Message ID: 698     Entry time: Wednesday, December 05, 2018, 10:22
Author: Aaron Schmidt 
Target/No.#:  
Source:  
Station:  
Module:  
Target Oven W.O#:  
Heat Shield W.O#:  
Status: Problems/Solutions 
Subject: RGA Filament & Ionizer replacement 

RGA Ionizer & Filament replaced as per SRS RGA manual.

The ionizer was found to be dirty and the previous RGA filament was broken. Initially the  old filament appeared to have warped and looked stuck to the anode grid/cage.

Both were replaced with clean parts. the support rods were also cleaned with acetone & methanol.

 

Parts used from the Replacement Ionizer Kit - SRS# O100RI

 

Repeller Cage - SRS Part# 7-00592

Anode Grid - SRS Part# 7-00591

Focus & Plate - SRS Part# 7-00590

ThO2/Ir Filament - SRS Part# 7-00611-720

 

Wednesday, December 05, 2018, 10:16: Roughing.

Probe bake out & RGA tuning required.

Wednesday, December 05, 2018, 11:08: Turbo pump on.

Wednesday, December 05, 2018, 11:13: Ion gauge on.

Wednesday, December 05, 2018, 13:38: TSG:BV1A Randomly shut off? TP1A tripped. Restarted.

Wednesday, December 05, 2018, 13:40: RGA switched on including the filament.


Thursday, December 06, 2018, 10:17: Vacuum @ 5.68e-07T on IG1 before Degas. Begin degas cycles 200 sec each, 1 min apart, 3 times.

Thursday, December 06, 2018, 10:30: Degas complete. Minimal change in IG1 Vac @ 5.70e-07T.

 

Attachment 1: 20181205_RGA-Old_I&F.jpg  1.452 MB  Uploaded Wednesday, December 05, 2018, 12:55  | Show | Hide all
Attachment 2: 20181205_RGA-Old_I&F-2.jpg  1.788 MB  Uploaded Wednesday, December 05, 2018, 12:56  | Show | Hide all
Attachment 3: 20181205_RGA-Old_I&F-Anode_view.jpg  1.665 MB  Uploaded Wednesday, December 05, 2018, 12:56  | Hide | Hide all
20181205_RGA-Old_I&F-Anode_view.jpg
Attachment 4: 20181205_RGA-OLD_I&F-Broken_Filament_&_Dirty_Anode.jpg  1.607 MB  Uploaded Wednesday, December 05, 2018, 12:57  | Show | Hide all
Attachment 5: 20181205_RGA-Old_I&F-Dirty_Anode-Focus_Plate.jpg  1.647 MB  Uploaded Wednesday, December 05, 2018, 12:57  | Show | Hide all
Attachment 6: 20181205_RGA-Old_I&F-Dirty.jpg  1.452 MB  Uploaded Wednesday, December 05, 2018, 12:57  | Show | Hide all
Attachment 7: 20181205_RGA-New_A&R.jpg  1.641 MB  Uploaded Wednesday, December 05, 2018, 12:58  | Show | Hide all
Attachment 8: 20181205_RGA-New_Anode_&_Filament.jpg  1.528 MB  Uploaded Wednesday, December 05, 2018, 12:59  | Show | Hide all
Attachment 9: 20181205_RGA-New_Anode_&_Repeller_cages.jpg  1.717 MB  Uploaded Wednesday, December 05, 2018, 13:03  | Show | Hide all
Attachment 10: 20181205_RGA-New_Filament_&_Ionizer_assy.jpg  1.776 MB  Uploaded Wednesday, December 05, 2018, 13:04  | Show | Hide all
Attachment 11: 20181205_RGA-New_Filament_&_Ionizer_assy-2.jpg  2.272 MB  Uploaded Wednesday, December 05, 2018, 13:05  | Show | Hide all
Attachment 12: TS-2018-12-05-13-40_RGA_Check.rga  4.400 MB  Uploaded Wednesday, December 12, 2018, 13:06
ELOG V2.9.2-2455