M9sol Logbooks Solenoid Daily log
  Solenoid logbook, Page 3 of 15  ELOG logo
Entry  Wed Apr 16 16:51:16 2008, Doug Preddy, Start Up, switched to parallel cold return 
 
Entry  Wed Apr 16 15:15:44 2008, Doug Preddy, Start Up, refrigerator started  
Turned on compressor last Tuesday. Ran purifier until today. Final value 0.15ppm.
Started cool down yesterday. Switched into parallel cold return this
morning, but PIC 229 would not allow the compressor to charge. The compressor tripped due to bad solenoid vacuum (solenoid was warming up). The controller
now seems to be working okay. 
Restarted in parallel warm mode at ~12:00
Entry  Wed Apr 16 15:12:01 2008, Doug Preddy, Normal Running, graphs to load 
There are three strip tools saved:
coils
powerlead
startup
Entry  Tue Apr 1 15:17:06 2008, Doug Preddy, Start Up, Starting up vacuum systems 
Monday afternoon there was no control of the vacuum system. The controls were working, but no output to devices. The "output" breaker in the PLC box was
found to be tripped. Reseting it allowed control to the vacuum systems.
I started the regeneration of BCP2 this morning, but hte output breaker tripped
again. reset it and continued to pump.
Started to pump out the solenoid also. Removed jumper on the BDP2 temperature read-back, bypassed BV2, and BCG2.
Entry  Mon Dec 3 14:59:58 2007, Doug Preddy, Quench Recovery, Not really a quench recovery, But... 
Not really a quench recovery, but...I wanted to send this out via e-mail. 
The contactor for the compressor building fan failed recently. Randy is 
looking after it, but parts will have to be ordered. The big door should 
Entry  Wed Nov 28 15:07:08 2007, Kevin Trithardt, Other, power lead regeneration 
Power leads cold again, power supply ramping up.
Entry  Tue Nov 27 21:02:08 2007, Doug Preddy, Normal Running, power lead ripples 
power leads were oscillating tonight. John K. came down and cycled CV142. 
(Thanks John!) Temps settling out.
Entry  Mon Nov 26 09:45:43 2007, Doug Preddy, Normal Running, power lead problems 
Power lead temperatures were rippling yesterday afternoon. I asked one of 
the operators to cycle CV142.
The temperatures were rippling again this morning, so I cycled CV142.
Entry  Fri Nov 23 08:41:45 2007, Doug Preddy, Normal Running, screen grab of powerlead temps Capture.jpg
 
Entry  Fri Nov 23 08:35:37 2007, Doug Preddy, Normal Running, power lead problems 
About 10 hours ago the power lead temperatures started to fluctuate and 
climb.
I cycled CV142 and the Temps have dropped and stablized.
Entry  Mon Nov 19 04:30:47 2007, Doug Preddy, Quench Recovery, switched to series mode 
 
Entry  Mon Nov 19 02:58:44 2007, Doug Preddy, Quench Recovery, Solenoid quenched on power lead over temperature 
I received a page at 1:30 this morning. The power lead temperatures were 
climbing. Cycling CV142 did not fix the problem. The solenoid quenched at 
1:50.
Entry  Fri Nov 9 18:56:20 2007, Doug Preddy, Maintenance, System Ready to Run 
M9BMP2 was replaced. Vacuum good. Solenoid is ramped up. :)
Entry  Fri Nov 9 16:24:33 2007, Doug Preddy, Maintenance, 2APLC Died 
thumb wheel panel all lite up. 2APLC rebooted by Klara (Thanks). M9BMP2 not 
pumping properly. Edi is looking at it. 
Entry  Fri Nov 9 14:50:03 2007, Doug Preddy, Quench Recovery, Solenoid cold 
Solenoid cold. Added two bottles of helium. Allowing to stablize. Will 
ramp power supply in half an hour.
Entry  Fri Nov 9 07:37:44 2007, Doug Preddy, Quench Recovery, switched to series mode 
switched to series mode.
Entry  Thu Nov 8 09:28:53 2007, Doug Preddy, Quench Recovery, started compressor 
And started flow to coldbox,solenoid in parallel warm return.

Entry  Wed Oct 24 12:04:06 2007, Doug Preddy, Maintenance, power lead and cryopump regeneration 
Regenerating M9B:CP2 as temperature was up to almost 20k.
Shut off flow to power leads to allow them to warm up. The positive and 
negative temps were beginning to creep up.
Entry  Tue Oct 9 13:16:44 2007, Doug Preddy, Quench Recovery, switched to series mode 
 
Entry  Tue Oct 9 12:58:22 2007, Doug Preddy, Quench Recovery, Solenoid Quench 2 
Refrigeration sytem shut off at ~ 12:30 today. No apparent reason. Only 
light on was a cold box fault. 
Restarted in parallel cold return mode.