OLIS Logbook General
  General, Page 16 of 26  Not logged in ELOG logo
ID Date Author Type Categorydown Subject
  270   Fri May 14 08:50:56 2010 Damien Gallop  22Ne2+ at 22.44keV for SEBT1

MWS continues to provide 22Ne2+ for pilot beam use. Beam quietly sitting on FC6. Stripchart attached. Also, Supernanogan plasma is on at low RF power for conditioning.

Attachment 1: Picture_5.png
Picture_5.png
  271   Mon May 17 09:09:50 2010 Damien Gallop  22Ne2+ at 22.44keV for SEBT1

Last day of this pilot beam to TUDA-II. Expect a beam switch tomorrow for sake of TIGRESS.

Attached is Stripchart of weekend beam operation. Pretty nice indeed. Top trace (red) is Q3 skimmer, representative of total beam out of the source  Blue trace is 22Ne2+ on FC6 with magnet field stabilization program active. Brown trace is NA-LCW temperature as measured at exit of Supernanogan plasma chamber. Note an inverse trend against ambient temperature. Something to investigate at some point.

Attachment 1: Picture_6.png
Picture_6.png
  272   Tue May 18 09:03:32 2010 Damien Gallop  22Ne2+ at 22.44keV for SEBT1 EOB

EOB for SEBT1 run at 08:30, taking 11Li from ITE and 22Ne2+ from OLIS when needed. Attempted emittance scans last evening, but no data showed, and no more time to investigate. So be it. Another time.

Attachment 1: Picture_6.png
Picture_6.png
Attachment 2: Picture_7.png
Picture_7.png
Attachment 3: Picture_8.png
Picture_8.png
  273   Tue May 18 09:15:16 2010 Damien Gallop  20Ne1+ at 40.8keV for TIGRESS setup

Prepare to switch beam from 22Ne2+ to 20Ne1+. Skimmers from 22Ne2+ run shown below. Table of tune options below. We will be using the Rick Tune.

Device Present Tune Scaled Scaled   Rick Tune Scaled   Web Tune Final
  22Ne2+ 20Ne2+ 20Ne1+   25Mg1+ 20Ne1+   20Ne1+ 20Ne1+
IOS:BIAS:VOL.VAL 22440 20400 40800   51000 40800   40800 40800
IOS:EE:VOL.VAL 1072 975 1949     3064   3064 1419
IOS:EESS:VOL.VAL 1810 1645 3290     0   0 0
IOS:XCB1:VOL.VAL 225 205 409     0   0 500
IOS:YCB1:VOL.VAL 268 244 487     0   0 140
IOS:CCB1:VOL.VAL 500 455 909     0   0 500
IOS:B1A:POS:VOL.VAL 3301 3001 6001     5693   5693 5933
IOS:B1A:NEG:VOL.VAL 3845 3496 6991     5693   5693 5933
IOS:XCB1AW:VOL.VAL 1658 1507 3014     4362   4362 3660
IOS:XCB1AE:VOL.VAL 2242 2038 4076     4362   4362 3660
IOS:YCB1A:VOL.VAL 412 375 749     0   0 340
IOS:Q1:POS:VOL.VAL 1215 1104 2208   -2425 -1940   0 1113
IOS:Q1:NEG:VOL.VAL 1215 1104 2208   -2425 -1940   0 1113
IOS:Q2:POS:VOL.VAL 148 134 269   -1115 -892   1428 892
IOS:Q2:NEG:VOL.VAL 148 134 269   -1115 -892   1428 892
IOS:Q3:POS:VOL.VAL 786 715 1429   0 0   2570 0
IOS:Q3:NEG:VOL.VAL 786 715 1429   0 0   2570 0
IOS:YCB3:VOL.VAL 397 361 721     0   0 500
IOS:MB:CUR.VAL 148 135 270     280   280 277
IOS:YCB4:VOL.VAL 316 287 575     0   0 470
IOS:CCB4:VOL.VAL 300 273 545     0   0 300
IOS:Q4:POS:VOL.VAL 1384 1258 2516   1333 1067   3392 1067
IOS:Q4:NEG:VOL.VAL 1384 1258 2516   1333 1067   3392 1067
IOS:Q5:POS:VOL.VAL 1254 1140 2280   0 0   2232 0
IOS:Q5:NEG:VOL.VAL 1254 1140 2280   0 0   2232 0
IOS:Q6:POS:VOL.VAL 0 0 0   3225 2580   0 2580
IOS:Q6:NEG:VOL.VAL 0 0 0   3225 2580   0 2580
IOS:YCB6:VOL.VAL 290 264 527     0   0 300

Saved tunes: 100518_1113.snapiosfc6 (strict) & 100518_1128.snapiosfc6 (tweaked). Emittance scan attached.

At noon, beam handed over to Sophia. She has questions about the defeats associated with the Q3 failure last year. Checking... done.

Summary of Q1/2/3 wiring status right now:

  • Defeat remains on Q1.
  • Q1+/- cables swapped at power supplies.
  • Q2 output is normal (+ drives L/R and - drives T/B) and is opposite to Q1 and Q3.
  • Q3 output is normal (+ drives T/B and - drives L/R).
  • All wiring at DBs is normal.

Sophia is continuing her tuning using IOS tune as provided.

Later... Sophia reports that this beam transports nicely. That said, the IOS tune is a little different now. Compare attached below.

Attachment 1: Picture_7.png
Picture_7.png
Attachment 2: grab.D19916.png
grab.D19916.png
Attachment 3: grab.l19920.png
grab.l19920.png
Attachment 4: grab.w19925.png
grab.w19925.png
Attachment 5: grab.z19931.png
grab.z19931.png
Attachment 6: grab.r19947.png
grab.r19947.png
Attachment 7: Picture_13.png
Picture_13.png
Attachment 8: Picture_14.png
Picture_14.png
  274   Wed May 19 09:55:51 2010 Damien Gallop  20Ne1+ at 40.8keV for TIGRESS

ISAC-I tuning done yesterday. TBC today into ISAC-II. Beam very nice overnight.

Attachment 1: Picture_6.png
Picture_6.png
  275   Mon May 24 22:42:51 2010 Keerthi Jayamanna  33S development

Tried to see the background before 33S installation. Found few problems: MCIS:YCB1 does not hold more than 58V.

Very difficult to transport the beam without YCB1. Need to find a solution. Open and fix it is not an option even though we have a ready spare. It will take two weeks to get the vacuum back.

------------------------------

Outgoing TIGRESS experiment was the occasion for another ISAC milestone, 8MeV/u through the SCC for the first time just around midnight last Friday, May 21st. This also represented the first 8MeV/u beam to an experiment. And we can proudly say that OLIS beam was used. TIGRESS run ended at 08:30 as scheduled. Attached is a trar graph of beam on FC6, snapped periodically throughout the run. As before, you can see a small daily variation likely related to temperature of NA-LCW. There were reports of HPL tests this past Thursday and Friday. Evidently no effect over here.

 

Attachment 1: Picture_1.png
Picture_1.png
  276   Tue May 25 00:19:18 2010 Keerthi Jayamanna  33S

 

Finally got something going. YCB1 short to the ground. YCB common plate connected to YCB1 power supply. We were lucky. Polarity was right. Ran the source with no 33S.
Saved file as “backgroundNo33S.dat” No optics set yet. SO2 connected not open yet. 33S 17% is in the oven, but not heated yet.
  277   Tue May 25 11:28:00 2010 Keerthi Jayamanna  33D dev

after 24 hours background spectrum saved as "BackgroungNo33S002.dat"

RF Amplitude :1.553 , RFoutput power= 0, Output power (dB)=0  

  278   Tue May 25 11:58:02 2010 Keerthi Jayamanna  33S dev

after 24 hours background spectrum saved as "BackgroungNo33S003.dat"

RF Amplitude :7.767 , RFoutput power= 5W, Output power (dB)=36.4

  279   Tue May 25 22:02:41 2010 Keerthi Jayamanna  33S spectrum

33S spectrum is saved as 33S_spectrum.dat 

  280   Wed May 26 18:16:10 2010 Damien Gallop  33S Development

Some 3enA of beam on FC6 presently. Let it sit.

Attachment 1: grab.Cm6004.png
grab.Cm6004.png
  281   Thu May 27 09:11:10 2010 Damien Gallop  33S Development and Handover

33S on FC6 overnight. Preparing to send beam to DRAGON now, and confirm isotope.

Source gas pressure gauge on canister = -20"Hg. MCIS pressure = 7.8e-8 Torr.

Attachment 1: Picture_2.png
Picture_2.png
  282   Fri May 28 11:02:59 2010 Damien Gallop  33S to DRAGON

Beam tuning in progress. Source gas canister at -20"Hg. MCIS pressure at 7.62e-8 Torr.

Attachment 1: Picture_1.png
Picture_1.png
  283   Mon May 31 09:18:39 2010 Damien Gallop  33S to DRAGON

Delivery of 33S6+ to DRAGON continues. A couple of RF tweaks done over the weekend. Source gas canister pressure at -20"Hg. MCIS pressure at 7.3e-8 Torr. At this rate, we could deliver 33S until approximately forever.

Attachment 1: Picture_2.png
Picture_2.png
  284   Tue Jun 1 09:12:39 2010 Damien Gallop  33S to DRAGON

33S6+ delivery to DRAGON continues. Looks like there was an RF tweak overnight. Source gas canister still at ~-20"Hg. MCIS vacuum at 7.4e-8 Torr.

Attachment 1: Picture_2.png
Picture_2.png
  285   Wed Jun 2 09:19:04 2010 Damien Gallop  33S to DRAGON

33S6+ delivery to DRAGON continues. It is Maintenance Day for everybody else today. Source gas canister pressure = -20"Hg. MCIS pressure = 7.7e-8 Torr.

  286   Thu Jun 3 09:29:11 2010 Damien Gallop  33S to DRAGON

33S6+ delivery to DRAGON continues until Saturday morning. All the signs are that beam is excellent. Source gas canister is still at ~-20"Hg, and MCIS vacuum is at 7.0e-8 Torr.

Attachment 1: Picture_1.png
Picture_1.png
  287   Fri Jun 4 09:30:48 2010 Damien Gallop  33S to DRAGON

Delivery of 33S6+ continues to DRAGON for the final day. Some online tuning of the line overnight. Source gas canister pressure remains at ~-20"Hg, and MCIS pressure is at 7.1e-8 Torr.

Attachment 1: Picture_2.png
Picture_2.png
  288   Mon Jun 7 10:36:14 2010 Damien Gallop  15N setup over the weekend

Keerthi set up the Supernanogan for 15N operation and destined for TIGRESS, using two source gas options: a miniature bottle of enriched 15N2 and a canister with a charge of research-grade 14N2. The canister is similar to the one we used for the 33S run to DRAGON, which finished Saturday morning. Canister is at -22"Hg and MCIS at 7.27e-8 Torr. A couple of beam stability tests were conducted at 15N4+ and 15N6+. Bob can choose which one he prefers to use. 15N4+ is selected and tuned through IOS up to the RFQ presently (~10:30 am).

Later... Bob reported at the Beam Delivery meeting that the 33S6+ beam was a very nice beam. We appreciate hearing it.

Attachment 1: Picture_2.png
Picture_2.png
  289   Tue Jun 8 09:27:54 2010 Damien Gallop  15N4+ to TIGRESS

User tuning began Sunday night. Beam physics expert chose to retire the IOS:Q1 defeat, as he did not know until yesterday that it existed. He discovered it after attempting to retune yesterday around noon from scratch and making a mess, notwithstanding Keerthi's clear instruction to leave it alone. He also does not know that Rick developed a tune for this, which provides cleaner transmission than the "book" tune. Pity. Presently there are some 75enA of beam at FC6. That should be enough. Note that there were some 90enA with the "defeat" tune in place, prior to changing the IOS tune. Canister is at ~-22"Hg, and MCIS pressure at ~7.0e-8 Torr at ~09:00.. ICR is tuning through presently.

Attachment 1: Picture_4.png
Picture_4.png
Attachment 2: Picture_5.png
Picture_5.png