|  Wed May  5 09:59:10 2004, Doug Preddy, Quench Recovery, started flow to power leads | 
| Started flow to power leads. TIC121 into auto mode
 | 
|  Wed May  5 07:55:37 2004, Doug Preddy, Quench Recovery, Solenoid quenched on power lead temp   | 
| Solenoid quenched on power lead temp. I ran down and managed to recover it with minimal temp rise in the solenoid. Coil temps are falling, liquid is
 building. No flow to the power leads yet.
 | 
|  Tue May  4 11:26:38 2004, Doug Preddy, Normal Running, power leads   | 
| Power cycled HIC 142 with solenoid power at 400 amps. Flows > 240mm and temps have dropped.
 | 
|  Tue May  4 11:20:57 2004, Doug Preddy, Normal Running, power leads going unstable | 
| power lead ripples increased over last couple of days. They are quickly dieing out now and the temps are startint to increase. PL flow is down to
 <180mm. I will power cycle HIC 142.
 | 
|  Fri Apr 30 01:53:34 2004, Doug Preddy, Quench Recovery, switched to parallel cold | 
|  | 
|  Thu Apr 29 23:43:15 2004, Doug Preddy, Quench Recovery, Starting flow to solenoid | 
| Solenoid quenched due to power leads at 9:00 tonight. Restarted flow in parallel warm mode.
 | 
|  Wed Apr 28 09:30:09 2004, Doug Preddy, Normal Running, power leads again | 
| Same story as yesterday. Ripples dieing out, flows down to <200 mm, temps increasing. Turned HIC 142 down to 0.4 then back up to 0.75. Temps dropped
 quickly. Then turned HIC 142 to 0 and back to 0.75. Temps dropping, flow >
 | 
|  Tue Apr 27 09:03:18 2004, Doug Preddy, Normal Running, Power leads | 
| Power leads are being monitored with the strip tool graphs. I saw the temps go from having a ripple to almost no ripple, but starting to increase in
 temperature. I also noticed that the flow had dropped to 200 mm. I power
 | 
|  Mon Apr 26 08:12:30 2004, Doug Preddy, Quench Recovery, Power leads flow | 
| Throttled back the LHe in until the power leads started to warm. My goal is to flow "warm" gas (hopefully 80K) through the leads to drive any frozen
 impurities out.
 | 
|  Mon Apr 26 07:48:46 2004, Doug Preddy, Quench Recovery, Series mode | 
| Switching over to series mode. | 
|  Sun Apr 25 21:41:00 2004, Willy Andersson, Quench Recovery, Trip on PL Temp. | 
| Got pager call at 16:20 this afternoon. Solenoid had tripped on PL temperature. Arrived at about 17:00. Did a print screen of the striptool
 graph and solenoid screen. See files Prtscrn1_april25_2004.doc and
 | 
|  Fri Apr 23 09:19:30 2004, Doug Preddy, Normal Running, CV211 | 
| PIC 229 was down to 10.8 bar this morning. PIC 229 was trying to open CV211 to charge the system. I realized that in restoring the control to CV211 I
 did not get the wire into the connector properly. Went back up to the
 | 
|  Thu Apr 22 10:39:16 2004, Doug Preddy, Normal Running, Yesterday | 
| Added two bottles of UHP helium to buffer tank. Had no liquifation rate, so optimized Ti121 and HIC141 . Finally put TIC121
 into manual mode at 0.65. Power lead fluctuation died out and liquid
 | 
|  Wed Apr 21 09:57:59 2004, Willy Andersson, Start Up, System status | 
| Buffer tank pressure is down to about 6 psig. Too low to keep up with liquefaction requirements to build LHe level. Doug Preddy
 requested that I help Curtis Ballard top up system with new
 | 
|  Wed Apr 21 08:16:18 2004, Doug Preddy, Start Up, switch to series mode | 
| Switched to series mode. | 
|  Wed Apr 21 06:48:31 2004, Doug Preddy, Start Up, JT bypass closed | 
| Closed cv155 | 
|  Wed Apr 21 06:39:17 2004, Doug Preddy, Start Up, over night | 
| Changed to parallel cold mode at 9:30. At about 4:30 the city water chiller was turned on. The air compressor had
 also tripped off. It was reset and restarted.
 | 
|  Tue Apr 20 07:33:03 2004, Doug Preddy, Start Up, PIC229 at 11bar | 
| PIC229 at 11 bar. Changed strip tool to 15 sec increments. | 
|  Tue Apr 20 07:06:41 2004, Doug Preddy, Start Up, morning day 2   | 
| Temps in solenoid are equalling out. Increasing PIC229 to 11 bar. | 
|  Mon Apr 19 15:06:49 2004, Doug Preddy, Start Up, Started turbines   | 
| Purifier run over the weekend. Both LN2 dewars empty, but purifier cold. Disabled CV211 and flowed all charge gas through purifier. Started turbine
 at 11:00am. set strip tool to monitor temps. Started M9BCP2. Pumped out
 |