Subject: Pump down empty chamber. Anode check 

TM2 Source tray still in.

Thursday, June 07, 2018, 09:20: Roughing.

Thursday, June 07, 2018, 12:42: Turbo pumps started.


Friday, June 08, 2018, 08:00: Anode run to 15 A. Short still apparent.

Friday, June 08, 2018, 08:12: Ion gauge, Turbos off.

Friday, June 08, 2018, 08:55: Venting.

Friday, June 08, 2018, 09:45: Anode connection investigated. Short found on the bracket once the anode wire was disconnected from the back screw (Picture)

Friday, June 08, 2018, 10:00: Removed 9-pin bracket. (Top connector left connected to TS. (Picture)

New Anode 9-pin connection/wire to be installed on the bracket. Test in the cold lab.

Friday, June 08, 2018, 10:12: Roughing.

Friday, June 08, 2018, 11:14: Turbo pumps started.

Friday, June 08, 2018, 12:31: Ion gauge on.


Tuesday, June 12, 2018, 06:18: Ion gauge & Turbo pumps off.

Tuesday, June 12, 2018, 07:40: Venting.

Attachment 1: 20180608_100402.jpg  960 kB  Uploaded Friday, June 08, 2018, 12:40  | Show | Hide all
Attachment 2: 20180608_100835.jpg  1.203 MB  Uploaded Friday, June 08, 2018, 12:40  | Show | Hide all
Attachment 3: 20180608_100840.jpg  960 kB  Uploaded Friday, June 08, 2018, 12:40  | Hide | Hide all
20180608_100840.jpg
Attachment 4: 20180613_080550.jpg  1.492 MB  Uploaded Wednesday, June 13, 2018, 09:52  | Show | Hide all
Attachment 5: 20180613_080605.jpg  1.428 MB  Uploaded Wednesday, June 13, 2018, 09:53  | Show | Hide all
ELOG V2.9.2-2455