Monday, March 02, 2020, 14:36: Roughing.
Monday, March 02, 2020, 15:08: Turbo pumps on.
Remove HV cage (top). Reroute FS#3 back to cool EE
Tuesday, March 03, 2020, 08:11: Waterlines open. Vacuum ok @ 6.52e-07T.
HV cage top on. Data Logger removed. Wiring grounded.
Tuesday, March 03, 2020, 09:00: Power / sources / HV on.
Tuesday, March 03, 2020, 09:03: Auto-ramping TGHT / TBHT to 300 / 150 A @ 4 A per min.
Device |
Temp
[C] |
Flow
[lpm] |
HS + MSP |
25.65 |
4.38 |
TBHT |
25.50 |
6.97
|
COIL + EE |
25.39 |
1.06 |
TGHT |
25.82 |
5.84 |
Tuesday, March 03, 2020, 09:08: BIAS / EE / EZL on.
Tuesday, March 03, 2020, 09:11: BIAS / EE / EZL @ 10 / 7.5 / -1 kV.
Tuesday, March 03, 2020, 09:15: BIAS / EE / EZL @ 20 / 12.5 / -1.5 kV. Read back on PS same / ok.
Tuesday, March 03, 2020, 09:17: BIAS / EE / EZL @ 30 / 12.5 / -1.5 kV.
Tuesday, March 03, 2020, 09:18: IV4 Opened.
Tuesday, March 03, 2020, 09:24: Optics on / set.
One spark noted. Reset optics.
Tuesday, March 03, 2020, 11:32: Auto-ramping TGHT / TBHT to 500 / 200 A @ 4 A per min.
One spark noted. Reset optics.
Tuesday, March 03, 2020, 12:47: Auto-ramping TGHT / TBHT to 580 / 230 A @ 4 A per min.
FM
Tuesday, March 03, 2020, 18:30: BIAS/EE/EZL @ 0 kV.
Tuesday, March 03, 2020, 18:31: TGHT / TBHT to 400 / 200 A @ 4 A per min.
Nothing observed on the emittance scan regardless of the tune
Wednesday, March 04, 2020, 09:09: Auto-ramping TGHT / TBHT to 500 / 210 A @ 4 A per min.
Wednesday, March 04, 2020, 10:47: Auto-ramping TGHT / TBHT to 580/ 230 A @ 4 A per min.
*Got busy in the morning, starting later than expected
Wednesday, March 04, 2020, 13:47: BIAS / EE / EZL @ 30 / 12.5 / -1.5 kV.
*Seminar at 2:30 I'll continue later
Wednesday, March 04, 2020, 16:08: Emittance rig troubleshoot (see e-log 778)
Wednesday, March 04, 2020, 19:00: BIAS/EE/EZL @ 0 kV.
Wednesday, March 04, 2020, 19:09: TBHT to 200A to try preserver some salts for tomorrow tests
Friday, March 06, 2020, 17:50: Auto-cooling 4 A/min .
Monday, March 09, 2020, 12:08: IV4 closed
Monday, March 09, 2020, 12:08: IG1 Off.
Monday, March 09, 2020, 12:09: Turbos off
Monday, March 09, 2020, 16:19: Venting |