Entry time:
Wed Apr 13 09:37:51 2011
Author
*
:
Type:
- please select -
Routine
Other
Category:
- please select -
General
Other
Subject:
09.20 TUDA2.PNG5 1.04E-06Torr See attachments 1 & 2 for EPICS log of pump down and overnight pumping 09.37 Vent TUDA 14.00 S2-2 and W positions swapped Coolant loop disconnected - needs to be re-tested 15.08 Time to 250mTorr ~12m Time from 250mTorr to 4.75-05Torr ~28m See attachment 3 15.36 FTS RS44 set point +5.5 deg C 17.00 TUDA2.PNG5 4.5E-06Torr 17.14 FTS RS44 set point +22 deg C 17.31 TUDA2.PNG5 6.2E-06Torr Cryo closed and turbo pump switched off. Will monitor leak-in rate from good vacuum (< 1E-05 Torr) overnight.