M9sol Logbooks Solenoid Daily log
  Solenoid logbook, Page 14 of 15  ELOG logo
New entries since:Wed Dec 31 16:00:00 1969
ID Date Author Operational Mode Subjectdown Text Attachments
  168   Thu Oct 12 19:54:10 2006 Willy AnderssonQuench RecoveryEnergized solenoid.Switched to series mode about 6 hours ago.
Cooling continued including 
power leads until LHe level at about 50%
  
  169   Fri Oct 13 00:13:57 2006 ExperimentorQuench RecoveryEnergized solenoid.Hi there,

This is Khashayar (experimenter). Lookin
  
  128   Wed Jun 14 16:27:01 2006 Willy AnderssonQuench RecoveryEnergized solenoidPowerleads appear stable.   
  130   Fri Jun 30 11:13:00 2006 Doug PreddyNormal RunningE-log is back up and runningI would have made an entry when it died, but-----   
  125   Tue Jun 13 15:52:46 2006 Willy AnderssonQuench RecoveryDoug has reduced and now closed CV155.Now in full // cold return mode.
Had just talked with Curtis, He said thta
he has already informed the 
  
  35   Wed Jul 28 21:26:19 2004 Doug PreddyQuench RecoveryDouble Power bumpTwo power bumps tonight. Everything tripped
of. Restarted cooling water 
pump, diff. pump under solenoid etc. Computer
  
  68   Sun Oct 17 05:03:13 2004 OtherStart UpCycle valve after oscillations begin(Sat night/Sun AM) Helium temperature started
oscillating, so
powered off magnet (precautionary) and cycled
  
  208   Thu Jul 5 11:01:47 2007 Kevin TrithardtOtherCurrent lead warm upDue to curent lead temperature occilation
Doug recommended warming up the 
current leads.
  
  143   Fri Jul 21 07:39:37 2006 Doug PreddyQuench RecoveryCool down continuesRaised PIC229 to 11Bar. Started to close HIC155.
Cooling rates improve.
  
  29   Fri Jun 25 12:36:04 2004 Willy AnderssonQuench RecoveryCompressor unloaded etc.At about 09:30 this morning I noticed the
sound coming from the compressor 
building seemed louder than usual. After
  
  40   Sat Aug 7 22:37:04 2004 Willy AnderssonQuench RecoveryCompressor tripped off.Got a call at about 10:00 this morning, the
whole system had tripped off. 
No power bumps or other disturbances. When
  
  124   Tue Jun 13 10:54:27 2006 Doug PreddyQuench RecoveryCompressor tripped offNo indicator lights etc as to why the compressor
tripped. The only  light 
was a cold box fault. I had to shut off the
  
  181   Wed Nov 15 11:09:28 2006 Doug PreddyQuench RecoveryCompressor tripped offCompressor tripped this morning at ~8:45.
I restarted the compressor, but 
turbine 1 would not spin up. I let it warm
  
  167   Thu Oct 12 11:53:04 2006 Willy AnderssonQuench RecoveryCompressor trip, 09:50 this morning.Main compressor tripped, refrigerator shutdown.
Checked with experimentors 
and main control room, no indications of
  
  106   Mon Jul 18 15:51:17 2005 Doug PreddyQuench RecoveryCompressor restarted, flow to solenoidRecovered the vacuum in the solenoid, then
started the compressor.
Started flow to solenoid in parallel warm
  
  213   Mon Aug 6 08:41:31 2007 Doug PreddyQuench RecoveryCold box trip at 6:00compressor off due to cold box trip at 6:00
this morning. Restarted system 
and started flow to solenoid in parallel
  
  4   Fri Apr 16 10:15:26 2004 Doug PreddyStart UpChanges to M9sol Vacuum PCYesterday and today there have been a number
of changes made to the M9 
Vacuum control computer. The computer now
  
  120   Fri May 26 08:45:38 2006 Doug PreddyStart UpChanged to series modeChanged to series mode.
Liquid level building.
  
  13   Fri Apr 23 09:19:30 2004 Doug PreddyNormal RunningCV211PIC 229 was down to 10.8 bar this morning.
PIC 229 was trying to open CV211 
to charge the system. I realized that in
  
  119   Fri May 26 07:29:54 2006 Doug PreddyStart UpCV155 closedCV155 closed.
Temps about 15K
  