Tuda logbooks
General
E1104
ISACII
E1195
E1103
E1147
S1287
S1203
S1233
S1284
S1447 23Na(a,p)
S1540
S1381
S1847 Catania
S1287
Fields marked with
*
are required
Entry time:
Thu Nov 14 20:33:52 2024
Author
*
:
Type:
- please select -
Routine
Other
Category:
- please select -
General
Other
Subject:
> > 09.20 TUDA2.PNG5 1.04E-06Torr > See attachments 1 & 2 for EPICS log of pump down and overnight > pumping > > 09.37 Vent TUDA > > 14.00 S2-2 and W positions swapped > Coolant loop disconnected - needs to be re-tested > > 15.08 Time to 250mTorr ~12m > Time from 250mTorr to 4.75-05Torr ~28m > See attachment 3 > > 15.36 FTS RS44 set point +5.5 deg C > > 17.00 TUDA2.PNG5 4.5E-06Torr > > 17.14 FTS RS44 set point +22 deg C > > 17.31 TUDA2.PNG5 6.2E-06Torr > > Cryo closed and turbo pump switched off. Will monitor leak-in rate > from good vacuum (< 1E-05 Torr) overnight. >
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
ELOG V2.9.2-2455