Message ID: 972     Entry time: Tuesday, June 30, 2015, 15:53
Author: David Wang 
Category: Conditioning Station 
Type: Development 
Module: TM2 
Target/Number:  
Subject: leak on TM2 at CS 

 I leak checked TM2 at CS this afternoon which including air to vacuum check on service cap and helium pressure test on water lines. There is a small leak on water line R ,U (COIL1) during the helium pressure test.. Below 20 psi helium, there is no leak on it. But , once the helium pressure is increased over 20 psi, it showed up on Leak Detector. At 60 psi helium, the leak rate is 9.4xE-9 atm.cc/sec. For detail information, See attachment.  All other water lines and air to vacuum seals are good.

Attachment 1: TM2_leak_check_2015-06-30.pdf  352 kB  | Show | Hide all
Attachment 2: leak_check_2015-06-30_007.JPG  2.313 MB  | Show | Hide all
Attachment 3: leak_check_2015-06-30_009.JPG  2.322 MB  | Hide | Hide all