Subject: Ramping up bias to 60kV with conditioning station supply 

Ramped up to 60kV with the new power supply, same as what we did last time with the standard test stand bias supply (see entry #627 ). Much more instability on the bias current readout (see attached screen shot), develops around 54kV. As well, the old supply read out 37uA at 60kV while the new supply reads out ~85uA. This could just be an EPICS read back issue though.

Attachment 1: vacuumbias_13022018.PNG  26 kB  | Show | Hide all