Tuda logbooks General E1104 ISACII E1195 E1103 E1147 S1287 S1203 S1233 S1284 S1447 23Na(a,p) S1540 S1381  S1847 Catania
  S1287  ELOG logo
Message ID: 59     Entry time: Tue Dec 4 05:40:10 2012
Author: Jessica Tomlinson 
Type: Other 
Category: Other 
Subject: analysis - single strip E vs t  

 I have attached some single strip energy vs time plots to show that the structure seen in the collective E vs t plots is in each strip. 

Attachments 1 and 2 show two different strips in S2-2 with no cuts. You can see that the 3 loci are different to one another in the single strips as well as in the combined S2-2 E vs t plot shown in attachment 3. Attachement 3 is plotted after applying the following cuts: equal energy in the front and back of each detector, low energy cut, must be a hit in each detector, multiplicity must be greater than 3. The 3 different loci are a beam pulse apart. The energy cal is 40keV/chan. 

Attachments 4 and 5 show energy vs time for single strips in S2-1 with no cuts. Attachment 6 shows the combined front strips in S2-1 after the same cuts as above for the combined S2-2 plot have been applied. Now there are 2 loci in the same beam pulse which are still there after the cuts. Not sure what the second line is. The energy cal in attachment 6 is 80keV/chan.

Attachment 1: s2077-evst-s22-1strip.png  121 kB  Uploaded Tue Dec 4 03:52:31 2012  | Show | Hide all | Show all
Attachment 2: s2090-evst-s22-1strip.png  121 kB  Uploaded Tue Dec 4 04:35:27 2012  | Show | Hide all | Show all
Attachment 3: S2-2-Evst-p93bk4.png  144 kB  Uploaded Tue Dec 4 04:35:47 2012  | Hide | Hide all | Show all
S2-2-Evst-p93bk4.png
Attachment 4: s2006-evst-s21-1strip.png  119 kB  | Show | Hide all | Show all
Attachment 5: s2017-Evst-s21-1strip.png  129 kB  | Hide | Hide all | Show all
s2017-Evst-s21-1strip.png
Attachment 6: s2-1-Evst-p93bk4.png  176 kB  | Show | Hide all | Show all