Ref:
https://elog.triumf.ca/TIS/Test-Stand/679
Pin configuration between top 9-pin connector on Source Tray and terminal in high voltage rack. Measured 2018-08-21 by Aaron and Anders
Grounded |
|
|
X |
|
|
|
|
|
|
Source Tray Side (looking head on) |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
8 |
9 |
High Voltage Cage Side |
5 |
4 |
3 |
2 |
1 |
9 |
8 |
7 |
6 |
Pins connected source tray side: 1, 4, 5, 6
ST 9-pin HV rack
Pin to L TGHT - 1 - Black / COM / 5
Pin to R TGHT - 6 - Red / 9
Pin to L TBHT - 4 - Black / COM / 2
Pin to R TBHT - 5 - Red / 1
Tuesday, May 07, 2019, 15:20: Roughing.
Turbo pumps switched on w/ Ion gauge around 4pm.
Wednesday, May 08, 2019, 08:05: Multi-meters tested & connected.
Wednesday, May 08, 2019, 08:09: Power / sources on.
Wednesday, May 08, 2019, 08:13: RGA on
Wednesday, May 08, 2019, 08:14: Auto-ramping TGHT / TBHT to 575 (580) / 230 A @ 2A per min.
Wednesday, May 08, 2019, 08:30: Voltage reading off on TBHT? Connection to source tray side is wrong.
Wednesday, May 08, 2019, 08:40: Halt auto ramping. Manually tap down TBHT / TGHT
Wednesday, May 08, 2019, 08:42: Sources / RGA off.
Wednesday, May 08, 2019, 08:43: Turbo pumps off.
Wednesday, May 08, 2019, 10:40: Roughing.
Wednesday, May 08, 2019, 11:01: Turbo pumps on.
Wednesday, May 08, 2019, 13:10: Power / sources on. Vacuum @ 4.37e-06T
Wednesday, May 08, 2019, 13:19: Auto-ramping TGHT to 575 (580) A @ 2A per min. However, the vacuum is not quite ready. Increment value set to 3.000e-06T to ramp. Once the vacuum is reached, the heaters will start to ramp automatically.
Wednesday, May 08, 2019, 14:23: Auto-ramping ongoing successfully. TGHT @ 32A +. Vacuum @ 2.66e-06T
Thursday, May 09, 2019, 07:05: Auto-ramping TGHT to 575 (580) A @ 2A per min from 400 A.
Monday, May 13, 2019, 08:00: TSG:TP1A appears to have tripped over the weekend. (Around 7pm Saturday)
Closed IV4 / reset. Reset TP1A. Reset BV1A. Turn on TP1A.
Monday, May 13, 2019, 08:32: TBHT & TGHT switched back on.
Ramped down around 10 PM.
Tuesday, May 14, 2019, 07:10: Power / RGA off.
Tuesday, May 14, 2019, 07:12: IV4 closed, Ion gauge & turbo pumps off.
Tuesday, May 14, 2019, 08:35: Waterlines purged. Venting. |