Tuda logbooks
General
E1104
ISACII
E1195
E1103
E1147
S1287
S1203
S1233
S1284
S1447 23Na(a,p)
S1540
S1381
S1847 Catania
S1233
Fields marked with
*
are required
Entry time:
Wed Dec 17 19:35:34 2025
Author
*
:
Type:
- please select -
Routine
Other
Category:
- please select -
General
Other
Subject:
<p> VME crate (master): VME V430 </p> <p>1x MVME2431 VME processor module, slot 1</p> <p>1x SBS master, slot 2</p> <p>1x SAC S9418 ACQ controller, slot 5 - trigger condition set here</p> <p>14x Silena 9418 ADC, slot 7-10 (geographic location doesn't matter only jumper address)</p> <p> </p> <p>VME crate (slave): VME430 - we didn't use this since the power supply died but we still needed a VME crate connected to the master crate</p> <p>1x SBS slave, slot 1 (needs to be plugged in!)</p> <p>2x CAEN V560 scalers</p> <p>4x CAEN V1190 TDC</p> <p> </p> <p>Debugging notes:</p> <p>Turn slave crates on first and master crate last</p> <p>VME crates must be of type V430 for MIDAS UK. We used V64x crates and unable to read scaler and TDC modules. Error messages result if you try to load modules in the wrong type caret into MIDAS UK. </p> <p>HOWEVER, SBS modules MUST be plugged in. Probably some firmware coding in MVME2431 tells MIDAS there are two VME crates. SIGBUS error results if slave crate is not connected to the master crate.</p>
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
ELOG V2.9.2-2455