Version 2 of ...
- re-test chamber/coolant loop with vacuum cycle - <b>done</b>
- install MSL Type W preamp - <b>done</b>
- to adjust position of support ring, temporarily loosen grub
screws of <b class="moz-txt-star"><span class="moz-txt-tag">*</span>all<span class="moz-txt-tag">*</span></b> support rings - secure <b class="moz-txt-star"><span class="moz-txt-tag">*</span>all<span class="moz-txt-tag">*</span></b> grub screws
when finished
- use 9/16" spanner and adjustable wrench for mechanical support
for Swagelok connectors - <b>done</b>
- test for gross leaks at air - <b>done</b>
- test preamp with pulser - <b>done</b>
- re-test coolant loop with vacuum cycle - <b>done</b>
- move linear translator from position #3 to #1 - <b>done</b>
- align target ladders & re-entrant flange window
- select appropriate collimator (TUDA chamber entrance flange)
- install & test Faraday cup
- Ortec 439 or EPICS readout
for the former, scaler readout is setup
for the latter, startup EPICS TUDA diagnostics
- install & test CCTV camera to view ZnS scintillator
- remember we will replace viewport with H2 feedthrough
- TV monitor in ISAC Control Room
- connect via cable #2
- use cable #1 for Keithley 610C Electrometer
- mark screen with position of 3mm dia. tuning aperture
- install additional thermocouple to monitor ambient gas temperature?
- TUDA S Cupboard ISAC-I for thermocouple & switch
- install & test re-entrant flange windows with He
- assumes bypass has been installed
- install detectors
- measure actual positions of detectors
- alpha & pulser calibration
- install detector shields for <b class="moz-txt-star"><span class="moz-txt-tag">*</span>all<span class="moz-txt-tag">*</span></b> detectors for beam tuning
- tune all required beam energies up front
- grab screenshots of each tune from ISAC Ops log to obtain
exact beam energy from ToF diagnostic
- remove detector shields
- RF delay timing (with beam)</pre>