Laura Lambert wrote: |
Laura Lambert wrote: |
Laura Lambert wrote: |
Thursday, December 07, 2017, 09:57
Each endcap is wrapped in a separate piece of graphite foil and placed in a target boat.
There are also 4 coated springs, 2 on each side of the stack of endcaps.
The boat is wrapped with 3-layers of Ta heatshields.
Chamber pumped down overnight.
OMEGA slope = 0.98
HS water valves opened, no leaks detected.
IGP1 = 5.9e-7 Torr
IGP2 = 4.0e-7 Torr
Tapping up manually.
Final desired temperature = 1800C
|
Thursday, December 07, 2017, 11:51
TGHT tripped off.
Turns out the vacuum limits have not been fully updated when they should have been and the old ones still apply.
IGP1 reached 8e-5 Torr which tripped off the system.
TGHT = 80A, 1.05V when the system tripped.
|
Friday, December 08, 2017, 15:29
Interlocks have been updated.
TGHT should now run up to 500mTorr before tripping off.
|
Friday, December 08, 2017, 15:41
IGP1 = 2.8e-7 Torr
IGP2 = 1.9e-7 Torr
Tapped up manually to 50A
Initiated auto-ramp at a rate of 0.2A/min up to 300A.
Vacuum limit stop at 3.0e-5 Torr, backing off at 4.0e-5 Torr.
Monday, December 11, 2017, 09:37
IGP1 = 2.4e-7
TGHT = 300A, 4.80V
OMEGA = 1175C
Tapping up manually...
Monday, December 11, 2017, 14:19
Tapped up manually to 1800C
IGP1 = 2.1e-6 Torr
IGP2 = 2.1e-6 Torr
TGHT = 545A, 9.90V
OMEGA = 1810C
Will remain at max setpoints overnight and cool down in the a.m.
Tuesday, December 12, 2017, 09:01
TGHT = 545A, 9.96V
OMEGA = 1833C
Remained at max setpoints for ~18hrs
IGP1 = 4.2e-7 Torr
IGP2 = 3.9e-7 Torr
Initiating auto-cool at 2.0A/min
Wednesday, December 13, 2017, 11:05
Vented in air and removed from evap2 this morning.
|