Message ID: 814
Entry time: Friday, December 05, 2014, 11:58
|
Author: |
David Wang |
Category: |
Conditioning Station |
Type: |
Development |
Module: |
TM2 |
Target/Number: |
|
Subject: |
helium pressure test on TM2 tube heater +(D) water lines. |
|
|
At CS, The leak detector is hooked directly to the TM2 top service cap. The base LR 0.0xE-9 atm.cc/sec, base pressure 0.0xE-4torr. Applied 80 psi helium to TM2 tube heater +(D) line . During 3 minutes applying time , I did not see any tiny change on LD readings. The line is perfectly vacuum sealed. See attachment pictures. |
|
|
|
|
|
|