Entry  Wed Jul 4 11:13:12 2007, Doug Preddy, Quench Recovery, test 
This is a test to make sure the e-mail works.
Entry  Sun Jul 1 16:09:37 2007, Doug Preddy, Normal Running, M9B:DP2 tripped off  
Diffusion pump tripped off and vacuum (and power lead temps) going up. I 
had an experimentor cycle HIC 142 at 4:15 this morning. When I got in 
today I restarted the pump and it is back to 6.0 x 10-7. Power lead temps 
also are going down.
Entry  Tue Jun 19 21:15:41 2007, Doug Preddy, Quench Recovery, Today's quench 
power lead temps caused a quench at 8:30 this morning. Recovery of system 
and flow to solenoid was started by 11:00. Solenoid was switched to series 
cold at 3:00. Solenoid and power leads were cold at ~6:30. power supply 
was to be energized by experimenters.
Entry  Sat Jun 16 15:05:03 2007, Doug Preddy, Normal Running, Solenoid power lead temp problem 
Power lead temps rippling and increasing. I asked an operator to cycle 
HIC142. This seems to have fixed the problem (for now).
Entry  Thu Jun 14 00:10:10 2007, Doug Preddy, Quench Recovery, recovering from earlier quench 
Solenoid quenched on power lead temperature last night at about 6:00. I 
restarted flow to the solenoid in parallel cold return. Switched to series 
mode and am building liquid. 
Entry  Tue Jun 12 12:48:25 2007, Doug Preddy, Maintenance, added one bottle UHP helium to system 
 
Entry  Thu May 31 11:51:49 2007, Doug Preddy, Quench Recovery, Switched to series mode 
 
Entry  Thu May 31 07:11:05 2007, Doug Preddy, Quench Recovery, switched to parallel cold return 
switched to parallel cold return earlier this morning. Vacuum is 5 x 10-5.
Entry  Wed May 30 11:27:32 2007, Doug Preddy, Quench Recovery, Problems recovering 
Various start up problems :
MV51 stuck open. Finally got it closed by hitting the valve with a hammer.
Vacuum is bad in cold box as MV902 tripped closed.
Vacuum in solenoid has gone bad. Suspect boil off from inside solenoid. So 
far it is not recovering.
Previously T1 would not spin up - frozen.
Entry  Wed May 30 08:38:38 2007, Doug Preddy, Quench Recovery, Solenoid quenched this morning 
Solenoid quenched this morning on power lead temp (I believe).
I restarted it at ~7:00, but the capacity control for the compressor would 
not respond. I shut the system back down and checked as many relays as I 
could. I then restarted the system and it is now cooling. I will go 
through the wiring diagrams and see if I can locate the bad relay.
Entry  Wed May 16 10:02:35 2007, Doug Preddy, Maintenance, warming up power leads 
Power lead temps have been slowly increasing over the last wekk. I ramped 
the solenoid down and am warming the power leads to >80k.
Entry  Wed May 2 16:28:16 2007, Doug Preddy, Normal Running, Energizing solenoid to 500 amps 
Power supply initially would not ramp up. I exercised the rate control 
selector switch and found that the power supply would ramp up at only the 
30 minute setting. It is ramping up at a rate of about 60 minutes.
Entry  Mon Apr 30 09:30:26 2007, Doug Preddy, Normal Running, problems over weekend 
The PLC Limit switch power supply was tripped off on Saturday. I came in 
and reset it and restarted all the vacuum systems that had tripped off. 
The power leads were found to be warming up. This morning they were at ~ 
45K. I stopped flow to the power leads and will allow them to warm up 
above 80K and then will recool them.
Entry  Fri Apr 27 10:32:07 2007, Doug Preddy, Quench Recovery, Reset quench relay logic box 
The quench relay logic controller was not functioning correctly. It showed 
a "Trip" light and no other lights. I cycled the power to the controller 
and it now seems to be working properly. This did cause a quench and I am 
now recovering the system.
Entry  Fri Apr 27 10:16:44 2007, Kevin Trithardt, Quench Recovery, Test 
Don't Panic -- There was no quench. I am testing the e-log.
I have added Kevin's name and e-mail to the M9 e-log. He will be e-mailed 
when a quench recovery log is submitted. His name is also included in the 
Author's list. 
Thanks, Doug
Entry  Fri Apr 27 10:10:07 2007, Doug Preddy, Normal Running, Solenoid Cold 
Solenoid cold, power leads cold. Alarm handler started.
Entry  Tue Apr 24 10:21:49 2007, Doug Preddy, Start Up, Started flow to turbines 
Started flow to turbines at about 10:00 in parallel warm mode.
Entry  Mon Apr 23 16:31:02 2007, Doug Preddy, Start Up, Started compressor 
Startewd compressor and purifer today.
Entry  Fri Apr 20 12:15:12 2007, Doug Preddy, Start Up, remote PLC switch 
I could not control the vacuum devices using EPICS. The buttons could be 
pushed, but nothing would respond. The PLC was in local mode. Thumbwheel 
to 100, push On/Open button to put it into remote mode.
Entry  Thu Apr 19 15:18:25 2007, Doug Preddy, Start Up, preparing to start the refrigerator 
Starting the vacuum systems in the coldbox and solenoid.