Conditioning Data Sheet,Data Entry,Units,notes,Notes for operator,,,, ,,,,,,,, Operator,Dave Neilson,,,,,,, Elog Number,1050,,,,,,, Target #,SiC#47,,,,,,, Target Oven WO#,45447-1,,,,,,, Heat Shield WO#,45235-23,,,,,,, ,,,,,,,, Conditioning current,,,,,,,, TGHT (default 440),440,A,,,,,, TBHT (default 290),290,A,,,,,, COIL (default 100),100,A,,,,,, ,,,,,,,, Conditioning voltages,,,,,,,, HV BIAS (max),30,kV,,,,,, Einzel Lens (BIAS - 17kV),17,kV,,,,,, Time at max,300,min,,,,,, Final Spark Period,300,min,,"On average, time between sparks at HV MAX",,,, ,,,,,,,, Anode Mode (30kV tune),,,,,,,, Gas (default Argon),Argon,,,,,,, Gas pressure (default 0.5 bar),0.5,bar,,,,,, measure_id,9785,,,tuning script data ID,,,, Anode voltage,100,V,,Local max from tuning script,,,, COIL current,100,A,,Local max from tuning script,,,, FC0 output,1.07E-07,A,,,,,, Mass Scan filename,2025-06-19 SiC#47 Argon Anode mass scan.txt,,,,,,, RPM0A filename,2025-06-19 SiC#47 Argon RPM0B 100nA,,,"right click plot, 'dump to file', enter file name ""YYYY-MM-DD_hh-mm RPMXX.txt""",,,, RPM0B filename,2025-06-19 SiC#47 Argon RPM0A 100nA,,,,,,, RPM6A filename,2025-06-19 SiC#47 Argon Anode RPM6A 33nA,,mass 40,,,,, RPM6B filename,2025-06-19 SiC#47 Argon Anode RPM6B 33nA,,mass 40,,,,, Gasless Mass Scan filename,2025-06-19 SiC#47 Gasless Anode mass scan.txt,,,,,,, ,,,,,,,, ,,,,,,,, Cathode Mode (30kV tune),,,,,,,, Gas (default Argon),Argon,,,,,,, Gas pressure (default 0.5 bar),0.5,bar,,,,,, measure_id,9783,,,tuning script data ID,,,, Cathode voltage,220,V,,Local max from tuning script,,,, COIL current,80,A,,Local max from tuning script,,,, FC0 output,1.13E-07,A,,,,,, Mass Scan filename,2025-06-19 SiC#47 Argon Cathode mass scan.txt,,,,,,, RPM0A filename,2025-06-19 SiC#47 Argon Cathode RPM0A 33nA,,,"right click plot, 'dump to file', enter file name ""YYYY-MM-DD_hh-mm RPMXX.txt""",,,, RPM0B filename,2025-06-19 SiC#47 Argon Cathode RPM0B 33nA,,,,,,, RPM6A filename,2025-06-19 SiC#47 Argon Cathode RPM6A 33nA,,mass 40,,,,, RPM6B filename,2025-06-19 Sic#47 Argon Cathode RPM6B 33nA,,mass 40,,,,, Gasless Mass Scan filename,2025-06-19 SiC#47 Gasless Cathode mass scan.txt,,,,,,, ,,,,,,,, ,,,,,,,, Peak (m/q),Max (A),Mode,,,,,, 12,6.17E-10,ANODE,,,,,, 13,3.86E-11,ANODE,,,,,, 14,1.50E-10,ANODE,,,,,, 15,1.18E-10,ANODE,,,,,, 16,5.63E-10,ANODE,,,,,, 17,8.50E-11,ANODE,,,,,, 18,2.43E-10,ANODE,,,,,, 19,1.07E-11,ANODE,,,,,, 20,3.19E-10,ANODE,,,,,, 23,5.37E-11,ANODE,,,,,, 24,1.37E-11,ANODE,,,,,, 25,2.88E-11,ANODE,,,,,, 26,1.36E-10,ANODE,,,,,, 27,1.92E-10,ANODE,,,,,, 28,3.63E-09,ANODE,,,,,, 29,1.44E-10,ANODE,,,,,, 30,3.03E-11,ANODE,,,,,, 31,1.19E-10,ANODE,,,,,, 36,1.51E-11,ANODE,,,,,, 38,1.90E-11,ANODE,,,,,, 40,2.59E-09,ANODE,,,,,, 39,1.34E-10,ANODE,,,,,, 41,6.93E-11,ANODE,,,,,, 42,2.49E-11,ANODE,,,,,, 43,3.32E-11,ANODE,,,,,, 44,3.23E-10,ANODE,,,,,, 46,1.27E-11,ANODE,,,,,, 45,4.35E-11,ANODE,,,,,, 50,1.32E-11,ANODE,,,,,, 51,1.32E-11,ANODE,,,,,, 52,1.03E-11,ANODE,,,,,, 53,1.12E-11,ANODE,,,,,, 55,1.66E-11,ANODE,,,,,, 56,1.95E-11,ANODE,,,,,, 12,3.35E-10,CATHODE,,,,,, 14,8.50E-11,CATHODE,,,,,, 15,1.03E-11,CATHODE,,,,,, 16,2.85E-10,CATHODE,,,,,, 17,4.49E-11,CATHODE,,,,,, 18,1.28E-10,CATHODE,,,,,, 20,3.03E-10,CATHODE,,,,,, 23,9.28E-11,CATHODE,,,,,, 27,1.90E-11,CATHODE,,,,,, 28,2.05E-09,CATHODE,,,,,, 29,3.66E-11,CATHODE,,,,,, 30,1.03E-11,CATHODE,,,,,, 31,1.12E-11,CATHODE,,,,,, 36,1.12E-11,CATHODE,,,,,, 40,2.32E-09,CATHODE,,,,,, 39,1.12E-10,CATHODE,,,,,, 44,6.15E-11,CATHODE,,,,,,