Message ID: 814     Entry time: Friday, December 05, 2014, 11:58
Author: David Wang 
Category: Conditioning Station 
Type: Development 
Module: TM2 
Target/Number:  
Subject: helium pressure test on TM2 tube heater +(D) water lines. 

At CS, The leak detector is hooked directly to the TM2 top service cap. The base LR 0.0xE-9 atm.cc/sec, base pressure 0.0xE-4torr. Applied 80 psi helium to TM2 tube heater +(D) line . During 3 minutes applying time , I did  not see any tiny change on LD readings. The line is perfectly vacuum sealed.  See attachment pictures.