86
Sun Nov 28 10:15:18 2004
Willy Andersson
Quench Recovery
Powerlead quench ?
Got page very early this morning. Couldn't
come in til 08:30. System was
in idle running mode and someone had reset
85
Thu Nov 25 07:21:37 2004
Doug Preddy
Quench Recovery
started power lead flow
Curtis switched to series mode last night
at 7:00.
Startedflow to power leads at 7:00 this morning.
84
Wed Nov 24 14:58:42 2004
Doug Preddy
Quench Recovery
today
Came in this morning to find the helium return
line to the buffer tank
frosted up. MV51 has been stuck open the
83
Wed Nov 24 07:28:38 2004
Other
Mission Impossible
82
Wed Nov 24 07:28:30 2004
Other
Mission Impossible
81
Tue Nov 23 15:10:06 2004
Doug Preddy
Quench Recovery
switched to parallel cold return mode
80
Tue Nov 23 11:35:44 2004
Doug Preddy
Quench Recovery
Another power bump
Compressor tripped off again due to a power
bump. Meson hall water off, so
vacuum pumps off again.
79
Tue Nov 23 09:30:15 2004
Doug Preddy
Quench Recovery
started flow to solenoid
Started flow to solenoid in parallel warm
return.
78
Tue Nov 23 07:47:00 2004
Doug Preddy
Quench Recovery
Solenoid quench Nov 21
Solenoid quenched Nov 21 due to alarge power
bump. Everything tripped off
including vacuum and water pumps. Cold box
77
Thu Nov 4 13:33:49 2004
Doug Preddy
Normal Running
Power supply problems
Last night the power supply was ramping up
at a very slow rate. This
morning it was energized,but with no current
76
Mon Nov 1 07:17:44 2004
Doug Preddy
Normal Running
power lead flow gone bad
power temps as follows:
LHE_IN 23
PL_POS 36
75
Fri Oct 29 14:25:23 2004
Doug Preddy
Normal Running
power lead flows dropping
Flow is down to <190 mm. temps are starting
to climb. Cycled HIC142
74
Tue Oct 26 09:45:24 2004
Doug Preddy
Normal Running
power leads warming
Flow to power leads dropped to <80 mm. Cycled
HIC142. Temps dropping, flows
>250.
73
Mon Oct 25 15:44:15 2004
Doug Preddy
Normal Running
Power lead temps starting to ripple
Power lead temps starting to ripple at a 1.5min
period. PL flows are
both>250 mm.
72
Fri Oct 22 09:15:06 2004
Doug Preddy
Start Up
started flow to power leads
Started flow to power leads HIC142 at 0.8.
71
Thu Oct 21 15:41:29 2004
Doug Preddy
Start Up
switched to series mode
70
Thu Oct 21 07:16:35 2004
Doug Preddy
Start Up
switched to parallel cold return
69
Wed Oct 20 09:38:17 2004
Doug Preddy
Start Up
restarting system
Solenoid tripped off Monday morning due to
high compressor pressure.
Regenerated purifier on Tuesday. Started
68
Sun Oct 17 05:03:13 2004
Other
Start Up
Cycle valve after oscillations begin
(Sat night/Sun AM) Helium temperature started
oscillating, so
powered off magnet (precautionary) and cycled
67
Sun Oct 17 04:26:29 2004
Other
Start Up
cycling 142 when temp hit 10.68
> I tried cycling the valve twice -- at about
11:30 and again around 1:00,
> trying both the buttons and the dial --