003098                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:08.145838	                      0.000000260                      0.000003120                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:09.146802	                      0.000000260                      0.000003098                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:10.146584	                      0.000000258                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:11.146460	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:12.146294	                      0.000000256                      0.000003098                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:13.146305	                      0.000000258                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:14.146042	                      0.000000258                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:15.145915	                      0.000000258                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:16.145794	                      0.000000258                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:17.145705	                      0.000000256                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:18.145906	                      0.000000258                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:19.145434	                      0.000000258                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:20.145323	                      0.000000253                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:21.145202	                      0.000000256                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:22.145076	                      0.000000260                      0.000003098                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:23.146118	                      0.000000260                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:24.145825	                      0.000000258                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:25.155723	                      0.000000258                      0.000003120                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:26.155570	                      0.000000258                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:27.155437	                      0.000000256                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:28.155279	                      0.000000256                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:29.155155	                      0.000000256                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:30.155132	                      0.000000256                      0.000003098                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:31.155083	                      0.000000258                      0.000003098                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:32.154768	                      0.000000258                      0.000003098                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:33.154675	                      0.000000258                      0.000003098                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:34.154594	                      0.000000253                      0.000003120                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:35.155716	                      0.000000256                      0.000003120                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:36.156306	                      0.000000258                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:37.156322	                      0.000000256                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:38.156054	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:39.155899	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:40.313843	                      0.000000258                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:41.313756	                      0.000000253                      0.000003120                      0.000001208                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:42.313502	                      0.000000253                      0.000003098                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:43.313407	                      0.000000256                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:44.313288	                      0.000000258                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:45.313123	                      0.000000258                      0.000003098                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:46.313028	                      0.000000256                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:47.312863	                      0.000000256                      0.000003098                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:48.313869	                      0.000000258                      0.000003120                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:49.314706	                      0.000000256                      0.000003098                      0.000001230                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:50.315547	                      0.000000253                      0.000003120                      0.000001186                            26.32                             0.42                             5.49                             0.01
11/20/2013 10:36:51.315404	                      0.000000260                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:52.315285	                      0.000000256                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:53.316161	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:54.316185	                      0.000000260                      0.000003098                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:55.317074	                      0.000000256                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:56.316803	                      0.000000258                      0.000003120                      0.000001208                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:57.316655	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:58.316082	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:36:59.316373	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:37:00.316469	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             5.49                             0.01
11/20/2013 10:37:01.316114	                      0.000000256                      0.000003120                      0.000001230                            27.46                             0.42                             0.61                             0.01
11/20/2013 10:37:02.316025	                      0.000000260                      0.000003098                      0.000001230                            26.32                             0.42                             0.61                             0.01
11/20/2013 10:37:03.316970	                      0.000000260                      0.000003120                      0.000001230                             8.01                             0.42                             0.61                             0.01
11/20/2013 10:37:04.316771	                      0.000000256                      0.000003120                      0.000001230                             8.01                             0.42                             0.61                             0.01
11/20/2013 10:37:05.316651	                      0.000000253                      0.000003098                      0.000001230                             9.15                             0.79                             4.27                             0.01
11/20/2013 10:37:06.316545	                      0.000000256                      0.000003098                      0.000001230                             9.15                             0.79                             4.27                             0.01
11/20/2013 10:37:07.316358	                      0.000000256                      0.000003120                      0.000001230                            12.58                             0.79                             6.10                             0.01
11/20/2013 10:37:08.317387	                      0.000000256                      0.000003120                      0.000001208                            12.58                             0.79                             5.49                             0.01
11/20/2013 10:37:09.318100	                      0.000000256                      0.000003076                      0.000001208                            10.29                             0.91                             5.49                             0.01
11/20/2013 10:37:10.317973	                      0.000000256                      0.000003098                      0.000001230                            12.58                             0.73                             5.49                             0.01
11/20/2013 10:37:11.317854	                      0.000000256                      0.000003120                      0.000001208                         12098.87                             6.77                            30.51                             0.01
11/20/2013 10:37:12.317785	                      0.000000253                      0.000003120                      0.000001208                          9037.53                             5.37                             8.54                             0.01
11/20/2013 10:37:13.318646	                      0.000000256                      0.000003120                      0.000001208                          6778.43                             4.08                             8.54                             0.01
11/20/2013 10:37:14.318517	                      0.000000258                      0.000003120                      0.000001208                          5105.28                             3.41                             3.05                             0.01
11/20/2013 10:37:15.318456	                      0.000000258                      0.000003120                      0.000001208                          3855.57                             2.86                             3.05                             0.01
11/20/2013 10:37:16.319259	                      0.000000256                      0.000003120                      0.000001230                          2921.72                             1.95                             0.61                             0.01
11/20/2013 10:37:17.319100	                      0.000000256                      0.000003120                      0.000001208                          2237.35                             1.58                             0.61                             0.01
11/20/2013 10:37:18.320004	                      0.000000253                      0.000003120                      0.000001230                          1826.50                             1.15                             1.83                             0.01
11/20/2013 10:37:19.320985	                      0.000000256                      0.000003120                      0.000001208                          1478.59                             1.15                             1.83                             0.01
11/20/2013 10:37:20.320764	                      0.000000258                      0.000003120                      0.000001208                          1197.07                             1.03                             3.66                             0.01
11/20/2013 10:37:21.320666	                      0.000000253                      0.000003120                      0.000001230                           969.32                             1.03                             3.66                             0.01
11/20/2013 10:37:22.321547	                      0.000000258                      0.000003120                      0.000001230                           785.07                             1.03                             4.27                             0.01
11/20/2013 10:37:23.322366	                      0.000000258                      0.000003120                      0.000001230                           635.15                             1.03                             4.27                             0.01
11/20/2013 10:37:24.322396	                      0.000000258                      0.000003120                      0.000001230                           513.84                             0.91                             4.27                             0.01
11/20/2013 10:37:25.322137	                      0.000000256                      0.000003120                      0.000001230                           415.42                             0.91                             4.27                             0.01
11/20/2013 10:37:26.321987	                      0.000000256                      0.000003142                      0.000001230                           335.31                             0.91                             4.88                             0.01
11/20/2013 10:37:27.321928	                      0.000000253                      0.000003120                      0.000001208                           270.08                             0.91                             4.88                             0.01
11/20/2013 10:37:28.322956	                      0.000000256                      0.000003098                      0.000001230                           216.29                             0.91                             4.88                             0.01
11/20/2013 10:37:29.323592	                      0.000000256                      0.000003098                      0.000001230                           179.67                             0.91                             4.88                             0.01
11/20/2013 10:37:30.323486	                      0.000000256                      0.000003142                      0.000001230                           141.90                             0.91                             4.88                             0.01
11/20/2013 10:37:31.323450	                      0.000000258                      0.000003120                      0.000001230                           108.72                             0.91                             4.88                             0.01
11/20/2013 10:37:32.323271	                      0.000000258                      0.000003120                      0.000001208                            89.26                             0.91                             4.88                             0.01
11/20/2013 10:37:33.324362	                      0.000000258                      0.000003098                      0.000001208                            70.95                             0.91                             4.88                             0.01
11/20/2013 10:37:34.324005	                      0.000000256                      0.000003120                      0.000001208                            53.78                             0.91                             5.49                             0.01
11/20/2013 10:37:35.323838	                      0.000000256                      0.000003120                      0.000001208                            53.78                             0.91                             1.83                             0.01
11/20/2013 10:37:36.324765	                      0.000000258                      0.000003120                      0.000001230                            33.18                             0.91                             1.83                             0.01
11/20/2013 10:37:37.324726	                      0.000000258                      0.000003120                      0.000001230                            17.16                             0.91                             1.22                             0.01
11/20/2013 10:37:38.325505	                      0.000000258                      0.000003120                      0.000001186                            17.16                             0.91                             1.22                             0.01
11/20/2013 10:37:39.325336	                      0.000000258                      0.000003098                      0.000001208                             4.57                             0.36                             1.22                             0.01
11/20/2013 10:37:40.325348	                      0.000000260                      0.000003120                      0.000001230                             4.57                             0.36                             1.22                             0.01
11/20/2013 10:37:41.325141	                      0.000000258                      0.000003120                      0.000001230                            12.58                             0.79                             1.22                             0.01
11/20/2013 10:37:42.325021	                      0.000000256                      0.000003120                      0.000001230                            12.58                             0.79                             1.22                             0.01
11/20/2013 10:37:43.325865	                      0.000000253                      0.000003120                      0.000001230                             9.15                             0.91                             0.61                             0.01
11/20/2013 10:37:44.325763	                      0.000000258                      0.000003120                      0.000001230                             9.15                             0.91                             0.61                             0.01
11/20/2013 10:37:45.325605	                      0.000000258                      0.000003098                      0.000001230                            10.29                             0.91                             0.61                             0.01
11/20/2013 10:37:46.325508	                      0.000000253                      0.000003120                      0.000001230                            10.29                             0.91                             0.61                             0.01
11/20/2013 10:37:47.325388	                      0.000000256                      0.000003098                      0.000001230                             5.72                             0.91                           998.55                             0.04
11/20/2013 10:37:48.326442	                      0.000000258                      0.000003120                      0.000001230                             5.72                             0.91                          3000.53                             0.09
11/20/2013 10:37:49.326181	                      0.000000256                      0.000003098                      0.000001252                             8.01                             0.91                          4500.80                             0.23
11/20/2013 10:37:50.325945	                      0.000000256                      0.000003120                      0.000001230                             8.01                             0.91                          5001.90                             0.23
11/20/2013 10:37:51.326819	                      0.000000256                      0.000003120                      0.000001230                             5.72                             0.91                          5568.93                             0.23
11/20/2013 10:37:52.326754	                      0.000000256                      0.000003120                      0.000001230                             5.72                             0.91                          7003.28                             0.23
11/20/2013 10:37:53.327593	                      0.000000256                      0.000003142                      0.000001230                             5.72                             0.91                          8003.66                             0.23
11/20/2013 10:37:54.327448	                      0.000000256                      0.000003120                      0.000001208                            13.73                             0.91                          9502.70                             0.53
11/20/2013 10:37:55.328351	                      0.000000260                      0.000003120                      0.000001230                            13.73                             0.91                         10506.14                             0.35
11/20/2013 10:37:56.328292	                      0.000000258                      0.000003120                      0.000001208                            17.16                             0.91                         11781.79                             0.38
11/20/2013 10:37:57.328116	                      0.000000256                      0.000003120                      0.000001208                            17.16                             0.91                         13006.17                             0.44
11/20/2013 10:37:58.328287	                      0.000000256                      0.000003120                      0.000001208                             9.15                             0.91                         14505.83                             0.44
11/20/2013 10:37:59.328830	                      0.000000256                      0.000003120                      0.000001208                             8.01                             0.91                         16503.54                             0.58
11/20/2013 10:38:00.328696	                      0.000000256                      0.000003120                      0.000001230                             8.01                             0.91                         18006.86                             0.58
11/20/2013 10:38:01.328586	                      0.000000256                      0.000003098                      0.000001230                            12.58                             0.91                         19506.52                             0.69
11/20/2013 10:38:02.329506	                      0.000000258                      0.000003098                      0.000001208                            12.58                             0.91                         21007.40                             0.79
11/20/2013 10:38:03.330336	                      0.000000258                      0.000003120                      0.000001208                            12.58                             0.91                         22507.66                             0.79
11/20/2013 10:38:04.330331	                      0.000000258                      0.000003098                      0.000001208                             4.57                             0.91                         23011.21                             0.80
11/20/2013 10:38:05.330125	                      0.000000256                      0.000003120                      0.000001208                             4.57                             0.91                         24009.15                             0.81
11/20/2013 10:38:06.329942	                      0.000000256                      0.000003120                      0.000001208                            12.58                             0.79                         25008.31                             0.81
11/20/2013 10:38:07.329857	                      0.000000258                      0.000003098                      0.000001208                            12.58                             0.79                         25010.14                             0.81
11/20/2013 10:38:08.330714	                      0.000000258                      0.000003098                      0.000001208                             5.72                             0.91                         25010.14                             0.81
11/20/2013 10:38:09.330602	                      0.000000260                      0.000003098                      0.000001230                             5.72                             0.91                         25011.97                             0.81
11/20/2013 10:38:10.330506	                      0.000000253                      0.000003098                      0.000001230                             5.72                             0.73                         25011.97                             0.81
11/20/2013 10:38:11.331338	                      0.000000258                      0.000003098                      0.000001230                             5.72                             0.73                         25011.97                             0.81
11/20/2013 10:38:12.331188	                      0.000000258                      0.000003120                      0.000001252                             6.86                             0.73                         25011.97                             0.81
11/20/2013 10:38:13.332192	                      0.000000258                      0.000003120                      0.000001252                             6.86                             0.73                         25011.97                             0.81
11/20/2013 10:38:14.331985	                      0.000000256                      0.000003120                      0.000001252                            10.29                             0.85                         25011.97                             0.81
11/20/2013 10:38:15.331812	                      0.000000256                      0.000003120                      0.000001208                            10.29                             0.85                         25012.58                             0.81
11/20/2013 10:38:16.331676	                      0.000000256                      0.000003120                      0.000001208                            13.73                             0.85                         25012.58                             0.81
11/20/2013 10:38:17.331599	                      0.000000258                      0.000003120                      0.000001208                            13.73                             0.85                         25012.58                             0.81
11/20/2013 10:38:18.331562	                      0.000000260                      0.000003098                      0.000001208                            10.29                             0.85                         25012.58                             0.81
11/20/2013 10:38:19.331349	                      0.000000256                      0.000003120                      0.000001230                            10.29                             0.85                         25012.58                             0.81
11/20/2013 10:38:20.331173	                      0.000000256                      0.000003120                      0.000001230                             8.01                             0.85                         25012.58                             0.81
11/20/2013 10:38:21.331084	                      0.000000258                      0.000003120                      0.000001230                            10.29                             0.85                         25012.58                             0.81
11/20/2013 10:38:22.331021	                      0.000000256                      0.000003120                      0.000001230                            10.29                             0.85                         25012.58                             0.81
11/20/2013 10:38:23.330936	                      0.000000256                      0.000003120                      0.000001230                             9.15                             0.85                         25012.58                             0.81
11/20/2013 10:38:24.330710	                      0.000000260                      0.000003120                      0.000001230                             9.15                             0.85                         25012.58                             0.81
11/20/2013 10:38:25.330593	                      0.000000256                      0.000003120                      0.000001208                             8.01                             0.73                         25012.58                             0.81
11/20/2013 10:38:26.330462	                      0.000000258                      0.000003120                      0.000001208                             8.01                             0.73                         25012.58                             0.81
11/20/2013 10:38:27.330338	                      0.000000256                      0.000003120                      0.000001208                            10.29                             0.91                         25012.58                             0.81
11/20/2013 10:38:28.331226	                      0.000000253                      0.000003120                      0.000001230                            10.29                             0.91                         25012.58                             0.81
11/20/2013 10:38:29.331092	                      0.000000258                      0.000003120                      0.000001230                            13.73                             0.91                         25012.58                             0.81
11/20/2013 10:38:30.331004	                      0.000000258                      0.000003098                      0.000001230                            13.73                             0.91                         25012.58                             0.81
11/20/2013 10:38:31.330795	                      0.000000258                      0.000003098                      0.000001208                             6.86                             0.91                         25012.58                             0.81
11/20/2013 10:38:32.330772	                      0.000000260                      0.000003120                      0.000001208                             6.86                             0.91                         25012.58                             0.81
11/20/2013 10:38:33.331565	                      0.000000260                      0.000003098                      0.000001208                             9.15                             0.79                         25012.58                             0.81
11/20/2013 10:38:34.331463	                      0.000000258                      0.000003098                      0.000001230                             9.15                             0.79                         25012.58                             0.81
11/20/2013 10:38:35.331347	                      0.000000256                      0.000003120                      0.000001208                            14.87                             0.61                         25012.58                             0.81
11/20/2013 10:38:36.331211	                      0.000000258                      0.000003120                      0.000001208                            14.87                             0.61                         25012.58                             0.81
11/20/2013 10:38:37.331045	                      0.000000258                      0.000003098                      0.000001230                            13.73                             0.85                         25012.58                             0.81
11/20/2013 10:38:38.330950	                      0.000000256                      0.000003098                      0.000001230                            13.73                             0.85                         25012.58                             0.81
11/20/2013 10:38:39.330833	                      0.000000258                      0.000003120                      0.000001186                             6.86                             0.73                         25012.58                             0.81
11/20/2013 10:38:40.330708	                      0.000000256                      0.000003120                      0.000001208                             6.86                             0.73                         25012.58                             0.81
11/20/2013 10:38:41.330581	                      0.000000258                      0.000003098                      0.000001230                             9.15                             0.85                         25012.58                             0.81
11/20/2013 10:38:42.330469	                      0.000000253                      0.000003098                      0.000001186                             6.86                             0.73                         25012.58                             0.81
11/20/2013 10:38:43.331408	                      0.000000256                      0.000003120                      0.000001230                             6.86                             0.85                         25012.58                             0.81
11/20/2013 10:38:44.331413	                      0.000000256                      0.000003120                      0.000001230                          1010.52                             2.99                         25012.58                             0.81
11/20/2013 10:38:45.332046	                      0.000000256                      0.000003120                      0.000001208                          1791.02                             5.61                         25012.58                             0.81
11/20/2013 10:38:46.331944	                      0.000000258                      0.000003098                      0.000001230                          2512.01                             3.90                         25012.58                             0.81
11/20/2013 10:38:47.331830	                      0.000000258                      0.000003120                      0.000001208                          3010.98                             9.94                         25012.58                             0.81
11/20/2013 10:38:48.332698	                      0.000000256                      0.000003120                      0.000001230                          4012.35                             4.51                         25012.58                             0.81
11/20/2013 10:38:49.332575	                      0.000000258                      0.000003120                      0.000001230                          5012.58                             7.75                         25012.58                             0.81
11/20/2013 10:38:50.332461	                      0.000000253                      0.000003098                      0.000001230                          6013.96                             6.65                         25012.58                             0.81
11/20/2013 10:38:51.332339	                      0.000000256                      0.000003098                      0.000001230                          7017.62                            11.41                         25012.58                             0.81
11/20/2013 10:38:52.332236	                      0.000000256                      0.000003098                      0.000001230                          8514.53                            10.43                         25012.58                             0.81
11/20/2013 10:38:53.333086	                      0.000000256                      0.000003098                      0.000001230                          9515.90                            17.09                         25012.58                             0.81
11/20/2013 10:38:54.332965	                      0.000000256                      0.000003098                      0.000001230                         11016.25                            14.95                         25012.58                             0.81
11/20/2013 10:38:55.332817	                      0.000000256                      0.000003120                      0.000001230                         12019.91                            16.35                         25012.58                             0.81
11/20/2013 10:38:56.332686	                      0.000000253                      0.000003098                      0.000001186                         13292.51                            16.41                         25012.58                             0.81
11/20/2013 10:38:57.332571	                      0.000000256                      0.000003120                      0.000001186                         14518.19                            15.93                         25012.58                             0.81
11/20/2013 10:38:58.333428	                      0.000000256                      0.000003120                      0.000001230                         15519.56                            15.68                         25012.58                             0.81
11/20/2013 10:38:59.333281	                      0.000000256                      0.000003120                      0.000001230                         16520.94                            16.11                         25012.58                             0.81
11/20/2013 10:39:00.333153	                      0.000000256                      0.000003098                      0.000001208                         18018.99                            24.23                         25012.58                             0.81
11/20/2013 10:39:01.333093	                      0.000000253                      0.000003142                      0.000001208                         19022.65                            22.70                         25012.58                             0.81
11/20/2013 10:39:02.333017	                      0.000000260                      0.000003120                      0.000001230                         20521.85                            21.30                         25012.58                             0.81
11/20/2013 10:39:03.333874	                      0.000000256                      0.000003098                      0.000001230                         22022.20                            23.25                         25012.58                             0.81
11/20/2013 10:39:04.333737	                      0.000000258                      0.000003098                      0.000001208                         23469.90                            26.06                         25012.58                             0.81
11/20/2013 10:39:05.333680	                      0.000000256                      0.000003120                      0.000001230                         24522.77                            24.71                         25012.58                             0.81
11/20/2013 10:39:06.333393	                      0.000000256                      0.000003120                      0.000001230                         26023.11                            28.99                         25012.58                             0.81
11/20/2013 10:39:07.333417	                      0.000000258                      0.000003120                      0.000001208                         27523.46                            29.35                         25012.58                             0.81
11/20/2013 10:39:08.334148	                      0.000000258                      0.000003120                      0.000001230                         29026.09                            31.31                         25012.58                             0.81
11/20/2013 10:39:09.334056	                      0.000000256                      0.000003120                      0.000001230                         30523.00                            32.53                         25012.58                             0.81
11/20/2013 10:39:10.333927	                      0.000000253                      0.000003120                      0.000001230                         31525.52                            31.67                         25012.58                             0.81
11/20/2013 10:39:11.333973	                      0.000000256                      0.000003120                      0.000001230                         32528.03                            30.15                         25012.58                             0.81
11/20/2013 10:39:12.334739	                      0.000000256                      0.000003120                      0.000001230                         33801.78                            36.19                         25012.58                             0.81
11/20/2013 10:39:13.335695	                      0.000000253                      0.000003120                      0.000001230                         35026.32                            35.33                         25012.58                             0.81
11/20/2013 10:39:14.335425	                      0.000000253                      0.000003120                      0.000001230                         36474.02                            39.06                         25012.58                             0.81
11/20/2013 10:39:15.336686	                      0.000000253                      0.000003098                      0.000001230                         37526.89                            38.20                         25012.58                             0.81
11/20/2013 10:39:16.337275	                      0.000000258                      0.000003098                      0.000001230                         38527.12                            38.20                         25012.58                             0.81
11/20/2013 10:39:17.338221	                      0.000000253                      0.000003098                      0.000001208                         39974.82                            39.61                         25012.58                             0.81
11/20/2013 10:39:18.338918	                      0.000000253                      0.000003098                      0.000001186                         41027.69                            41.44                         25012.58                             0.81
11/20/2013 10:39:19.338838	                      0.000000256                      0.000003098                      0.000001230                         42029.06                            41.74                         25012.58                             0.81
11/20/2013 10:39:20.338621	                      0.000000256                      0.000003098                      0.000001208                         43028.15                            49.80                         25012.58                             0.81
11/20/2013 10:39:21.338512	                      0.000000256                      0.000003098                      0.000001208                         44526.20                            53.40                         25012.58                             0.81
11/20/2013 10:39:22.338428	                      0.000000258                      0.000003098                      0.000001230                         45528.72                            48.34                         25012.58                             0.81
11/20/2013 10:39:23.339288	                      0.000000258                      0.000003098                      0.000001186                         46528.95                            48.46                         25012.58                             0.81
11/20/2013 10:39:24.339252	                      0.000000253                      0.000003098                      0.000001230                         47529.18                            52.36                         25012.58                             0.81
11/20/2013 10:39:25.339125	                      0.000000256                      0.000003098                      0.000001230                         48975.73                            52.67                         25012.58                             0.81
11/20/2013 10:39:26.338885	                      0.000000256                      0.000003120                      0.000001208                         50026.32                            52.85                         25012.58                             0.81
11/20/2013 10:39:27.338831	                      0.000000256                      0.000003098                      0.000001208                         51027.69                            55.35                         25012.58                             0.81
11/20/2013 10:39:28.339727	                      0.000000258                      0.000003120                      0.000001230                         52025.63                            56.09                         25012.58                             0.81
11/20/2013 10:39:29.339894	                      0.000000256                      0.000003120                      0.000001230                         53027.00                            56.45                         25012.58                             0.81
11/20/2013 10:39:30.340396	                      0.000000256                      0.000003120                      0.000001230                         54030.67                            62.74                         25012.58                             0.81
11/20/2013 10:39:31.340369	                      0.000000260                      0.000003120                      0.000001230                         54594.87                            66.28                         25012.58                             0.81
11/20/2013 10:39:32.340350	                      0.000000253                      0.000003120                      0.000001230                         55528.72                            64.69                         25012.58                             0.81
11/20/2013 10:39:33.341106	                      0.000000260                      0.000003120                      0.000001230                         56029.98                            69.58                         25012.58                             0.81
11/20/2013 10:39:34.340924	                      0.000000260                      0.000003098                      0.000001252                         56031.12                            69.09                         25012.58                             0.81
11/20/2013 10:39:35.340837	                      0.000000319                      0.000003758                      0.000001604                         56032.27                            68.90                         25012.58                             0.81
11/20/2013 10:39:36.340732	                      0.000000306                      0.000003714                      0.000001626                         56032.27                            70.61                         25012.58                             0.81
11/20/2013 10:39:37.341651	                      0.000000300                      0.000003582                      0.000001604                         56032.27                            69.88                         25012.58                             0.81
11/20/2013 10:39:38.342419	                      0.000000291                      0.000003538                      0.000001560                         56032.27                            67.68                         25012.58                             0.81
11/20/2013 10:39:39.342271	                      0.000000291                      0.000003494                      0.000001560                         56032.27                            68.54                         25012.58                             0.81
11/20/2013 10:39:40.342168	                      0.000000289                      0.000003494                      0.000001560                         56032.27                            68.17                         25012.58                             0.81
11/20/2013 10:39:41.341996	                      0.000000289                      0.000003472                      0.000001494                         56032.27                            66.77                         25012.58                             0.81
11/20/2013 10:39:42.341916	                      0.000000284                      0.000003450                      0.000001494                         56032.27                            67.99                         25012.58                             0.81
11/20/2013 10:39:43.342793	                      0.000000284                      0.000003450                      0.000001472                         56032.27                            66.10                         25012.58                             0.81
11/20/2013 10:39:44.342661	                      0.000000282                      0.000003384                      0.000001450                         56032.27                            67.01                         25012.58                             0.81
11/20/2013 10:39:45.342577	                      0.000000282                      0.000003384                      0.000001450                         56032.27                            67.32                         25012.58                             0.81
11/20/2013 10:39:46.343125	                      0.000000282                      0.000003362                      0.000001450                         56032.27                            67.13                         25012.58                             0.81
11/20/2013 10:39:47.343318	                      0.000000282                      0.000003384                      0.000001450                         56032.27                            65.24                         25012.58                             0.81
... 10427 more lines ...