M9sol Logbooks Solenoid Daily log
  Solenoid logbook, Page 3 of 15  ELOG logo
ID Date Authordown Operational Mode Subject Text Attachments
  65   Sat Oct 16 17:13:19 2004 Other ss   
  66   Sat Oct 16 17:18:50 2004 Other lame    
  67   Sun Oct 17 04:26:29 2004 OtherStart Upcycling 142 when temp hit 10.68> I tried cycling the valve twice -- at about
11:30 and again around 1:00,
> trying both the buttons and the dial --
  
  68   Sun Oct 17 05:03:13 2004 OtherStart UpCycle valve after oscillations begin(Sat night/Sun AM) Helium temperature started
oscillating, so
powered off magnet (precautionary) and cycled
  
  82   Wed Nov 24 07:28:30 2004 Other Mission Impossible   
  83   Wed Nov 24 07:28:38 2004 Other Mission Impossible   
  170   Sun Oct 15 07:09:41 2006 OtherNormal Runningcryopump bcp2 tripDonald Arseneau reporting

Not really solenoid realated, but FYI...
  
  192   Fri Apr 27 10:16:44 2007 Kevin TrithardtQuench RecoveryTestDon't Panic -- There was no quench. I am testing
the e-log.
I have added Kevin's name and e-mail to the
  
  208   Thu Jul 5 11:01:47 2007 Kevin TrithardtOtherCurrent lead warm upDue to curent lead temperature occilation
Doug recommended warming up the 
current leads.
  
  212   Tue Jul 31 07:53:57 2007 Kevin TrithardtNormal Runningcycled power leadsgot a page for the power lead temp
came down and cycled the power leads.. did
not seem to make a difference..
  
  241   Wed Nov 28 15:07:08 2007 Kevin TrithardtOtherpower lead regenerationPower leads cold again, power supply ramping
up.
  
  275   Mon Mar 29 09:25:52 2010 Kevin TrithardtStart UpStarted cool down this morning    
  282   Tue May 18 15:04:36 2010 Kevin TrithardtMaintenanceSolenoid ramping up    
  58   Sun Oct 10 17:36:35 2004 ExperimentorStart Uppower leadsAt 5:30 pm sundat, power lead temp is about
10.6K.  script tool still 
notworking
  
  59   Mon Oct 11 03:56:14 2004 ExperimentorStart Upcv142 valvenoticed that the lead temp rose to 24K over
the past 4 hours. had control 
room personnle cycle the cv142 valve. temperature
  
  60   Mon Oct 11 11:49:34 2004 ExperimentorStart Upcv142 valvenoticed temp rise, which occured only about
5 hours after the 
last "reset" .... the time between resets
  
  61   Mon Oct 11 22:32:11 2004 ExperimentorStart Upcv142 valvetemp of leads begins its exponential rise
again (caught it at 12K), approx 
11 hours after last reset.  this time asked
  
  62   Tue Oct 12 02:05:31 2004 ExperimentorStart Upcv142 valveafter only two hours since last lead cycling,
they warm up quickly. 
ops cycles the valve, but more quickly this
  
  64   Wed Oct 13 08:17:00 2004 ExperimentorStart Upcycling 142 when temp hit 10.68I tried cycling the valve twice -- at about
11:30 and again around 1:00,
trying both the buttons and the dial -- but
  
  99   Thu Jun 9 06:53:40 2005 ExperimentorOtherTrip at 6:45 Today. Solenoid quenched at 6:45.  All temperature
monitors were flat until trip.  
Hearing compressor stop was first warning.
  
ELOG V2.9.2-2455