Tuesday, June 24, 2014, 15:14, Anders Mjos, South Hot-Cell, Standard Operation, TM3, TiC#4, Conductor insulator broken close to right TGHT terminal
|
Wednesday, June 25, 2014, 12:49, Anders Mjos, South Hot-Cell, Standard Operation, TM3, TiC#4, Conductor insulator broken close to right TGHT terminal
|
|
Message ID: 689
Entry time: Wednesday, June 25, 2014, 12:49
In reply to: 687
|
Author: |
Anders Mjos |
Category: |
South Hot-Cell |
Type: |
Standard Operation |
Module: |
TM3 |
Target/Number: |
TiC#4 |
Subject: |
Conductor insulator broken close to right TGHT terminal |
|
|
Anders Mjos wrote: |
When the Containment Box was opened today by Chad for a target change, we found the conductor insulator block B (ITD0054) to be broken close to the right TGHT terminal. Chad was nevertheless able to remove the target. See DocuShare and attached file for photos.
An attempt to replace the insulator will be made tomorrow with a spare.
A small sticky piece, likely off plastic material, was found next to the left main source tray insulator. The piece was removed from the Containment Box. See link above and attachment for photos.
|
The two insulators ITD0053 and ITD0054 was replaced this morning by Chad. Photos on DocuShare. All bolts torqued with torque wrench see attached PDF for values. |
|