  206   Wed Jul 4 11:13:12 2007 Doug PreddyQuench Recoverytest
This is a test to make sure the e-mail works.
  95   Wed Jun 1 23:48:02 2005 Doug PreddyQuench Recoverythe first quench
I recieved a call from ops at 22:00 tonight that the solenoid had alarmed. 
I came in and found a cold box fault, with the compressor tripped off. I 
was unable to restart the compressor. I cleared the cryostat quench and 
compressor trips. The system looked as though it was ready to start. I 
pushed the compressor "ON" button, but as soon as the main contactor 
closed, the compressor tripped again. After a few trys and a bit of 
searching, I tried resetting the compressor over-pressure switch. That 
allowed me to restart the compressor. Once the high pressure comes up I'll 
start flow to the turbines.
  84   Wed Nov 24 14:58:42 2004 Doug PreddyQuench Recoverytoday
Came in this morning to find the helium return line to the buffer tank 
frosted up. MV51 has been stuck open the whole time. That exlains the 
cooldown rate and other problem. Cycled MV51 a few times and it now seems 
closed. Switched back to parallel cold return this moring at 8:00. Rates 
are much better. The E-log was also down this morning.
  56   Thu Oct 7 13:33:20 2004 Doug PreddyStart Upturned off purifier
turned off purifier
  196   Wed May 16 10:02:35 2007 Doug PreddyMaintenancewarming up power leads
Power lead temps have been slowly increasing over the last wekk. I ramped 
the solenoid down and am warming the power leads to >80k.
  110   Mon Oct 31 07:48:57 2005 Willy AnderssonQuench Recoverywhat a fine entry
Keep IT safe boys