Message ID: 625     Entry time: Wednesday, March 19, 2014, 15:46
Author: David Wang 
Category: ITW 
Type: Standard Operation 
Module: TM4 
Target/Number:  
Subject: TM4 and ITE air to vacuum leak check and TM4 water lines helium test. 

With TM4 located in ITW, All air to vacuum seals  on ITW primary and secondary vacuum vessels have been leak  checked with leak detector hooked directly onto the TM4 service cap.The base LR :00xE-9atm.cc/sec. The ITW IG1 was 1.16xE-6 torr which was good. No leak has been found during the checking.Also, the TM4 water lines(except HS) had been  pressurized with 70 psi helium on each for 2 minutes. There was no response during the test. The ITW and TM4 are good on leak checking.See attachment pictures