M9sol Logbooks Solenoid Daily log
  Solenoid logbook, Page 7 of 15  ELOG logo
ID Datedown Author Operational Mode Subject Text Attachments
  166   Thu Oct 5 14:53:54 2006 Doug PreddyStart UpSwitched to series mode    
  165   Thu Oct 5 11:51:53 2006 Doug PreddyStart UpAdde 2 bottles UHP HeliumValved off purifiers, but left LN2 on.   
  164   Thu Oct 5 10:26:26 2006 Doug PreddyStart Upswitched to parallel cold return    
  163   Wed Oct 4 18:10:18 2006 Willy AnderssonStart UpQuick check on systemSystem in good // warm return mode, temps
coming down nicely. BCP2 cryopump 
does not look too healthy, temperature is
  
  162   Wed Oct 4 07:29:04 2006 Doug PreddyStart UpStarting cool downStarted compressor yesterday with flow through
the purifier overnight.
Starting tubines this morning in parallel
  
  161   Thu Aug 17 19:45:27 2006 Willy AnderssonQuench Recoveryswitch to series coolingTemps of flow returning from solenoid flat
lined at about 10K.
Gradually closed CV155.
  
  160   Thu Aug 17 08:04:44 2006 Doug PreddyQuench RecoverySolenoid tripped at about 3:00 this morningrestarting system.
Compressor would not start. Idid not see
anything that would not let it 
  
  159   Tue Aug 15 15:39:27 2006 Willy AnderssonQuench Recoveryswitched to parallel cold returnReturn flow temps from solenoid about 40K,
TIC121 at about 40K so switching 
to //cold return.
  
  158   Tue Aug 15 08:45:56 2006 Doug PreddyQuench Recoveryrestarted the compressorSolenoid tripped off Sunday. Restarted compressor
and cool down of cold box.
  
  157   Thu Aug 10 20:26:09 2006 Willy AnderssonQuench Recoverysolenoid energizedPowerleads seem stable.   
  156   Thu Aug 10 16:33:11 2006 Willy AnderssonQuench RecoveryFlow to solenoid tripped on powerlead temperature.Power lead temps were drifting up, tried to
tweak TIC121 and HIC141 to 
unstall powerlead flow. Got quench. Restarted
  
  155   Thu Aug 10 08:33:44 2006 Doug PreddyNormal RunningRunning low on heliumTemperture okay, but there are fluctuation
in the power leads. PIC229 is 
trying to charge. The buffer tank pressure
  
  154   Wed Aug 9 07:50:49 2006 Doug PreddyQuench RecoverySwitched to series mode.Cool down continues.   
  153   Wed Aug 9 06:59:43 2006 Willy AnderssonQuench Recoverymore // cold return stuffAbout 21:00 last night the solenoid temps
had not changed much, pretty well 
flat lined. I guess my reducing CV155 stalled
  
  152   Wed Aug 9 06:44:58 2006 Willy AnderssonQuench Recovery// cold mode cooling continues.After restarting flow to the solenoid yesterday
in // cold return mode 
TIC121 just refused to go lower than about
  
  151   Tue Aug 8 07:27:49 2006 Doug PreddyQuench RecoverySolenoid Quench at about 6:00 this morningNo indication of problem. Turbines still running,
but no flow to solenoid. 
Started flow. Solenoid ~20K.
  
  150   Sat Jul 29 09:38:06 2006 Doug PreddyQuench Recoverypower leads cooledopened power lead flow about 40 minutes ago.
~12K will let them stablize. 
LHE level >20%
  
  149   Sat Jul 29 05:30:28 2006 Doug PreddyQuench Recoveryswitched to series modeTemps bottomed out at ~16K. Switching to series
mode.
  
  148   Fri Jul 28 20:39:40 2006 Doug PreddyQuench RecoveryRestarting refrigeratorSolenoid quenched at about 6:00 tonight.
I restarted the compressor, then the cold
box and turbines. When cold box 
  
  147   Sat Jul 22 06:34:14 2006 ExperimentorQuench Recoverylooking goodDonald Arseneau reporting.

It is looking pretty good, so I came back
  