Message ID: 609     Entry time: Tuesday, February 25, 2014, 10:00
Author: David Wang 
Category: Conditioning Station 
Type: Repair 
Module: TM2 
Target/Number:  
Subject: helium pressure test on TM2 tube heater (+) water line with blank off(C-seal) installed on WB. 

I did the helium pressure  test  on TM2 tube heater (+) water line  with blank off(C-seal) installed on  water block at CS.The leak detector was hooked directly to the top of the module this time. So base leak rate and base pressure were at the bottom scale on LD. 30 -80 psi helium was applied to the D lines(both side)  with10 psi   increasing each time. There is no any tiny response hasbeen found during the test. The lines are perfectly sealed. Also i sprayed the helium on the top of the service cap .All  air to vacuum seals on the top are good either. See attachment pictures.

Attachment 1: 004.JPG  2.359 MB  | Show | Hide all
Attachment 2: 005.JPG  2.197 MB  | Show | Hide all
Attachment 3: 006.JPG  2.843 MB  | Show | Hide all
Attachment 4: 007.JPG  2.384 MB  | Show | Hide all
Attachment 5: 008.JPG  3.614 MB  | Show | Hide all
Attachment 6: 009.JPG  2.664 MB  | Show | Hide all
Attachment 7: 010.JPG  3.421 MB  | Show | Hide all
Attachment 8: 011.JPG  3.088 MB  | Hide | Hide all