Message ID: 18     Entry time: Tuesday, February 21, 2012, 09:56
Author: Isaac Earle 
Category: Repair 
Type: Target 2 
Specific: Target 
Subject: Air Supply Check Valve & Q4 Flow Transducer 

 The faulty air pressure supply check valve for M9BB, M20BB, and T2 Profile Monitor was replaced.

The rotor, shaft, and o-ring were replaced on the Q4 flow transducer to address erroneous readout that began last week.
The attached plot shows the flow readout before failure, during the failed state, and after repair.

Attachment 1: T2_Q4_Flow_Transducer.JPG  63 kB  | Hide | Hide all