  4   Fri Apr 16 10:15:26 2004 Doug PreddyStart UpChanges to M9sol Vacuum PCYesterday and today there have been a number
of changes made to the M9 
Vacuum control computer. The computer now
  
  5   Mon Apr 19 15:06:49 2004 Doug PreddyStart UpStarted turbinesPurifier run over the weekend. Both LN2 dewars
empty, but purifier cold. 
Disabled CV211 and flowed all charge gas
 april191500.BMP 
  6   Tue Apr 20 07:06:41 2004 Doug PreddyStart Upmorning day 2Temps in solenoid are equalling out. Increasing
PIC229 to 11 bar. 
 april200705.BMP 
  7   Tue Apr 20 07:33:03 2004 Doug PreddyStart UpPIC229 at 11barPIC229 at 11 bar. Changed strip tool to 15
sec increments.
  
  8   Wed Apr 21 06:39:17 2004 Doug PreddyStart Upover nightChanged to parallel cold mode at 9:30.
At about 4:30 the city water chiller was
turned on. The air compressor had 
  
  9   Wed Apr 21 06:48:31 2004 Doug PreddyStart UpJT bypass closedClosed cv155   
  10   Wed Apr 21 08:16:18 2004 Doug PreddyStart Upswitch to series modeSwitched to series mode.   
  11   Wed Apr 21 09:57:59 2004 Willy AnderssonStart UpSystem statusBuffer tank pressure is down to about 6 psig.
Too low to keep up
with liquefaction requirements to build LHe
  
  42   Thu Aug 19 10:28:32 2004 Doug PreddyStart Upafter M9bv5 FIXOver the last couple of days we replaced the
gate on the M9BV5 valve. The 
old one had worn the teflon sliders on the
  
  52   Mon Oct 4 09:16:29 2004 Doug PreddyStart UpStarted turbinesStarted flow to the turbimes with the solenoid
in parallel warm return.
  
  53   Tue Oct 5 07:18:16 2004 Doug PreddyStart Upadjust compressor discharge pressureadjusted compressor discharge pressure to
11 Bar
  
  54   Wed Oct 6 07:57:01 2004 Doug PreddyStart Upparallel cold returnSwitched to parallel cold return at 4:00 yesterday
afternoon.
6:30 this morning started closing cv155 as
  
  55   Wed Oct 6 09:25:54 2004 Doug PreddyStart Upswitch to series modeswitched to series mode   
  56   Thu Oct 7 13:33:20 2004 Doug PreddyStart Upturned off purifierturned off purifier   
  58   Sun Oct 10 17:36:35 2004 ExperimentorStart Uppower leadsAt 5:30 pm sundat, power lead temp is about
10.6K.  script tool still 
notworking
  
  59   Mon Oct 11 03:56:14 2004 ExperimentorStart Upcv142 valvenoticed that the lead temp rose to 24K over
the past 4 hours. had control 
room personnle cycle the cv142 valve. temperature
  
  60   Mon Oct 11 11:49:34 2004 ExperimentorStart Upcv142 valvenoticed temp rise, which occured only about
5 hours after the 
last "reset" .... the time between resets
  
  61   Mon Oct 11 22:32:11 2004 ExperimentorStart Upcv142 valvetemp of leads begins its exponential rise
again (caught it at 12K), approx 
11 hours after last reset.  this time asked
  
  62   Tue Oct 12 02:05:31 2004 ExperimentorStart Upcv142 valveafter only two hours since last lead cycling,
they warm up quickly. 
ops cycles the valve, but more quickly this
  
  63   Tue Oct 12 13:00:08 2004 Curtis BallardStart Upcv142 valveCurtis closes and opens slowly at about 11:00
am this morning. Wants to see 
if that works better than the sudden on/off
  