Message ID: 456     Entry time: Wednesday, September 11, 2019, 15:16
Author: Marla Cervantes 
Target Material:  
Target Oven W.O#: 44062-4 
Source: LP-SIS 
Status: Done 
Subject: TaC coated target container calibration/conditioning  

 Wednesday, September 11, 2019, 15:12

A target container coated with TaC elog 455 was loaded with an empty graphite cartridge and set it in the EVAP2

The system is currently pumping down.

The target container was wrapped with 3 new Ta heat shields.

The target container will be heated to 2000 C and the temperature will be held for 4 hours.


 The autoramp was set to:

TGHT 1 A/min, 450 A

TBHT 0.5 A/min, to 230 A


At 500 A on the TGHT, 20 additional amps were programmed to increased at 1 A/min, however, the autoramp increased the current to 520 A without stepping (see figure A). 


The final conditions were:

TGHT: 9.95 V, 565.59 A

TBHT: 4.18 V, 233.43 A

IGP1: 2.2 e-6 Torr


Friday, September 13, 2019, 11:16

After 4 hours the EVAP2 was cooled down

at a rate of TGHT 1 A/min and TBHT 0.5 A/min

The EVAP2 was vented with air.


 

Attachment 1: A.JPG  98 kB  Uploaded Friday, September 13, 2019, 12:25  | Show | Hide all
Attachment 2: B.JPG  95 kB  Uploaded Friday, September 13, 2019, 12:26  | Show | Hide all