E1104
ISACII
E1195
E1103
E1147
S1287
S1203
S1233
S1284
S1447 23Na(a,p)
S1540
S1381
S1847 Catania
S1287
Fields marked with
*
are required
Entry time:
Sat Jan 10 17:22:23 2026
Author
*
:
Type:
- please select -
Routine
Other
Category:
- please select -
General
Other
Subject:
<p>Attached plot shows rate of pressure increase in ~2l volume upstream of TUDA with Ni window #2 installed on reentrant flange. Comparison between</p> <p>a)pumping out TUDA and 2l volume, and</p> <p>b) filling TUDA with ~340mbar of He and pumping out 2l volume.</p> <p>In both cases, 2l volume isolated from pumps at start.</p> <p> </p> <p>Gradients are comparible, hence no detectable He leak through window.</p> <p> </p> <p>Will repeat with other windows and blank flange tomorrow.</p>
Encoding
:
HTML
ELCode
plain
Suppress Email notification