15-02-2024
The coil PS control card is fried. Roughly since the beginning of December. The PS was removed.
See attached image.
New PS was installed on Friday.
Tuesday, February 20, 2024, 08:01: Check Coil not responding.
Tuesday, February 20, 2024, 09:48: System check 2: TGHT / TBHT manually to 5 / 5 A. BIAS / EE / EZL to 1 / -0.5 / 1 kV. Anode to 50 V.
Coil set to 15 A, did not adjust.
Tuesday, February 20, 2024, 12:04: Coil PS SW1-7 switch in was in an "Off" position & wouldn't allow the voltage signal through to the PS. Switched to "on". Coil tapped to 7 A manually & back to 0 A ok.
Fixed*
|