Message ID: 896
Entry time: Thursday, April 23, 2015, 09:08
|
Author: |
David Wang |
Category: |
Conditioning Station |
Type: |
Maintenance |
Module: |
TM1 |
Target/Number: |
UCx12 |
Subject: |
Huge Leak on TM1 contaiment box window. |
|
|
There is a huge leak on TM1 containment box window. I did a 70 psi helium pressure test on all TM1 water circuits this morning. The LD base leak rate was 0.0XE-9 atm.cc/sec. When applying the 70 psi helium to containment box window circuit, the leak rate on LD jumped to 1XE-4 atm.cc/sec in 5 seconds.I released the helium source right away,t but the conditioning station is full of helium molecules already, and the leak rate on LD stays at 1xE-5 atm.cc /sec. The leak on this circuit is huge. Other water lines are good. |
|
|
|
|
|
|