Logbooks Lab Maintenance Evaporator_1 Evaporator_2 Laser cutter Target Production Test-Stand RH-ISAC RH-Cyclotron RH-Meson Hall RH-Beamlines RH-ARIEL
  Test-Stand  Not logged in ELOG logo
Message ID: 931     Entry time: Wednesday, June 08, 2022, 10:01
Author: Aaron Schmidt 
Target/No.#: Ta#64 
Source: LP-SIS 
Station: ITE 
Module: TM4 
Target Oven W.O#: 44691-12 
Heat Shield W.O#: 44378-13 
Status: In progress 
Subject: Conditioning Ta#64 & HV Test. 

Thermocouples connected. Test HV

Wednesday, June 08, 2022, 09:33: Roughing.

Wednesday, June 08, 2022, 10:07: TP1B on.

Wednesday, June 08, 2022, 10:13: TP1B up to speed. TP1A on.

Wednesday, June 08, 2022, 10:21: TP1A up to speed. IG1 on.

Wednesday, June 08, 2022, 13:06: Waterline FS#5 Open. Vacuum @ 3.32-e06T.

Wednesday, June 08, 2022, 13:22: BIAS / EE / EZL @ 30 / -1.5 / 13 kV. OK.

Wednesday, June 08, 2022, 13:26: One spark*.

Wednesday, June 08, 2022, 13:32: BIAS / EE / EZL @ 35 / -1.8 / 15 kV.

Wednesday, June 08, 2022, 13:33: Spark*

Wednesday, June 08, 2022, 13:38: BIAS / EE / EZL @ 40 / -2.0 / 17 kV

Wednesday, June 08, 2022, 13:42: Spark*. Outside the cage (EE) Feedthrough (K) Between a hex edge & one of the resistors on a plastic plate.

Wednesday, June 08, 2022, 13:50: BIAS / EE / EZL @ 45 / -2.3 / 19 kV

Wednesday, June 08, 2022, 14:18: We are back. Spark tripped the computer & shut it off while tapping up the HV. BIAS / EE / EZL @ 48.3 / 2.4 / 20 kV.

Wednesday, June 08, 2022, 14:22: BIAS / EE / EZL @ 50 / -2.5 / 20 kV.

Wednesday, June 08, 2022, 14:45: Spark* In the cage.

Wednesday, June 08, 2022, 14:57: Spark* In the cage.

Wednesday, June 08, 2022, 15:19: No sparks have occurred off the thermocouples.  Begin tapping down the HV.

Wednesday, June 08, 2022, 15:29: HV off. TGHT / TBHT Check @ 5 / 5 A OK.

Wednesday, June 08, 2022, 15:31: IV4 open.


Thursday, June 09, 2022, 05:14: Auto-ramping TGHT / TBHT to 350 / 150 A @ 2 A per min. 

Thursday, June 09, 2022, 08:21: BIAS / EE / EZL @ 30 / -1.5 / 13 kV. OK.

Thursday, June 09, 2022, 08:22: Auto-ramping TGHT / TBHT to 350 / 150 A @ 2 A per min

Thursday, June 09, 2022, 08:29: Optics set. Standard tune.

Thursday, June 09, 2022, 10:22: Beginning wire & mass scans.

Thursday, June 09, 2022, 10:30: Adjust TSG:CB1 (ok)

Thursday, June 09, 2022, 11:40: Mass & wire scans complete. Tune set to standard. HV tests begin. BIAS / EE / EZL @ 35 / -1.8 / 15 kV.

Thursday, June 09, 2022, 11:45: BIAS / EE / EZL @ 40 / -2.0 / 17 kV

Thursday, June 09, 2022, 13:01: BIAS / EE / EZL @ 45 / -2.3 / 19 kV

Thursday, June 09, 2022, 13:13: Small spark* Outside the chamber, but in the cage.

Thursday, June 09, 2022, 13:22: BIAS / EE / EZL @ 50 / -2.5 / 20 kV. (One spark during the tap up. In the cage)

Thursday, June 09, 2022, 14:03: Small spark, barely audible.

Thursday, June 09, 2022, 14:11: Large spark in the cage* Computer was shut off. Got it back around 14:16

Thursday, June 09, 2022, 14:20: Reset steering, optics, IG6.

Thursday, June 09, 2022, 14:28: Begin tapping down HV

Thursday, June 09, 2022, 14:34: HV to 0.

Thursday, June 09, 2022, 14:36: Closed IV4. (Reset w/ screwdriver)


Friday, June 10, 2022, 09:58: Auto-Cooling @ 2 A per min.  


Monday, June 13, 2022, 10:16: IG1 off. Turbo pumps off.

Attachment 1: Ta#64_Conditioning_HV..docx  384 kB  Uploaded Wednesday, June 08, 2022, 16:30
Attachment 2: Ta#64_Conditioning_&_HV_test_-_2.docx  1.481 MB  Uploaded Thursday, June 09, 2022, 15:37
Attachment 3: ER-580-230-30kV-M7.txt  3 kB  Uploaded Thursday, June 09, 2022, 15:37  | Show | Hide all
Attachment 4: ER-580-230-30kV-M23.txt  3 kB  Uploaded Thursday, June 09, 2022, 15:37  | Show | Hide all
Attachment 5: ER-580-230-30kV-M40.txt  3 kB  Uploaded Thursday, June 09, 2022, 15:38  | Hide | Hide all
value (Voltage offset      40.0000)
     75.0000
      1.0000
     95.0000
  29978.0220
21
21
./emit.dat
 -0.1080 -0.0702 -0.0894 -0.0940 -0.0763 -0.0891 -0.0916 -0.0824 -0.0919 -0.0919 -0.0864 -0.0909 -0.0925 -0.0848 -0.0876 -0.0742 -0.0781 -0.0784 -0.0330 0.0363 -0.0443
 -0.0702 -0.0681 -0.0867 -0.0867 -0.0684 -0.0854 -0.0839 -0.0781 -0.0839 -0.0864 -0.0815 -0.0824 -0.0891 -0.0790 -0.0818 -0.0732 -0.0720 -0.0363 0.0253 -0.0330 -0.0677
 -0.0720 -0.0616 -0.0613 -0.0690 -0.0635 -0.0772 -0.0592 -0.0607 -0.0833 -0.0854 -0.0583 -0.0616 -0.0885 -0.0656 -0.0601 -0.0693 -0.0369 0.0461 -0.0443 -0.0659 -0.0668
 -0.0729 -0.0638 -0.0650 -0.0696 -0.0650 -0.0809 -0.0690 -0.0635 -0.0830 -0.0842 -0.0653 -0.0620 -0.0882 -0.0696 -0.0650 -0.0107 0.0418 -0.0403 -0.0778 -0.0671 -0.0677
 -0.0711 -0.0644 -0.0641 -0.0677 -0.0656 -0.0815 -0.0681 -0.0641 -0.0854 -0.0833 -0.0644 -0.0629 -0.0867 -0.0641 -0.0217 0.0269 -0.0574 -0.0662 -0.0763 -0.0644 -0.0702
 -0.0708 -0.0644 -0.0641 -0.0684 -0.0635 -0.0754 -0.0650 -0.0647 -0.0839 -0.0858 -0.0644 -0.0650 -0.0830 -0.0156 0.0656 -0.0583 -0.0739 -0.0641 -0.0714 -0.0626 -0.0677
 -0.0705 -0.0638 -0.0647 -0.0644 -0.0613 -0.0732 -0.0653 -0.0635 -0.0818 -0.0867 -0.0647 -0.0629 -0.0189 0.0793 -0.0455 -0.0687 -0.0739 -0.0650 -0.0696 -0.0629 -0.0671
 -0.0702 -0.0629 -0.0650 -0.0668 -0.0638 -0.0717 -0.0647 -0.0653 -0.0839 -0.0854 -0.0620 -0.0150 0.0168 -0.0470 -0.0616 -0.0690 -0.0742 -0.0641 -0.0693 -0.0647 -0.0668
 -0.0708 -0.0653 -0.0635 -0.0638 -0.0644 -0.0757 -0.0653 -0.0650 -0.0824 -0.0739 -0.0104 0.0470 -0.0809 -0.0674 -0.0629 -0.0668 -0.0751 -0.0647 -0.0690 -0.0644 -0.0668
 -0.0711 -0.0659 -0.0632 -0.0632 -0.0638 -0.0726 -0.0629 -0.0626 -0.0781 -0.0153 0.0430 -0.0549 -0.0864 -0.0629 -0.0668 -0.0659 -0.0735 -0.0638 -0.0790 -0.0620 -0.0696
 -0.0693 -0.0629 -0.0635 -0.0650 -0.0641 -0.0699 -0.0650 -0.0583 -0.0293 -0.0052 -0.0430 -0.0708 -0.0864 -0.0635 -0.0629 -0.0668 -0.0751 -0.0626 -0.0748 -0.0632 -0.0641
 -0.0684 -0.0632 -0.0632 -0.0659 -0.0626 -0.0711 -0.0482 -0.0140 -0.0064 -0.0757 -0.0644 -0.0684 -0.0879 -0.0644 -0.0635 -0.0690 -0.0714 -0.0638 -0.0778 -0.0635 -0.0659
 -0.0687 -0.0629 -0.0629 -0.0662 -0.0650 -0.0620 -0.0150 0.0009 -0.0671 -0.0848 -0.0641 -0.0659 -0.0873 -0.0659 -0.0626 -0.0665 -0.0745 -0.0641 -0.0742 -0.0653 -0.0650
 -0.0693 -0.0623 -0.0632 -0.0644 -0.0546 -0.0302 -0.0064 -0.0461 -0.0839 -0.0836 -0.0671 -0.0690 -0.0867 -0.0641 -0.0635 -0.0674 -0.0769 -0.0653 -0.0711 -0.0644 -0.0644
 -0.0690 -0.0641 -0.0620 -0.0470 -0.0217 -0.0235 -0.0583 -0.0632 -0.0821 -0.0821 -0.0635 -0.0650 -0.0851 -0.0653 -0.0638 -0.0665 -0.0754 -0.0638 -0.0739 -0.0635 -0.0653
 -0.0690 -0.0644 -0.0595 -0.0278 -0.0299 -0.0604 -0.0632 -0.0632 -0.0833 -0.0821 -0.0632 -0.0681 -0.0867 -0.0632 -0.0653 -0.0677 -0.0751 -0.0650 -0.0797 -0.0638 -0.0644
 -0.0671 -0.0629 -0.0235 -0.0284 -0.0525 -0.0711 -0.0662 -0.0632 -0.0821 -0.0845 -0.0632 -0.0674 -0.0885 -0.0620 -0.0638 -0.0677 -0.0732 -0.0629 -0.0632 -0.0653 -0.0638
 -0.0684 -0.0134 -0.0436 -0.0604 -0.0632 -0.0693 -0.0677 -0.0644 -0.0806 -0.0839 -0.0629 -0.0644 -0.0879 -0.0650 -0.0616 -0.0684 -0.0711 -0.0635 -0.0775 -0.0647 -0.0656
 -0.0302 -0.0485 -0.0513 -0.0665 -0.0653 -0.0705 -0.0659 -0.0653 -0.0836 -0.0839 -0.0665 -0.0665 -0.0842 -0.0629 -0.0613 -0.0668 -0.0760 -0.0656 -0.0635 -0.0641 -0.0620
 -0.0568 -0.0470 -0.0629 -0.0644 -0.0647 -0.0708 -0.0674 -0.0629 -0.0824 -0.0824 -0.0632 -0.0662 -0.0854 -0.0623 -0.0623 -0.0665 -0.0735 -0.0629 -0.0674 -0.0620 -0.0656
 -0.0610 -0.0650 -0.0635 -0.0616 -0.0623 -0.0687 -0.0659 -0.0626 -0.0854 -0.0815 -0.0638 -0.0665 -0.0879 -0.0635 -0.0641 -0.0677 -0.0739 -0.0623 -0.0665 -0.0610 -0.0656
 
Attachment 6: ER-580-230-30kV-M86.txt  3 kB  Uploaded Thursday, June 09, 2022, 15:38  | Show | Hide all
Attachment 7: ER-580-230-30kV-M133.txt  3 kB  Uploaded Thursday, June 09, 2022, 15:38  | Show | Hide all
ELOG V2.9.2-2455