Salts:
Li - 10 uL
Na - 10 uL
K - 5 uL
Rb - 10 uL
Cs - 15 uL
Wednesday, June 10, 2020, 08:55: Roughing.
Wednesday, June 10, 2020, 09:30: Turbo pumps on.
Wednesday, June 10, 2020, 09:35: Ion gauge on.
Wednesday, June 10, 2020, 13:52: HS waterline opened. Vacuum ok @ 4.11e-06T.
Wednesday, June 10, 2020, 13:59: All main PS on. E-rig PS on. Optics on.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
27.21 |
5.81 |
TBHT |
27.04 |
8.83
|
COIL + EE |
26.93 |
2.76 |
TGHT |
27.04 |
6.58 |
Wednesday, June 10, 2020, 14:02: Sources on.
Thursday, June 11, 2020, 07:15: Auto-ramping TGHT / TBHT to 580 / 230 A @ 2 A per min.
Thursday, June 11, 2020, 07:29: IV4 opened. BIAS / EZL / EE @ 30 / 13 / -1.5 kV.
Thursday, June 11, 2020, 09:56: BIAS reduced to 20 kV due to sparking / out gassing. MB1 having issues?
Thursday, June 11, 2020, 14:24: Halted auto ramp due to the vacuum. TGHT @ 500 A. Vacuum @ 7.08e-06T.
Thursday, June 11, 2020, 14:26: TBHT reduced to 200 A.
Thursday, June 11, 2020, 19:09: Auto-cooling @ 1 A per min.
Friday, June 12, 2020, 10:43: Auto-ramping TGHT / TBHT to 300 / 150 A @ 2 A per min.
Monday, June 15, 2020, 03:00: Vacuum @ 2.66e-06T. TGHT / TBHT @ 300 / 150 A. Auto-ramping TGHT / TBHT to 580 / 230 A @ 2 A per min.
Monday, June 15, 2020, 05:30: Tap up BIAS. BIAS / EZL / EE @ 30 / 13 / -1.5 kV. TGHT @ 512 A. Vacuum @ 6.11e-06T.
Monday, June 15, 2020, 06:46:
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
26.40 |
7.51 |
TBHT |
26.08 |
10.15
|
COIL + EE |
25.88 |
3.53 |
TGHT |
26.06 |
7.31 |
Mass scans complete w/ TGHT / TBHT @ 512 / 230 A. 30 kV.
Mass, wire, & emittance scans taken on: M7, M23, M38, M86, M125.
Monday, June 15, 2020, 09:52: Auto-cooling @ 1 A per min.
Monday, June 15, 2020, 21:23: BIAS / EE / EZL @ 0 kV.
Tuesday, June 16, 2020, 06:57: IV4 closed & reset.
Tuesday, June 16, 2020, 07:00: Sources / Power off. IG1 off. Turbo pumps off.
Tuesday, June 16, 2020, 08:06: Waterlines purged. N bottle @ 1450 PSI. Venting. |