Message ID: 817     Entry time: Tuesday, October 27, 2020, 08:54
Author: Aaron Schmidt 
Target/No.#: Nb#9  
Source: LP-SIS 
Station: ITW 
Module: TM2 
Target Oven W.O#: 44508-2 
Heat Shield W.O#: 43515-4 
Status: In progress 
Subject: Nb#9 Conditioning 

Tuesday, October 27, 2020, 08:40: Roughing.

Tuesday, October 27, 2020, 09:13: TP1B on.

Tuesday, October 27, 2020, 10:13: TP1A on.

Tuesday, October 27, 2020, 12:59: Vacuum @ 4.55e-06T

Tuesday, October 27, 2020, 13:01: Waterlines opened. Vacuum @ 4.26e-06T.

Tuesday, October 27, 2020, 13:03: Power / sources on.

Tuesday, October 27, 2020, 13:05: Optics on. IV4 open. TGHT / TBHT check ok.


Wednesday, October 28, 2020, 04:30: Auto-ramping TGHT / TBHT  to 350 / 150 A @ 2 A per min. 

Wednesday, October 28, 2020, 06:30:

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 26.66 5.67
 TBHT 26.51

8.40

 COIL + EE 26.43 2.59
 TGHT 26.52 6.44

Wednesday, October 28, 2020, 06:42: BIAS / EE / EZL @ 30 / -1.5 / 13 kV.

Wednesday, October 28, 2020, 07:33: Auto-ramping TGHT / TBHT  to 580 / 230 A @ 2 A per min.

Wednesday, October 28, 2020, 13:03:

 Device Temp
[C]
 Flow
[lpm]
 HS + MSP 27.25 6.62
 TBHT 26.90 9.27
 COIL + EE 26.68 3.08
 TGHT 26.88 6.85

 


Thursday, October 29, 2020, 15:19: HV to 0 kV. Begin auto-cooling @ 1 A per min.


Large Vacuum spike noted 5:30 approx. Tripped TS.

Monday, November 02, 2020, 08:32: Vacuum @ 2.65e-06T. PS/optics off (Sources already tripped)

Monday, November 02, 2020, 08:33: IV4 closed.

https://elog.triumf.ca/TIS/Test-Stand/818 - Leak Checks (FS#4 confirmed)

Monday, November 02, 2020, 09:27: IG1 off. Turbo pumps off.

Monday, November 02, 2020, 10:38: Venting.

Attachment 1: Nb#9_Conditioning.docx  1.136 MB  Uploaded Monday, November 09, 2020, 11:42
Attachment 2: ER-580-230-30kV-M133.txt  3 kB  Uploaded Monday, November 09, 2020, 11:42  | Show | Hide all
Attachment 3: ER-580-230-30kV-M23.txt  3 kB  Uploaded Monday, November 09, 2020, 11:42  | Hide | Hide all
value (Voltage offset      40.0000)
     75.0000
      1.0000
     95.0000
  29992.6740
21
21
./emit.dat
 -0.1556 -0.1151 -0.0980 -0.1022 -0.0198 0.0833 -0.1035 -0.1001 -0.1001 -0.1025 -0.1016 -0.0986 -0.0958 -0.0995 -0.1019 -0.0970 -0.1007 -0.0995 -0.1007 -0.0992 -0.1028
 -0.1199 -0.1111 -0.0977 -0.1028 -0.0909 -0.0772 0.1352 -0.0201 -0.0970 -0.0946 -0.0937 -0.0974 -0.0958 -0.0912 -0.1001 -0.0934 -0.0912 -0.0937 -0.1004 -0.0970 -0.0995
 -0.1169 -0.0909 -0.0912 -0.0934 -0.0946 -0.0937 -0.0830 -0.0598 0.1181 -0.0879 -0.0873 -0.0958 -0.0952 -0.0894 -0.0916 -0.0955 -0.0891 -0.0946 -0.0928 -0.0958 -0.0964
 -0.1175 -0.0961 -0.0986 -0.0916 -0.0958 -0.0958 -0.0928 -0.0934 -0.0891 0.0388 0.0153 -0.0916 -0.0937 -0.0912 -0.0900 -0.0961 -0.0909 -0.0931 -0.0912 -0.0992 -0.0967
 -0.1175 -0.0940 -0.0989 -0.0940 -0.0946 -0.0958 -0.0946 -0.0934 -0.0934 -0.0946 -0.0815 0.0415 -0.0793 -0.0934 -0.0900 -0.0922 -0.0934 -0.0961 -0.0934 -0.0955 -0.0955
 -0.1190 -0.0934 -0.0980 -0.0952 -0.0922 -0.0970 -0.0964 -0.0940 -0.0961 -0.0961 -0.0906 -0.0897 -0.0519 0.0281 -0.0879 -0.0922 -0.0928 -0.0934 -0.0961 -0.0974 -0.0998
 -0.1163 -0.0925 -0.0958 -0.0952 -0.0931 -0.0952 -0.0943 -0.0949 -0.0946 -0.0949 -0.0992 -0.0916 -0.0928 -0.0882 -0.0555 -0.0568 -0.0922 -0.0919 -0.0980 -0.0961 -0.1016
 -0.1151 -0.0974 -0.0925 -0.0912 -0.0931 -0.0925 -0.0903 -0.0934 -0.0961 -0.0931 -0.0940 -0.0937 -0.0934 -0.0925 -0.0876 -0.0815 -0.0214 -0.0726 -0.0894 -0.0974 -0.1001
 -0.1163 -0.0940 -0.0943 -0.0937 -0.0946 -0.0928 -0.0909 -0.0934 -0.0937 -0.0931 -0.0909 -0.0922 -0.0937 -0.0925 -0.0922 -0.0894 -0.0912 -0.0778 -0.0385 -0.0949 -0.1007
 -0.1151 -0.0937 -0.0964 -0.0964 -0.0943 -0.0961 -0.0922 -0.0952 -0.0934 -0.0940 -0.0928 -0.0909 -0.0931 -0.0934 -0.0937 -0.0928 -0.0925 -0.0934 -0.0897 -0.0684 -0.0696
 -0.1160 -0.0916 -0.0940 -0.0955 -0.0952 -0.0952 -0.0919 -0.0946 -0.0928 -0.0931 -0.0983 -0.0922 -0.0934 -0.0912 -0.0931 -0.0925 -0.0922 -0.0919 -0.0980 -0.0934 -0.0961
 -0.1163 -0.0934 -0.0912 -0.0937 -0.0928 -0.0912 -0.0912 -0.0916 -0.0961 -0.0900 -0.0931 -0.0909 -0.0952 -0.0922 -0.0906 -0.0955 -0.0897 -0.0928 -0.0958 -0.0949 -0.0937
 -0.1151 -0.0940 -0.0922 -0.0934 -0.0940 -0.0919 -0.0916 -0.0925 -0.0949 -0.0922 -0.0931 -0.0919 -0.0949 -0.0922 -0.0906 -0.0970 -0.0919 -0.0900 -0.0931 -0.0958 -0.0974
 -0.1147 -0.0937 -0.0946 -0.0925 -0.0934 -0.0916 -0.0900 -0.0922 -0.0928 -0.0928 -0.1047 -0.0916 -0.0943 -0.0912 -0.0925 -0.0934 -0.0925 -0.0934 -0.0928 -0.0977 -0.0992
 -0.1166 -0.0934 -0.0952 -0.0937 -0.0909 -0.0919 -0.0934 -0.0912 -0.0906 -0.0946 -0.0937 -0.0922 -0.0943 -0.0925 -0.0916 -0.0916 -0.0928 -0.0906 -0.0943 -0.0946 -0.0964
 -0.1184 -0.0922 -0.0928 -0.0934 -0.0906 -0.0919 -0.0922 -0.0912 -0.0916 -0.0934 -0.0931 -0.0882 -0.0925 -0.0919 -0.0928 -0.0909 -0.0916 -0.0916 -0.0977 -0.0925 -0.1025
 -0.1166 -0.0934 -0.0906 -0.0934 -0.0946 -0.0891 -0.0909 -0.0916 -0.0903 -0.0900 -0.0989 -0.0909 -0.0912 -0.0906 -0.0912 -0.0922 -0.0897 -0.0916 -0.0919 -0.0925 -0.0934
 -0.1160 -0.0958 -0.0940 -0.0946 -0.0946 -0.0906 -0.0888 -0.0928 -0.0916 -0.0912 -0.0946 -0.0928 -0.0949 -0.0916 -0.0934 -0.0925 -0.0909 -0.0931 -0.0909 -0.0952 -0.0970
 -0.1154 -0.0937 -0.0934 -0.0955 -0.0934 -0.0928 -0.0906 -0.0919 -0.0925 -0.0916 -0.0888 -0.0928 -0.0940 -0.0897 -0.0922 -0.0919 -0.0934 -0.0940 -0.0967 -0.0937 -0.1010
 -0.1151 -0.0919 -0.0974 -0.0934 -0.0925 -0.0940 -0.0940 -0.0912 -0.0919 -0.0928 -0.0903 -0.0900 -0.0928 -0.0903 -0.0912 -0.0922 -0.0943 -0.0937 -0.0964 -0.0940 -0.0946
 -0.1141 -0.0912 -0.0934 -0.0937 -0.0919 -0.0919 -0.0928 -0.0922 -0.0900 -0.0931 -0.1035 -0.0894 -0.0912 -0.0906 -0.0900 -0.0925 -0.0922 -0.0885 -0.0922 -0.0952 -0.1013