M9sol Logbooks Solenoid Daily log
  Solenoid logbook, Page 10 of 15  ELOG logo
Entry  Thu Oct 26 14:14:57 2006, Doug Preddy, Quench Recovery, power lead quench 
solenoid tripped on power lead temp. I was here and caught it. System 
cooling.
Entry  Tue Oct 31 03:52:29 2006, Doug Preddy, Quench Recovery, power lead quench at 2:30 this morning 
Compressor still on. Starting flow to solenoid
Entry  Tue Oct 31 05:36:14 2006, Doug Preddy, Quench Recovery, switch to series mode 
 
Entry  Tue Oct 31 08:37:34 2006, Doug Preddy, Quench Recovery, Energizing solenoid 
 
Entry  Fri Nov 3 12:09:51 2006, Doug Preddy, Normal Running, Fine tuning 
The power lead temperatures have been climbing at a rate of ~ 0.02K very 8 
hours. The negative lead started to show some rippling this morning. The 
positive flow had dropped from >250mm to 240mm. I cycled CV142 and the 
Entry  Tue Nov 7 10:35:32 2006, Doug Preddy, Normal Running, Maintenance day 
I ramped down the solenoid power suplly and shut off flow to the power 
leads. I will allow them to warm up and (hopefully) recover the flow rates. 
I am also regenerating the M9BCP@ cryopump.
Entry  Wed Nov 15 11:09:28 2006, Doug Preddy, Quench Recovery, Compressor tripped off 
Compressor tripped this morning at ~8:45. I restarted the compressor, but 
turbine 1 would not spin up. I let it warm for an hour and it started. I 
started flow to the solenoind in parallel cold return.
Entry  Wed Nov 15 13:56:00 2006, Doug Preddy, Quench Recovery, switch to series mode 
 
Entry  Tue Nov 28 10:32:04 2006, Doug Preddy, Quench Recovery, recovering after looong power outage 
The power has been out for about 24 hours. Recovering coldbox, solenoid and 
beamline vacuum.PLC is down for the vacuum system. Started compressor and 
am flowing warm helium through the soleniod to warm it up while the vacuum 
Entry  Fri Jan 26 15:10:37 2007, Doug Preddy, Maintenance, new control computer installed 
I have installed a new control computer at the M9 solenoid control panel
Entry  Mon Mar 5 14:59:41 2007, Doug Preddy, Maintenance, Password 
On the weekend someone used this log to send spam messages. I have 
installed a password for the write privilege. Please contact me at 6399 
for the password.
Entry  Thu Apr 19 15:18:25 2007, Doug Preddy, Start Up, preparing to start the refrigerator 
Starting the vacuum systems in the coldbox and solenoid.
Entry  Fri Apr 20 12:15:12 2007, Doug Preddy, Start Up, remote PLC switch 
I could not control the vacuum devices using EPICS. The buttons could be 
pushed, but nothing would respond. The PLC was in local mode. Thumbwheel 
to 100, push On/Open button to put it into remote mode.
Entry  Mon Apr 23 16:31:02 2007, Doug Preddy, Start Up, Started compressor 
Startewd compressor and purifer today.
Entry  Tue Apr 24 10:21:49 2007, Doug Preddy, Start Up, Started flow to turbines 
Started flow to turbines at about 10:00 in parallel warm mode.
Entry  Fri Apr 27 10:10:07 2007, Doug Preddy, Normal Running, Solenoid Cold 
Solenoid cold, power leads cold. Alarm handler started.
Entry  Fri Apr 27 10:32:07 2007, Doug Preddy, Quench Recovery, Reset quench relay logic box 
The quench relay logic controller was not functioning correctly. It showed 
a "Trip" light and no other lights. I cycled the power to the controller 
and it now seems to be working properly. This did cause a quench and I am 
Entry  Mon Apr 30 09:30:26 2007, Doug Preddy, Normal Running, problems over weekend 
The PLC Limit switch power supply was tripped off on Saturday. I came in 
and reset it and restarted all the vacuum systems that had tripped off. 
The power leads were found to be warming up. This morning they were at ~ 
Entry  Wed May 2 16:28:16 2007, Doug Preddy, Normal Running, Energizing solenoid to 500 amps 
Power supply initially would not ramp up. I exercised the rate control 
selector switch and found that the power supply would ramp up at only the 
30 minute setting. It is ramping up at a rate of about 60 minutes.
Entry  Wed May 16 10:02:35 2007, Doug Preddy, Maintenance, warming up power leads 
Power lead temps have been slowly increasing over the last wekk. I ramped 
the solenoid down and am warming the power leads to >80k.