RGA Ionizer & Filament replaced as per SRS RGA manual.
The ionizer was found to be dirty and the previous RGA filament was broken. Initially the old filament appeared to have warped and looked stuck to the anode grid/cage.
Both were replaced with clean parts. the support rods were also cleaned with acetone & methanol.
Parts used from the Replacement Ionizer Kit - SRS# O100RI
Repeller Cage - SRS Part# 7-00592
Anode Grid - SRS Part# 7-00591
Focus & Plate - SRS Part# 7-00590
ThO2/Ir Filament - SRS Part# 7-00611-720
Wednesday, December 05, 2018, 10:16: Roughing.
Probe bake out & RGA tuning required.
Wednesday, December 05, 2018, 11:08: Turbo pump on.
Wednesday, December 05, 2018, 11:13: Ion gauge on.
Wednesday, December 05, 2018, 13:38: TSG:BV1A Randomly shut off? TP1A tripped. Restarted.
Wednesday, December 05, 2018, 13:40: RGA switched on including the filament.
Thursday, December 06, 2018, 10:17: Vacuum @ 5.68e-07T on IG1 before Degas. Begin degas cycles 200 sec each, 1 min apart, 3 times.
Thursday, December 06, 2018, 10:30: Degas complete. Minimal change in IG1 Vac @ 5.70e-07T.
|