M9sol Logbooks Solenoid Daily log
  Solenoid logbook, Page 7 of 15  ELOG logo
ID Date Author Operational Mode Subjectdown Text Attachments
  64   Wed Oct 13 08:17:00 2004 ExperimentorStart Upcycling 142 when temp hit 10.68I tried cycling the valve twice -- at about
11:30 and again around 1:00,
trying both the buttons and the dial -- but
  
  67   Sun Oct 17 04:26:29 2004 OtherStart Upcycling 142 when temp hit 10.68> I tried cycling the valve twice -- at about
11:30 and again around 1:00,
> trying both the buttons and the dial --
  
  212   Tue Jul 31 07:53:57 2007 Kevin TrithardtNormal Runningcycled power leadsgot a page for the power lead temp
came down and cycled the power leads.. did
not seem to make a difference..
  
  59   Mon Oct 11 03:56:14 2004 ExperimentorStart Upcv142 valvenoticed that the lead temp rose to 24K over
the past 4 hours. had control 
room personnle cycle the cv142 valve. temperature
  
  60   Mon Oct 11 11:49:34 2004 ExperimentorStart Upcv142 valvenoticed temp rise, which occured only about
5 hours after the 
last "reset" .... the time between resets
  
  61   Mon Oct 11 22:32:11 2004 ExperimentorStart Upcv142 valvetemp of leads begins its exponential rise
again (caught it at 12K), approx 
11 hours after last reset.  this time asked
  
  62   Tue Oct 12 02:05:31 2004 ExperimentorStart Upcv142 valveafter only two hours since last lead cycling,
they warm up quickly. 
ops cycles the valve, but more quickly this
  
  63   Tue Oct 12 13:00:08 2004 Curtis BallardStart Upcv142 valveCurtis closes and opens slowly at about 11:00
am this morning. Wants to see 
if that works better than the sudden on/off
  
  170   Sun Oct 15 07:09:41 2006 OtherNormal Runningcryopump bcp2 tripDonald Arseneau reporting

Not really solenoid realated, but FYI...
  
  98   Thu Jun 2 09:16:40 2005 Doug PreddyQuench Recoverycontinuing to coolSee screen grab  02-06-05-7.32.49 AM.jpg 
  50   Mon Sep 13 13:56:06 2004 Doug PreddyShut Downcompressor offcompressor off   
  141   Thu Jul 20 13:27:21 2006 Willy AnderssonQuench Recoverycold return mode continues...cr_c1 and cr_12 are both at about 45K and
heading slowly down, so the 
gamble of going to // cold return worked.
  
  142   Thu Jul 20 13:34:26 2006 Willy AnderssonQuench Recoverycold return mode continues...Even though cr_c1 and cr_c12 have begun to
tick down, cr_c7 and cr_c6 are 
actually ticking up. This is probably a wave
  
  218   Tue Sep 18 18:48:11 2007 Curtis BallardMaintenancecold box leak checkThe leak check of the cold box diff pump fittings.
 Found a leak at the 
base of the diff pump on the tee that is
  
  252   Fri Jul 25 21:48:15 2008 Doug PreddyQuench Recoveryanother quenchSystem was stable and solenoid ramped up at
4:00 today. At about 7:30 it tripped off
again. I came in and restarted the compressor
  
  42   Thu Aug 19 10:28:32 2004 Doug PreddyStart Upafter M9bv5 FIXOver the last couple of days we replaced the
gate on the M9BV5 valve. The 
old one had worn the teflon sliders on the
  
  53   Tue Oct 5 07:18:16 2004 Doug PreddyStart Upadjust compressor discharge pressureadjusted compressor discharge pressure to
11 Bar
  
  88   Wed Dec 1 12:55:56 2004 Doug PreddyNormal Runningadded one bottle helium gasbuffer tank pressure is up to 6.5 psi.   
  201   Tue Jun 12 12:48:25 2007 Doug PreddyMaintenanceadded one bottle UHP helium to system    
  12   Thu Apr 22 10:39:16 2004 Doug PreddyNormal RunningYesterdayAdded two bottles of UHP helium to buffer
tank.
Had no liquifation rate, so optimized Ti121
  