Message ID: 609
Entry time: Tuesday, February 25, 2014, 10:00
|
Author: |
David Wang |
Category: |
Conditioning Station |
Type: |
Repair |
Module: |
TM2 |
Target/Number: |
|
Subject: |
helium pressure test on TM2 tube heater (+) water line with blank off(C-seal) installed on WB. |
|
|
I did the helium pressure test on TM2 tube heater (+) water line with blank off(C-seal) installed on water block at CS.The leak detector was hooked directly to the top of the module this time. So base leak rate and base pressure were at the bottom scale on LD. 30 -80 psi helium was applied to the D lines(both side) with10 psi increasing each time. There is no any tiny response hasbeen found during the test. The lines are perfectly sealed. Also i sprayed the helium on the top of the service cap .All air to vacuum seals on the top are good either. See attachment pictures. |
|
|
|
|
|
|
|
|
|
|